Home > Philips > Ultrasound System > Philips Epiq 5 Manual

Philips Epiq 5 Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Philips Epiq 5 Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 619 Philips manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							 Table 4.46  
    FILM SESSION SOP CLASS N -CREATE REQUEST ATTRIBUTES  
    Attribute Name  Tag  VR  Value  
    Presence 
    of Value Source  
    Number of 
    Copies  (2000,0010)
     IS  Default 1.  User defined in Device Configuration.  ALWAYS  USER  
    Print Priority  (2000,0020)  CS HIGH   ALWAYS AUTO  
    Medium Type  (2000,0030)  CS BLUE FILM, CLEAR FILM or PAPER*  ALWAYS USER  
    Film Destination  (2000,0040)  CS MAGAZINE or PROCESSOR*  ALWAYS USER  
    Film Session 
    Label  (2000,0050)
     LO Philips Medical Systems   ALWAYS AUTO  
    *Dependent on the specific printer selected  
    Table  4.47  summarizes the behavior of Hardcopy AE when encountering status codes in an  N -CREATE response.  
    Table  4.47  
    FILM SESSION SOP CLASS N -CREATE  RESPONSE STATUS HANDLING BEHAVIOR  
    Service  Status   Further Meaning
     Error Code   Behavior 
    Success  Success  0000  The SCP has completed the operation 
    successfully.  
     
    Warning  Attribute Value Out of 
    Range
      0116H
      System continues operations.  
    Warning Attribute List Error  0107H  Same as above. 
    *  *  Any other status 
    code.
     The Association is aborted and the print -job fails.   
     
    4.2.3.3.1.4  SOP Specific Conformance for the Film Box SOP Class  
    Hardcopy AE supports the following DIMSE operations for the Film Box SOP Class:  
    —  N-CREATE  
    —  N-ACTION  
     
    Details of the supported attributes and status handling behavior are described in the following subsections.  
    4.2.3.3.1.4.1  Film Box SOP Class Operations (N -CREATE)  
    Table  4.48  lists the attributes s upplied in an N-CREATE Request.  
    E7 and E5/A70 and A50 1.5.x.x    DICOM Conformance Statement 000490000000111   Rev  A   Page 51   
    						
    							 Table 4.48  
    FILM BOX SOP CLASS N -CREATE REQUEST ATTRIBUTES  
    Attribute Name  Tag  VR  Value  
    Presence 
    of Value Source  
    Image Display 
    Format  (2010,0010)
     ST STANDARD
    \1,1  
    2 or  
    CUSTOM \xxx depending on printer.    
    Default is displayed, and is user editable.  Edit 
    only when a valid substitute value is known.  ALWAYS
     AUTO/USER  
    Referenced Film 
    Session 
    Sequence (2010,0500)
     SQ   ALWAYS AUTO  
    >Referenced 
    SOP Class UID  (0008,1150)
     UI  1.2.840.10008.5.1.1.1   ALWAYS AUTO  
    >Referenced 
    SOP Instance 
    UID  (0008,1155)
     UI  From created Film Session SOP Instance  ALWAYS AUTO  
    Film Orientation   (2010,0040)  CS PORTRAIT or LANDSCAPE  ALWAYS USER  
    Film Size ID  (2010,0050)  CS Depends on configuration file selected. 
     
    DICOM Defined Terms plus US_ LETTER.  ALWAYS
     USER  
    Magnification 
    T ype  (2010,0060)
     CS Default Value = NONE, depending on printer  ANAP  AUTO 
    Min Density  (2010,0120)  US Default value displayed, user editable  ALWAYS AUTO/USER  
    Max Density (2010,0130)  US Default value displayed, user editable  ALWAYS AUTO/USER  
    Trim  (2010,0140) CS NO   ALWAYS AUTO  
    Configuration 
    Information  (2010,0150)
     ST Default value displayed, user editable.  Edit only 
    when a valid substitute value is 
    known.  ALWAYS
     AUTO/USER  
     
    Table 4.49  summarizes the behavior of Hardcopy AE when encountering status codes in an  N -CREATE response.   
    Table  4.49  
    FILM BOX SOP CLASS N -CREATE RESPONSE  STATUS HANDLING BEHAVIOR  
    Service  Status   Further Meaning
     Error Code  Behavior 
    Success  Success  0000  The SCP has completed the operation 
    successfully.  
     
    Warning Requested Max Density outside of 
    printer’s operating range
     B605H
      The N
    -CREATE operation is considered 
    successful but the status meaning is logged.    
    *  *  Any other 
    status code.
     The Association is aborted and the job failed.  
     
     
    2 EPIQ pe rforms page compositing of n- up prints (i.e., 2x2, 2,3, etc.) internally and prints the pages as a single Image 
    Box per page; hence STANDARD \1,1  
    E7 and E5/A70 and A50 1.5.x.x    DICOM Conformance Statement 000490000000111   Rev  A   Page 52  
                                                            
    						
    							 4.2.3.3.1.4.2  Film Box SOP Class Operations (N -ACTION)  
    The Hardcopy AE issues an N -ACTION Request to instruct the Print SCP to print the contents of the Film Box.   
    Table  4.50  summarizes the behavior of Hardcopy AE when encountering status codes in an N -ACTION response.  
    Ta ble  4.50  
    FILM BOX SOP CLASS N -ACTION RESPONSE STATUS HANDLING BEHAVIOR 
    Service  Status   Further Meaning
     Error Code  Behavior 
    Success  Success  0000  The SCP has completed the operation 
    successfully.  The film has been accepted for 
    printing.
     
    Warning  Film Box SOP Instance hierarchy does not 
    contain Image Box SOP Instances (empty 
    page)
      B603H
      The Association is aborted and the job is 
    failed.  
     
    Failure Unable to create Print Job SOP Instance; 
    print queue is full.
      C602
      Same as B603H above.    
    *  *  Any other 
    status code.
     Same as B603H above.  
     
     
    4.2.3.3.1.5 SOP Specific Conformance for the Image Box SOP Class  
    Hardcopy AE supports the following DIMSE operations for the Image Box SOP Class:  
    —  N-SET  
     
    Details of the supported attributes  and status handling behavior are described in the following subsections.  
    4.2.3.3.1.5.1  Image Box SOP Class Operations (N -SET)  
    Table  4.51  lists the attributes supplied in an N -SET R equest.  
    Table  4.51  
    IMAGE BOX SOP CLASS N -SET REQUEST ATTRIBUTES 
    Attribute Name  Tag  VR  Value  Presence 
    of Value  Source
     
    Image Position  (2020,0010)  US 1  ALWAYS AUTO  
    Polarity   (2020,0020) CS NORMAL   ALWAYS AUTO  
    Basic Grayscale 
    Image Sequence  (2020,0110)
     SQ Used for BW  (Monochrome2) print   ALWAYS*  AUTO  
    Basic Color Image 
    Sequence   (2020,0111)
     SQ Used for Color (RGB) print  ALWAYS* AUTO  
    >Samples Per Pixel  (0028,0002)  US 1 for Monochrome2 
    3 for RGB
      ALWAYS
     AUTO  
    >Photometric 
    Interpretation  (0028,0004)
     CS MONOCHROME2
     
    RGB    ALWAYS
     AUTO  
    Planar Configuration  (0028,0006)  US Always “01”, only used for RGB print.  ANAP  AUTO 
    E7 and E5/A70 and A50 1.5.x.x    DICOM Conformance Statement 000490000000111   Rev  A   Page 53   
    						
    							 
    >Rows   (0028,0010) US Depends on film size, number of rows for 
    entire sheet of film 
      ALWAYS
     Printer 
    Configuration 
    File
     
    >Columns   (0028,0011) US Depends on film size, number of columns 
    for entire sheet of film
      ALWAYS
     Printer 
    Configuration 
    File
     
    >Bits Allocated  (0028,0100)  US 8  ALWAYS AUTO  
    >Bits Stored  (0028,0101)  US 8  ALWAYS AUTO  
    >High Bit   (0028,0102) US 7  ALWAYS AUTO  
    >Pixel Representation  (0028,0103)  US 0  ALWAYS AUTO  
    >Pixel Data   (7FE0,0010)  OW Pixels of rendered film sheet.  ALWAYS AUTO  
    * Mutually exclusive attributes  
    Table  4.52  summarizes the behavior of Hardcopy AE when encountering status codes in an  N -SET response.  
    Table  4.52  
    IMAGE BOX SOP CLASS N -SET RESPONSE STATUS  HANDLING BEHAVIOR 
    Service 
    Status Further Meaning Error Code  Behavior 
    Success  Success  0000  The SCP has completed the operation 
    successfully.  
     
    Failure  Insufficient memory in printer to store 
    the image.
      C605
      The Association is aborted and the job is 
    failed.  
     
    *  *  Any other status 
    code.
      Same as C605 above.
     
     
    4.2.3.4 Association Acceptance Policy  
    The Hardcopy Application Entity does not accept Associations.  
    4.2.4  Verification Application Entity specification  
    4.2.4.1  SOP Class  
    EPIQ/Affiniti  provides Standard Conformance to the following SOP Class:  
    Table  4.53  
    SOP CLASSES FOR AE VERIFICATION  
    SOP Class Name   SOP Class UID SCU  SCP 
    Verification   1.2.840.10008.1.1  Yes  Yes 
     
    4.2.4.2  Association  Establishment Policy  
    4.2.4.2.1  General  
    The DICOM standard application context name for DICOM 3.0 is always proposed:  
    E7 and E5/A70 and A50 1.5.x.x   DICOM Conformance Statement 000490000000111   Rev  A   Page 54   
    						
    							 Table 4.54  
    DICOM APPLICATION CONTEXT FOR AE VERIFICATION  
    Application Context Name  1.2.840.10008.3.1.1.1 
     
    4.2.4.2.2 Number of Associations  
    EPIQ/Affiniti  initiates one Association at a time for a Verification request.  
    Table  4.55  
    NUMBER OF ASSOCIATIONS INITIATED FOR AE VERIFICATION  
    Maximum number of simultaneous Associations  Up to 10, one for each configured 
    remote device
     
     
     
    Table  4.56  
    NUMBER OF ASSOCIATIONS ACCEPTED FOR AE VERIFICATION  
    Maximum number of simultaneous Associations  Unlimited, however, calli
    ng AE 
    must be already configured in 
    EPIQ/Affiniti . 
     
    4.2.4.2.3  Asynchronous Nature  
    EPIQ/Affiniti  does not support asynchronous communication (multiple outstanding transactions over a single 
    Association).  
    Table 4.57  
    ASYNCHRONOUS NATURE AS A SCU FOR AE VERIFICATION  
    Maximum number of outstanding asynchronous transactions  1 
     
    4.2.4.2.4  Implementation Identifying Information  
    The implementation information for this Application Entity is:  
    Table  4.58  
    DICOM IMPLEMENTATION CLASS AND VERSION FOR AE VERIFICATION  
    Implementation Class UID  1.3.46.670589.14.10000.105 
    Implementation Version Name  EPIQ-AFFIN_1.05  
     
    E7 and E5/A70 and A50 1.5.x.x    DICOM Conformance Statement 000490000000111   Rev  A   Page 55   
    						
    							 
    4.2.4.3 Association Initiation Policy  
    4.2.4.3.1  Activity  – Verify  Remote  SCP 
    4.2.4.3.1.1  Description and Sequencing of Activities  
    The user selecting the “Verify” button on the Device configuration page initiates the verification request to the device 
    whose data has just been configured.  This tool allows the user to ensure all data (AE Title, Port and IP Address) was 
    correctly entered and the remote device may be contacted. It uses C -Echo and verifies the remote device supports all 
    configured SOP Classes.  Any SOP Cl asses requested that are not supported will report “failed”.  Operations may 
    continue, but objects of the type that are not supported will not be exported.  
    EPIQ/Affiniti  initiates an Association in order to issue:  
    —  C-ECHO request according to the Verification SOP Class.  
    1. Open Association
    2. C-ECHO Request
    4. Close Association
    Verification
    SCU
    Verification
    SCP
    3. C-ECHO Response
      
    Figure 8a  
    SEQUENCING OF ACTIVI TY – ISSUE VERIFY  
    4.2.4.3.1.2  Proposed Presentation Contexts  
    EPIQ/Affiniti  will propose Presentation Contexts as shown in the following table:  
       
    E7 and E5/A70 and A50 1.5.x.x    DICOM Conformance Statement 000490000000111   Rev  A   Page 56   
    						
    							 Table 4.59  
    PROPOSED PRESENTATION CONTEXTS FOR ACTIVITY VERIFICATION  
    Presentation Context Table  
    Abstract Syntax  Transfer Syntax   
     
    Role    
    Ext.  
    Neg.  
    Name   UID  Name List  UID List 
    Verification   1.2.840.10008.1.1  Implicit VR Little Endian
     
    Explicit VR Little Endian  1.2.840.10008.1.2
     
    1.2.840.10008.1.2.1  SCU
     None  
     
    4.2.4.3.1.3  SOP Specific Conformance for Verification  
    When Verification of a remote AE is requested in PSC –  Network/DICOM  – DICOM Settings, the Association 
    Negotiation Request message contents  for each DICOM device will include Presentation Contexts for the following 
    SOP Classes:  
    Table 4.60  
    SOP CLASS VERIFICATION BY DICOM DEVICE  
    Device Type SOP Classes Requested Additional Notes 
    DICOM Storage Server US Image Storage  
    US Image Storage (Retired)  
    US Multiframe Storage  
    US Multiframe Storage (Retired)  
    Selected  non-ultrasound image storage  
    Philips  Private 3D Presentation State  
    Comprehensive  SR Storage 
    Storage Commitment Push Model  
    Study Root Query Ret rieve-FIND  
    Study Root Query Retrieve -MOVE  
    Verification   If SR export to a Storage Server 
    is 
    desired  & supported,  select Export 
    SR  before  verification.  
    Storage Commitment requires 
    configuration of  both a Storage 
    server and a Storage Commitment 
    server, and association of the 
    Storage server to the Storage 
    Commitment server in the PSC –  
    Network/DICOM  – DICOM Selection 
    page.  
    DICOM Commit Server  Storage Commitment  Push Model  
    Verification   Storage Commitment requires 
    configuration of both a Storage 
    server and a 
    Storage Commitment 
    server, and association of the 
    Storage server to the Storage 
    Commitment server in the PSC –  
    Network/DICOM  – DICOM Selection 
    page.  
    DICOM PPS Server  Modality Performed Procedure Step  
    Verification   To activate this server, it must be 
    select
    ed in the PSC –  
    Network/DICOM  – DICOM Selection 
    page.  
    DICOM Worklist Server  Modality Worklist  
    Verification   To activate this server, it must be 
    selected in the PSC –
     
    Network/DICOM  – DICOM Selection 
    page.  
    E7 and E5/A70 and A50 1.5.x.x    DICOM Conformance Statement 000490000000111   Rev  A   Page 57   
    						
    							 
    DICOM Structured Report Server Comprehensive Struct ured Report Storage 
    Verification   Configuration of this device is only 
    required if SRs are to be 
    exported to 
    a  dedicated Structured Report 
    Server  Storage SCP . 
    To activate this server, it must be 
    selected in the PSC –  
    Network/DICOM  – DICOM Selection 
    page.  
    DICOM BW Printer  Basic Grayscale META Print  
    Verification   All 
    BW  printers configure this entry.  
    If the printer supports both BW and 
    Color, then this must be configured 
    to allow BW on that printer.  
    DICOM Color Printer Basic Color META Print  
    Verification   May
     be the same printer if color is 
    also supported.  
     
    Table 4.61  summarizes the behavior of EPIQ/Affiniti  when receiving status codes in a C-ECHO response.    
    Table  4.61  
    VERIFICATION C -ECHO RESPONSE STATUS HANDLING BEHAVIOR  
    Service  Status  Further Meaning
     Error Code  Behavior 
    Success    0000  Device Status is set to: Verified   
    Refused Out of Resources  A700  Device Status is set to:  Not Verified  
    Failed Unable to Process  C000 – CFFF   Device Status is set to: Not Verified  
    *  *  Any other status code. Device Status is set to: Not Verified  
     
    Results of a user-initiated Verify are reported as one of three outcomes:  
    •   Successfully Verified  – All SOP Classes proposed by  EPIQ/Affiniti were accepted AND   
    •   Verify Failed  – Server could not be reached OR remote AE is not running OR no SOP Classes proposed by 
    EPIQ/Affiniti  were accepted OR C -ECHO did not return “Success”  
    •   Partia lly Verified –  Verification SOP Class and at least one other proposed SOP Class were accepted AND C -
    ECHO returned “Success”   AND at least one proposed SOP Class was not accepted  
     
    4.2.4.4  Association Acceptance Policy  
    4.2.4.4.1  Activity  – Verification By Remote AE  
    4.2.4.4.1.1  Description and Sequencing of Activities  
    The system listens on the port configured on the Global System Configuration screen for Verification requests initiated 
    by other remote devices.  The calling device AE must already be configured as a remote device i n EPIQ/Affiniti  or the 
    association is rejected.  
    E7 and E5/A70 and A50 1.5.x.x    DICOM Conformance Statement 000490000000111   Rev  A   Page 58   
    						
    							 
    1. Open Association2. C-ECHO Request
    4. Close Association
    Verification  SCPVerification SCU
    3. C-ECHO Response
     
    Figure 8b  
    SEQUENCING OF ACTIVI TY – VERIFICATION BY REMO TE AE 
    4.2.4.4.1.2  Accepted Presentation Contexts  
    EPIQ/Affiniti  will accept  Presentation Contexts as shown in the following table:  
    Table  4.62  
    ACCEPTABLE PRESENTAT ION CONTEXTS FOR  
    ACTIVITY VERIFICATIO N BY REMOTE AE 
    Presentation Context Table  
    Abstract Syntax  Transfer Syntax   
     
    Role    
    Ext.  
    Neg.  
    Name   UID  Name List  UID List 
    Verification   1.2.840.10008.1.1  Implicit VR Little Endian
     
    Explicit VR Little Endian  1.2.840.10008.1.2
     
    1.2.840.10008.1.2.1  SCP
     None  
     
    4.2.4.4.1.3  SOP Specific Conformance for Verification  
    The association request will be rejected if the remote AE is not configured on  EPIQ/Affiniti 
    4.3   NETWORK INTERFACES  
    4.3.1  Physical Network Interface 
    EPIQ/Affiniti  supports both wired and wireless  network interface as follows: 
    Table  4.63  
    SUPPORTED PHYSICAL NETWORK INTERFACE  
    Ethernet 10/100/1000  BaseT, RJ-45,  UTP, STP; AutoDetect Speed, Full or Half Duplex  
    Wireless 802.11 G/N using optionally provided wireless adapter only  
    E7 and E5/A70 and A50 1.5.x.x   DICOM Conformance Statement 000490000000111   Rev  A   Page 59   
    						
    							 
     
    If both wired and wireless networks are configured, EPIQ/Affiniti will select the best available network automatically as 
    followed:  
    •   The  Wired network has priority and it will be used if available  
    •   The Wireless network will be automatically used if the wired network is not available.  
    4.3.2  Additional Protocols  
    EPIQ/Affiniti  supports the following additional protocols:  
    •   Dynamic Host Configuration Pr otocol (DHCP):  If enabled by the user, DHCP is used to acquire network 
    configuration parameters, such as IP address, subnet mask, gateway address, and DNS server addresses 
    from the DHCP server.  
    •   Domain Name System (DNS):  If configured, DNS is used to look up server network addresses specified by 
    hostname 
    •   Wi-Fi Protected Access II (WPA2) Personal (PSK):  Wireless encryption method with shared- key 
    authentication. Provides reasonable wireless security in small to medium wireless network environments.  
    •   Wi-Fi Pro tected Access II (WPA2) Enterprise:  W ireless encryption method with centralized authentication. 
    Provides very good wireless security in medium to large wireless network environments with centralized 
    authentication server support.  
    4.3.3  IPv4 and IPv6 Support  
    EPI Q/Affiniti  supports IPv4 and IPv6 internet protocol s. 
    When IPv4 is used  
    •   EPIQ/Affiniti’s network address may be specified as either static IP addresses or DHCP  
    •   Remote servers’ network addresses may  be specified as either IP address or DNS hostname 
    When  IPv6 is used  
    •  EPIQ/Affiniti’s network address must be generated automatically as link local address, router discovery 
    address, or assigned by DHCP.  Static IPv6 address entry is not  supported.  
    •  Remote server’ network addresses must be specified by DNS hostname; fixed IPv6 address entry is not  
    supported. 
    •  IPv6 versions of DHCP client, DNS client, ICMP, etc. are supported while IPv6 is being used.  
    •   IPv6 may be disabled if name resolution to an IPv4 device is problematic in a dual -protocol IPv4/IPv6 network.  
     
    4.4   CONFIGURATION   
    4.4.1  AE Title/Presentation Address Mapping  
    Philips Support Connect, or PSC, is the primary interface on  EPIQ/Affiniti for configuring DICOM services.  It is 
    reached by pressing the “Support” key on the  EPIQ/Affiniti control panel.  
    Additional acquisit ion and export settings are configured in the Setups interface.  This interface is reached by selecting 
    Utilities  – Setups  – Acquisition/Capture.  Ensure correct settings are made in “Archive/Print” to associate the selected 
    Acquisition Button(s) to the sel ected configured devices and continue to the Touchscreen button selections as well.  
    Please note to check each touchscreen button individually.  
    4.4.1.1  Local AE Title  
    All local AEs use the same AE Title and TCP/IP Port configured via the  PSC – Network/DICOM  – DICO M Settings 
    screen. The system listens on the configured Port for Verification requests  from remote AE’s,  Storage Commitment N -
    E7 and E5/A70 and A50 1.5.x.x    DICOM Conformance Statement 000490000000111   Rev  A   Page 60   
    						