3 INTRODUCTION 3.1 AUDIENCE This document is intended for hospital staff, health care system integrators, software designers or implementers. It is assumed that the reader has a working understanding of DICOM. 3.2 REMARKS DICOM, by itself, does not guarantee interoperability. However, the Conformance Statement facilitates a first - level validation for interoperability between different applications supporting the same DICOM func tionality. This Conformance Statement is not intended to replace validation with other DICOM equipment to ensure proper exchange of information intended. The scope of this Conformance Statement is to facilitate communication between Philips Medical Systems and other vendors’ Medical equipment. The Conformance Statement should be read and understood in conjunction with the DICOM Standard [DICOM]. However, by itself it is not guaranteed to ensure the desired interoperability and successful interconnectivity. The user should be aware of the following important issues: — The comparison of different conformance statements is the first step towards assessing interconnectivity between Philips Medical Systems and non - Philips Medical Systems equipment. — Test procedures should be defined to validate the desired level of connectivity. — The DICOM standard will evolve to meet the users’ future requirements. Philips Medical Systems is actively involved in developing the standard further and therefore reserves the right t o make changes to its products or to discontinue its delivery. — This DICOM Conformance Statement reports the implementation of four ultrasound systems, EPIQ 7 and EPIQ 5 , Affiniti 70 and Affiniti 50. Also referred to as E7 and E5; A70 and A50. 3.3 DEFINITIONS, TERMS A ND ABBREVIATIONS Definitions, terms and abbreviations used in this document are defined within the different parts of the DICOM standard. Abbreviations and terms are as follows: AE Application Entity AET Application Entity Title ASCE Association Control Service Element CD- R Compact Disk Recordable CID Context ID. Reference to a DICOM code “context group” defined in DICOM PS3.16 CSE Customer Service Engineer DICOM Digital Imaging and Communications in Medicine FSC File-Set Creator FSU File-Set Updater FSR File-Set Reader GSDF Grayscale Standard Display Function HIPAA United States “Healthcare Insurance Portability and Accountability Act of 1996” E7 and E5/A70 and A50 1.5.x.x DICOM Conformance Statement 000490000000111 Rev A Page 11
IOD Information Object Definition ISO International Standard Organization LOINC Logical Observ ation Identifiers Names and Codes MPPS Modality Performed Procedure Step MW L Modality Worklist R Required Key Attribute O Optional Key Attribute PDIR Patient Directory screen on EPIQ/Affiniti PDU Protocol Data Unit PDE Patient Data Entry screen on EPIQ/Aff initi PS3.x Reference to part ‘x’ the DICOM 3.0 Standard PSC Philips Support Connect. It is the primary interface on EPIQ/Affiniti for configuring DICOM services. It is reached by pressing the “Support” key on the EPIQ/Affiniti control panel. SCU Service Class User (DICOM client) SCP Service Class Provider (DICOM server) SOP Service -Object Pair SNOMED Systematized Nomenclature of Medicine (SRT) SPS Scheduled Procedure Step SR Structured Report TID Template ID. Reference to a DICOM Structured Report “template” defined in DICOM PS3.16 U Unique Key Attribute US Ultrasound USB Universal Serial Bus The following upper -case abbreviations represent specific Attributes. Refer to PS3.3 Section 8 for a description: CV Code Value (0008,0100) CSD Coding Scheme Desi gnator (0008,0102) CM Code Meaning (0008,0104) CSV Coding Scheme Version (0008,0103) 3.4 REFERENCES [DICOM] Digital Imaging and Communications in Medicine (DICOM), NEMA PS 3.1- 3.20, 2011, available free at http://medical.nema.org/standard.html Integrated Healthcare Enterprise (IHE) Radiology Technical Framework, Vol. 1, Integration Profi les, version 11.0 Final Text July 24, 2012 Integrated Healthcare Enterprise (IHE) Radiology Technical Framework, Vol. 2, Transactions, version 11.0 Final Text July 24, 2012 Integrated Healthcare Enterprise (IHE) Radiology Technical Framework, Vol. 3, Transacti ons continued, version 11.0 Final Text July 2 4, 2012 Integrated Healthcare Enterprise (IHE) Cardiology Technical Framework , Vol. 1, Integration Profiles, Revision 4 .0, Final Text, August 5, 2011 E7 and E5/A70 and A50 1.5.x.x DICOM Conformance Statement 000490000000111 Rev A Page 12
Integrated Healthcare Enterprise (IHE) Cardiology Technical Framework, Vol. 2, Transactions, Revision 4.0, Final Text, August 5, 2011 Integrated Healthcare Enterprise (IHE) Cardiology Technical Framework S upplement, Stress Testing Workflow, Trial Implementation Revision 2.1, October 15, 2010 Joint VA / DoD DICOM Conformance Requirements for Digital Acquisition Modalities, Version 3.0, September 28, 2005 E7 and E5/A70 and A50 1.5.x.x DICOM Conformance Statement 000490000000111 Rev A Page 13
4 NETWORKING 4.1 IMPLEMENTATION MODEL 4.1.1 Applicati on Data Flow Send Images, Reports, Commits; Query ImagesStorage Application Entity Update Worklist Acquire Images Workflow Application Entity Film ImagesHardcopy Application Entity DICOM Standard Interface Local AE's Connected AE's Receives Images, Reports, Commits; Returns Commitments Sends images Receives queries Provides Worklist Items Receives MPPS Create / Set Prints Film Sheets E7 and E5/A70 and A50 1.5.x.x DICOM Conformance Statement 000490000000111 Rev A Page 14
Figure 1: APPL IC AT ION DAT A FLOW DI AG R AM — The Storage Application Entity sends Images and Structured Reports to a remote AE. It is associated with the local real -world activity “Acquire” on EPIQ/Affiniti . Sending of images may occur automatically when configured to do so, either “After Each Print/Acquire” or “At End of Exam” w ith or without “Send on Demand”, with separate export destination configurati on of each acquisition control. The system can also be configured to automatically end and send an exam after a specified time of system inactivity. In addition, exam data may be sent manually upon request of the user. If configured to send “After Each Print/Acquire” , images are transferred immediately after acquisition and Structured Reports and Private 3D Presentation States are transferred at the end of the exam. If configured for send “ At End Exam, images and Structured Reports are transferred when the exam is ended. An exam may be sent manually by user selection from the Patient Directory (PDIR), including exams that have been imported or retrieved into EPIQ/Affiniti from other sources. Further, individually -selected images acquired on EPIQ/Affiniti may be sent manually from image rev iew. If the remote AE is configured for Storage Commitment , the Storage AE will request Storage Commitment immediately after the instances are sent to the Storage device associated with the Storage Commitment device. If a commitment response is successful ly obtained for all instances acquired in the exam , this information is recorded in the local database, placing a checkmark in the commit portion of the Patient Directory display and signaling the Auto- delete function that the exam qualifies for deletion. There are several available mechanisms for importing images acquired on other devices, including Ultrasound images and images of other modalities: • When configured, images may be retrieved using the Query/Retrieve service • DICOM images may be imported direct ly from local media • DICOM images may be pushed onto EPIQ/Affiniti by another Storage SCU device without being initiated by the EPIQ/Affiniti user Send on Demand allows for sending acquired images and/or Structured Report during an active exam without ending the exam or in- progress MPPS. Print images will be sent without requiring a full print page to be filled. Send on Demand is invoked by pressing an icon located at the bottom center of the system display after acquiring the first image of the study . — The Workflow Application Entity receives Modality Worklist (MWL ) information from and sends Modality Performed Procedure Step (MPPS) information to remote AEs. It is associated with the local real -world activities “Update W orklist”, “Patient Search” and “Acquire Images”. When either the “Update Worklist” or “Patient Search” local real -world activity is performed, the Workflow Application Entity queries a remote AE for worklist items and provides the set of worklist items matching the query request. ”Update Worklist” is performed as a result of an operator request or can be is performed automatically at specific time intervals. “Patient Search” is manually initiated. Modality Performed Procedure Steps are created and updated with the following real -world events: o MPPS N -Create, Status = IN PROGRESS: Acquisition of images will result in automated creation of an MPPS Instance managed by a remote AE. o MPPS N -Set, Status = COMPLETE Completion of the MPPS is performed as the result of an operator action of ending the exam. o MPPS N -Set, Status = DISCONTINUED “Cancel Exam” causes the “Discontinued” status to be sent. A “Paused” exam does not initiate an MPPS event. Images and SRs may be appended to an ended exam. There are two fundamental methods to perform append: E7 and E5/A70 and A50 1.5.x.x DICOM Conformance Statement 000490000000111 Rev A Page 15
o Append from Patient Directory Select an Ended study from the Patient Directory and press c hoose “Append.” Two options are available: • “Add” (if less than 24 hours old), allows images to be added to the original Study, using the same Study Instance UID, and a new Performed Procedure Step SOP Instance UID. If more than 24 hours old, only “Create” will be available. • “Create”, which will create a new Study Instance UID. o Append from Image Review Open the Study in Review from the Patient Directory. Default behavior (Add to existing study vs. Create new study) is selecta ble via a configuration screen for Append from Image Review – Create Study Instance UID o For Exams appended by creating the image from Image Review and changing visualization of the existing images, for example, changing chroma map, added images would have the same Acquisition DateTime value of the original images in which the images were acquired. This is not the case for append from Patient Directory, which creates completely new images with the current date and time. o Study Date on the Report page refers to the date of an exam in which the evidences were acquired. This is export ed as DICOM Study Date. This includes all Appended and Non- Appended exams. o For Exams appended from Image Review, the Report footer has the statement “Appended: “followed by the date in which the exam is started for Appending/Appended exams from Image Revie w. o SR Vendors shall refer to DICOM Study Date to determine the date of the original exam in which the evidences were acquired and refer to DICOM Performed Procedure Step Start Date for the date of Appended exam. — The Hardcopy Application Entity prints images on a remote AE (Printer or print server). It is associated with the local real -world activity ”Acquire” depending on configuration. The user action creates a Film Session containing one or more Film Boxes and Image Boxes composed from images acquired by the user. The AE creates and sends fully rendered pages according to the user’s selected formatting choices in a single Image Box per page . This Image Box is rather large compared to sending individual Image Boxes to the printer. If the user has both BW and Color DICOM printer s configured and selected, and is using send “After Each Print/Acquire”, the images containing no Color Flow or Chroma data will be sent to the BW printer, all others will be sent to the Color print er. Exam data is sent to all selected Store, Print and Workflow destinations simultaneously in accordance with system configuration of send “After Each Print/Acquire” or “At End of Exam” . If Send on Demand had been invoked, images and Structured Reports modified since the last Send on Demand are sent at End Exam. 4.1.2 Functional Definition of AEs 4.1.2.1 Functional Definition of Storage Application Entity The existence of a Network Store queue with associated network destination will activate the Storage AE. An asso ciation request is sent to the destination AE and upon successful negotiation of a Presentation Context , the image transfer is started. If the association cannot be opened, the related queue is set to a “Failed” state, indicated by a Red dot on the Network Icon, and can be restarted by the user via the queue management interface. The user may need to cancel the queue, and then restart manually. Multiframe (loop) objects will be transferred first, then single frames when configured for End of Exam. When “Im age Export Format” is selected as “monochrome”, single frame images that have no Color Flow Doppler or Chroma maps applied, will export as grayscale using Monochrome2 Photometric Interpretation. Single frame images that include Color Flow Doppler will be s ent as RGB. If the non-active region of a 2D/scrolling image has a chroma map, it will be sent as monochrome. The Storage AE is also responsible for accepting images for storage in EPIQ/Affiniti , though any of the following mechanisms: • Import from removabl e media E7 and E5/A70 and A50 1.5.x.x DICOM Conformance Statement 000490000000111 Rev A Page 16
• Retrieve using the Query/Retrieve service • Unsolicited push from a Storage SCU to EPIQ/Affiniti acting as a Storage SCP. This mechanism requires that the Storage SCU to have previously been configured as an acceptable DICOM Storage Device. 4.1.2.2 Function al Definition of Workflow Application Entity Update Worklist attempts to download a Modality Worklist from a Modality W orklist SCP with studies matching the search criteria. Default query criteria are Modality of US (ultrasound) and Scheduled Procedure S tep Start Date of the current date. Additional changes to Modality Worklist search include customizing to search for a different AE Title, Station Name, System Location and Custom Modality. If the W orkflow AE establishes an Association to a remote AE, it w ill transfer all matching worklist items via the open Association. The results of a successful Worklist Update will overwrite the data in the Worklist display. Specific queries for Patient Last Name, Patient ID, Accession #, Exam Date or Date Range, and Requested Procedure ID may also be performed using the Patient Search. There is no queue management for Modality Worklist. Note: A wildcard (broad) patient query can be performed by entering a * in the Patient Name field. The Workflow AE performs t he creation of a Modality Performed Procedure Step (MPPS) i nstance automatically when the first image of a study is acquired. MPPS message queues are listed along with Image and Structured Report queues in the Network status window. The user may also change patient identification information while a MPPS is in progress. If this is done, the prev ious MPPS is discontinued and a new MPPS is created with the changed patient information. The Storage AE is informed, and if images acquired under the previous patient information have already been transferred (because send “ After Each Print/Acquire” or “Send On Demand” ) the images are regenerated and sent again with updated patient information and new DICOM UID’ s. 4.1.2.3 Functional Definition of Hardcopy Application Entity The existence of a print queue will activate the Hardcopy AE. An association is established with the printer(s) and the printer’s status determined. If the printer is operating normally, the film sheets will be printed. If the printer is not operating normally, the print queue will set to a “Failed” state and can be restarted by the user via the queue management interface. In the case that a user has both a BW and a Color DICOM printer configured, during an exam with send “After Each Print/ Acquire” selected, the images that contain color data, i.e., Color Flow Doppler or Chroma, will be sent to the Color printer only, and all other images sent only to the BW printer. E7 and E5/A70 and A50 1.5.x.x DICOM Conformance Statement 000490000000111 Rev A Page 17
4.1.3 Sequencing of Real -World Activities StorageHardcopyWorkflowDepartment SchedulerPrinterImage Manager 1 Query Worklist 2 Receive Worklist 3 Create MPPS 9 Complete MPPS 5 Print Film Sheet 10 Image and SR Document Commit Request 12 Accept Commitment Response user update or query 7 Store SR Document 4 Store Image acquire first image end exam store completion acquireimage end exam sheet full commit status update 8 Discontinued MPPS cancel exam optional append11 Optional Append or Accept Stage 6 Print Film Sheet end exam acquireimages FIGURE 2A: SEQUENCING CONSTRAINTS – CONFIGURED TO SEND A FTER EACH PRINT/ACQUIRE Figures 2a and 2b illustrate normal scheduled workflow conditions. E7 and E5/A70 and A50 1.5.x.x DICOM Conformance Statement 000490000000111 Rev A Page 18
Notes: • Messages may slightly change order occasionally. • All selected store, print and workflow devices are sent data during the exam when configured for send “After Each Print/ Acquire” or at “At End of Exam.” • Selecting a study from Review for export will send to all selected devices. Other workflow situations (e.g. unscheduled procedure steps) will have other sequencing constr aints. Printing or storage could equally take place after image acquisition. Printing could be omitted completely if no printer is connected or hardcopies are not required. StorageHardcopyWorkflowDepartment SchedulerPrinterImage Manager 1 Query Worklist 2 Receive Worklist 4 Create MPPS 7 Discontinue or Complete MPPS 6 Print All Film Sheets 8 Image and SR Document Commit Request 10 Accept Commitment Response user update or query 5 Store All Images and Report end exam store completion end exam commit status update end exam cancel or end exam optional append 9 Optional Append (no PPS) 3 Acquire Images FIGURE 2B: SEQUENCING CONSTRAINTS – CONFIGURED TO SEN D AT END EXAM E7 and E5/A70 and A50 1.5.x.x DICOM Conformance Statement 000490000000111 Rev A Page 19
Figures 2c Illustrates Flow Changes to Send At End of Exam with Send on Demand (SoD) StorageHardcopyWorkflowDepartmentSchedulerPrinterImageManager 3 Create MPPS 7 Complete MPPS 4 Print Film Sheet 8 Image and SR Document Commit Request 9 Accept Commitment Response 6 Store Final SR Document 1 Acquire Images first storedimage end exam storecompletion end exam sheet full commit statusupdate Send onDemand2 Store Images, SR and Print Film Sheets since last SoD 5 Store Remaining Images, Send Remaining Film Sheet end exam FIGURE 2C: SEQUENCING CONSTRAINTS – SEND ON DEMAND CONFI GURATION E7 and E5/A70 and A50 1.5.x.x DICOM Conformance Statement 000490000000111 Rev A Page 20