    							 Attribute Name  Tag  VR  Value 
    Presence 
    of Value Source  
    Time(s) 
    Event Timer 
    Name(s)   (0008,2132)
     LO Timer
     
    ANAP  AUTO 
    View Code 
    Sequence  (0054
    ,0220)  SQ 
    Sequence that describes the view of the patient 
    anatomy in this image.  
    Only a single Item shall be permitted in this  
    Sequence. 
      
    Acquisition 
    DateTime  (0008,002A)
     DT The date and time that the acquisition of data 
    that resulted in this image started.
     ALWAYSU
    A
      AUTO
     
    Trigger Time  (0018,1060)  DS nnn msec.
     
    ANAP  AUTO 
    Heart Rate  (0018,1088)  IS  Beats per minute   ANAP  AUTO 
    Transducer Data  (0018,5010)  LO Transducer name   ALWAYS AUTO  
    Processing 
    Function   (0018,5020)
     LO Imaging optimization name.    ALWAYS AUTO  
    * View name and View Number attributes are also in use by General Imaging Protocol   
     
    Table  9.17  
    VOI LUT MODULE OF CREATED US SOP INSTANCES  
    Attribute Name  Tag  VR  Value  
    Presence 
    of Value Source  
    Window Center  (0028,1050)  DS Fixed  at  127   ANAP  AUTO 
    Window W idth  (0028,1051) DS Fixed  at  254   ANAP  AUTO 
     
     
    Table  9.18  
    SOP COMMON MODULE OF CREATED US IMAGE OR US MULTIFRAME IMAGE SOP INSTANCES  
    Attribute  Name   Tag
      VR  Value   Presence 
    of Value  Source
     
    SOP Class 
    UID  (0008,0016)
     UI  1.2.840.10008.5.1.4.1.1.6.1 or 
    1.2.840.10008.5.1.4.1.1.6 for US Image
     
    1.2.840.10008.5.1.4.1.1.3.1 or 
    1.2.840.10008.5.1.4.1.1.3 for US Multiframe Image  ALWAYS
     AUTO  
    SOP Instance 
    UID  (0008,0018)
     UI  Generated by device   ALWAYS AUTO  
    Specific 
    Character Set  (0008,0005)
     CS  Attribute only sent if an Extended or Replacement 
    Character Set is used
      ANAP
      AUTO 
    Instance 
    Creation Date  (0008,0012)
     DA    ALWAYSUA AUTO 
    Instance 
    Creation Time  (0008,0013)
     TM     ALWAYS AUTO  
     
    E7 and E5/A70 and A50 1.5.x.x    DICOM Conformance Statement 000490000000111   Rev  A   Page 81   
    						
    							 
    9.1.1.5.3 Secondary Capture Image Modules 
    Table 9.19  
    GENERAL IMAGE MODULE OF CREATED SECONDARY CAPTURE  SOP INSTANCES 
    Attribute Name  Tag  VR  Value   Presence of 
    Value   Source
     
    Instance 
    Number**  (0020,0013)
     IS  Generated by device, increments from  “1” in 
    each 
    Secondary Capture  series  ALWAYS
     AUTO  
    Content Date  (0008,0023)  DA     ALWAYSUA AUTO  
    Content Time  (0008,0033)  TM     ALWAYS AUTO  
    Image Type  (0008,0008)  CS  DERIVED/PRIMARY/   ALWAYS CONFIG  
    Acquisition 
    DateTime  (0008,002A)
     DT  The date and time that the acquisition of data 
    that resulted in this image started.
     ALWAYSUA
     AUTO  
    Derivation 
    Description  (0008,2111)
     ST  “Uncompressed” for 
    SC Image or “Low”, 
    “Medium” or “High” for USMF Image based on 
    configuration setting   ALWAYS
     AUTO  
    Burned  In 
    Annotation  (0028,0301)
     CS  See SC Multi -Frame Image Module Table  9.22    ALWAYS  AUTO  
    Lossy Image 
    Compression  (0028,2110)
     CS   “01” if image is lossy compressed, “00” if not.  ALWAYS  AUTO  
    Lossy Image 
    Compression 
    Ratio  (0028,2112)
     DS  For Lossy Compressed images, the following is 
    sent:
     
    Lossy Low = 100  
    Lossy Med = 95  
    Lossy High = 90   ANAP
      AUTO 
    Lossy Image 
    Compression 
    Method   (0028,2114)
     CS  For Lossy Compressed images, the attribute is 
    set to “ISO_10918_1”
      ANAP
      AUTO 
     
    Table 9.20  
    IMAGE PIXEL MODULE OF CREATED  SECONDARY CAPTURE SOP INSTANCES  
    Attribute Name  Tag  VR  Value   Presence of 
    Value   Source
     
    Samples per Pixel  (0028,0002)  US  Set to 1 for Secondary Capture Images,              
    Set to 3 for Multi-
    frame True Color SC Images  ALWAYS
     AUTO  
    Photometric 
    Interpretation  (0028,0004)
     CS  Uncompressed: “MONOCHROME2” or “RGB”
     
    Compressed:     “YBR_FULL_422”  ALWAYS
     AUTO  
    Rows  (0028,0010) US  Image height in pixels
    : 
      983 or 1050 for Secondary Capture Images  
      983 for Multi -frame True Color SC Images  ALWAYS
     CONFIG  
    Columns   (0028,0011)  US  Image width in pixels:
     
      1593 or 1920 for Secondary Capture Images  
      1593 for Multi -frame True Color SC Images  ALWAYS
     CONFIG  
    Bits Allocated  (0028,0100)  US  8 Bits per pixel.   ALWAYS AUTO  
    Bits Stored  (0028,0101)  US  8 Bits per pixel.   ALWAYS AUTO  
    E7 and E5/A70 and A50 1.5.x.x    DICOM Conformance Statement 000490000000111   Rev  A   Page 82   
    						
    							 Attribute Name  Tag  VR  Value  
    Presence of 
    Value Source  
    High Bit   (0028,0102)  US  High bit is 7   ALWAYS AUTO  
    Pixel 
    Representation  (0028,0103)
     US  “0” pixels are Unsigned I ntegers  ALWAYS AUTO  
    Pixel Data  (7FE0,0010)  OW  / 
    OB
       
    ALWAYS AUTO  
    Planar 
    Configuration  (0028,0006)
     US  Present  when image is RGB. Value is “0”.   ANAP  AUTO 
     
     
     
    Table 9.21  
    SC EQUIPMENT  MODULE OF CREATED  SECONDARY CAPTURE SOP INSTANCES  
    Attribute 
    Name Tag  VR  Value  Presence 
    of Value Source  
    Conversion 
    T ype  (0008,0064)
     CS  WSD = Workstation  ALWAYS AUTO  
    Modality  (0008,0060)  CS  Determined by the image modality being viewed at 
    the time of acquisition.
      ALWAYS
     AUTO  
     
    Table 9.22  
    SC MULTI -FRAM E  IM AG E MODULE OF CREATED  SECONDARY CAPTURE  SOP INSTANCES 
    Attribute 
    Name Tag  VR  Value  Presence 
    of Value Source  
    Burned  In 
    Annotation  (0028,0301)
     CS  Set to “
    NO”. Set to “YES” if “Burn Patient Information 
    Into Images” is set.    ALWAYS
     AUTO  
    Frame 
    Increment 
    Pointer   (0028,0009)
     AT 0x00181063:   Frame Time only  ALWAYS AUTO  
     
    Table 9.23  
    SOP COMMON MODULE OF CREATED SECONDARY CAPTURE  SOP INSTANCES  
    Attribute  Name   Tag
      VR  Value   Presence 
    of Value  Source
     
    SOP Class 
    UID  (0008,0016)
     UI  1.2.840.10008.5.1.4.1.1.7 for Secondary Capture 
    Image 
    1.2.840.10008.5.1.4.1.1.7.4 for Multi -frame True 
    Color Secondary Capture Image  ALWAYS
     AUTO  
    SOP Instance 
    UID  (0008,0018)
     UI  Generated by device   ALWAYS AUTO  
    Specific 
    Character Set  (0008,0005)
     CS  Attribute only sent if an Extended or Replacement 
    Character Set is used
      ANAP
      AUTO 
    E7 and E5/A70 and A50 1.5.x.x    DICOM Conformance Statement 000490000000111   Rev  A   Page 83   
    						
    							 
    Attribute 
    Name Tag  VR  Value  Presence 
    of Value Source  
    Instance 
    Creation Date  (0008,0012)
     DA    ALWAYSUA AUTO 
    Instance 
    Creation Time  (0008,0013)
     TM     ALWAYS AUTO  
     
     
    9.1.1.5.4  Comprehensive Structured Report Modules  
     
    Table  9.24  
    SR DOCUMENT SERIES MODULE OF CREATED COMPREHENSIVE SR SOP INSTANCES  
    Attribute Name  Tag  VR  Value   Presence 
    of Value  Source
     
    Modality   (0008,0060)  CS “SR”   ALWAYS AUTO  
    Series Instance 
    UID  (
    0020,000E )  UI  Auto -generated   ALWAYS AUTO  
    Series Number  (0020,0011)  IS  A number unique within the Study  ALWAYS AUTO  
    Referenced 
    Performed 
    Procedure Step 
    Sequence  (0008,1111)
     SQ Identifies the MPPS SOP Instance to which this 
    image is related
      ALWAYS
     MPPS  
    >Referenced 
    SOP Class UID  (0008,1150)
     UI  PPS SOP Class = “1.2.840.10008.3.1.2.3.3”  ALWAYS  MPPS  
    > Referenced 
    SOP Instance 
    UID  (0008,1155)
     UI  PPS Instance UID of the PPS generating this 
    document 
      ALWAYS
     MPPS  
     
     
    Table  9.25  
    SR DOCUMENT GENERAL MODULE OF CREATED COMPREHENSIVE SR SOP INSTANCES  
    Attribute Name  Tag  VR  Value   Presence of 
    Value  Source
     
    Instance Number  (0020,0013)  IS   Unique number    ALWAYS AUTO  
    Completion Flag  (0040,A491)  CS  PARTIAL   ALWAYS AUTO  
    Verification Flag  (0040,A493)  CS  UNVERIFIED   ALWAYS AUTO  
    Content Date  (0008,0023)  DA  Date content created.   ALWAYSUA AUTO  
    Content Time  (0008,0033)  TM  Time content created.   ALWAYS AUTO  
    Predecessor 
    Documents Sequence  (0040,A360)
     SQ  Used when Send of Demand is used and Send 
    Structured Reports with Send on Demand is 
    selected.  Will indicate the preceding SR sent in 
    the study.
      ANAP
      AUTO 
    >Study Instance UID  (0020,000D ) UI  Study’s UID   ANAP  AUTO 
    E7 and E5/A70 and A50 1.5.x.x    DICOM Conformance Statement 000490000000111   Rev  A   Page 84   
    						
    							 Attribute Name  Tag  VR  Value  
    Presence of 
    Value Source  
    >Referenced Series 
    Sequence   (0008,1115)
     SQ  Identifies the Series containing the referenced SR  ALWAYS  AUTO  
    >>Referenced SOP 
    Sequence   (0008,1199)
     SQ  SOP Instance UID for SR Series in the study  ALWAYS  AUTO  
    >>> Referenced SOP 
    Class  (0008,1150)
     UI   Comprehensive SR SOP Class
     
    1.2.840.10008.5.1.4.1.1.88.33  ALWAYS
     AUTO  
    >>> Referenced SOP 
    Instance UID  (0008,1155)
     UI   SOP Instance UID of the preceding SR in the study  ALWAYS  AUTO  
    Referenced Request 
    Sequence   (0040,A370)
     SQ  Identifies Requested Procedures being  
    fulfilled (completely or partially) by creation of this 
    Document. 
    ANAP   AUTO 
    >Study Instance UID  (0020,000D ) UI  Same value as in MWL or auto generated  ALWAYS MW L/
     
    AUTO  
    >Referenced Study 
    Sequence   (0008,1110)
     SQ  1 item per item in MW L, zero length if  unscheduled ANAP VNAP MW L  
    >>Referenced SOP 
    Class UID   (0008,1150)
     UI   Identifies the Referenced SOP Class  ANAP  MW L 
    >>Referenced SOP 
    Instance UID  (0008,1155)
     UI   Instance UID   ANAP  MW L 
    >Accession Number  (0008,0050)  SH  Same attribute of MWL or user PDE input.  VNAP  MW L/
     
    USER  
    >Placer Order 
    Number/Imaging 
    Service Request  (0040,2016)
     LO  Order Number of Imaging Service Request 
    assigned by placer
      VNAP
      MW L 
    >Filler Order 
    Number/Imaging 
    Service Request  (0040,2017)
     LO  Order Number of Imaging Service Request 
    assigned by
     filler   VNAP
      MW L 
    >Requested 
    Procedure ID  (0040,1001)
     SH  1 item per item in MW L, absent if unscheduled  ANAP  MW L 
    >Requested 
    Procedure 
    Description   (0032,1060)
     LO  1 item per item in MW L, absent if unscheduled  ANAP  MW L 
    >Requested 
    Procedure Code 
    Sequence   (0032,1064)
     SQ  1 item per item in MW L, zero length if unscheduled  VNAP   MW L 
    Performed Procedure 
    Code Sequence  (0040,A372)
     SQ  Codes of the performed procedure, zero length if 
    unscheduled
      VNAP
      AUTO/
     
    MW L  
     
     
     
     
    E7 and E5/A70 and A50 1.5.x.x   DICOM Conformance Statement 000490000000111   Rev  A   Page 85   
    						
    							 Table 9.26  
    SR DOCUMENT CONTENT MODULE OF CREATED COMPREHENSIVE SR SOP INSTANCES  
    Attribute Name  Tag  VR  Value  
    Presence 
    of Value Source  
    Content Template 
    Sequence   (0040,A504)
     SQ    ALWAYS AUTO  
    >Template Identifier  (0040,DB00)  CS  The Root Content Item identifies TID 5000 (OB
    -
    GYN),  5100 (Vascular). 5200 (Echo),  TID 5220 or 
    995300 (Ped Echo) , TID 995400 (Small Parts)  ALWAYS
     AUTO  
    >Mapping Resource  (0008,0105)  CS  DCMR   ALWAYS AUTO  
    Content Sequence  (0040,A730)  SQ    ALWAYS AUTO  
    >Relationship  Type  (0040,A010)  CS  Template ID 5000 for OB
    -GYN  
    Template ID 5100 for Vascular  
    Template ID 5200 for Adult Echo  
    Template ID 5220 for Ped,Fetal,Congenital  
    Template ID 995300 for private Ped   
    Template ID 995400 for private Small Parts  ALWAYS
     AUTO  
    Document  Relationship 
    Macro Table    
     Template ID 5000 for OB
    -GYN  
    Template ID 5100 for Vascular  
    Template ID 5200 for Adult Echo  
    Template ID 5220 for Ped,Fetal,Congenital  
    Template ID 995300 for private Ped  
    Template ID 995400 for private Small Parts  ANAP
     AUTO  
    Document  Content 
    Macro    
     Template ID 5000 for OB
    -GYN  
    Template ID 5100 for Vascular  
    Template ID 5200 for Adult Echo  
    Template ID 5220 for Ped,Fetal,Congenital  
    Template ID 995300 for private Ped  
    Template ID 995400 for private Small Parts  ALWAYS
     AUTO  
    Value Type   (0040,A040) CS  CONTAINER, always first attribute 
    of SR 
    Additional values used:  
    TEXT, NUM, DATE, CODE, IMAGE, SCOORD  ALWAYS
     AUTO  
    Concept Name Code 
    Sequence   (0040,A043)
     SQ    ALWAYS AUTO  
    >Code Value   (0008,0100)   125000, 125100,
     125200,  125195, 995300, or 
    995400   ALWAYS
     AUTO  
    >Coding Scheme 
    Designator   (0008,0102)
       DCM   ALWAYS AUTO  
    >Code Meaning  (0008,0104)    “OB
    -GYN Ultrasound Procedure Report”,  
    “Vascular Ultrasound Procedure Report”,  
    “Adult Echocardiography Procedure Report”  
    “Pediatric Ultrasound Procedure Report”  
    “ Pediatric Echocardiography Procedure Report ” 
    or “Small Parts Ultrasound Procedure Report”   ALWAYS
     AUTO  
    Continuity of Content  (0040,A050)  CS  SEPARATE   ALWAYS AUTO  
    Spatial Coordinates 
    Macro   
     If and only if Value Type (0040,A040) = SCOORD  ANAP  AUTO  
    E7 and E5/A70 and A50 1.5.x.x    DICOM Conformance Statement 000490000000111   Rev  A   Page 86   
    						
    							 Attribute Name  Tag  VR  Value  
    Presence 
    of Value Source  
    Graphic Data  (0070,0022)   Order set of Column \
     Row pairs defining positions 
    in a reference image   ANAP
     AUTO  
    Graphic Type  (0070,0023)   Polyline   ANAP AUTO  
    Numeric Measurement 
    Macro   
     
    Template ID 5000 for OB-GYN 
    Template ID 5100 for Vascular  
    Template ID 5200 for Adult Echo  
    Template ID 5220 for Ped,Fetal,Congenital  
    Template ID 995300 for private Ped  
    Template ID 995400 for private Small Parts 
    ALWAYS AUTO  
    Code Macro  
       
    Template ID 5000 for OB-GYN 
    Template ID 5100 for Vascular  
    Template ID 5200 for Adult Echo  
    Template ID 5220 for Ped,Fetal,Congenital  
    Template ID 995300 for private Ped  
    Template ID 995400 for private Small Parts  ALWAYS
     AUTO  
     
     
    Table  9.27  
    SOP COMMON MODULE OF CREATED CO MPOSITE SR SOP INSTANCES 
    Attribute Name  Tag  VR  Value   Presence 
    of Value  Source
     
    SOP Class UID  (0008,0016)  UI  1.2.840.10008.5.1.4.1.1.88.33  ALWAYS AUTO  
    SOP Instance UID  (0008,0018)  UI  Generated by device   ALWAYS AUTO  
    Specific Character 
    Set  (0008,0005)
     CS “None”, unless required by characters used  ALWAYS  CONFIG  
     
    9.1.2 Usage of Attributes from Received IOD’s  
    The attributes of images imported to  EPIQ/Affiniti are stored with level 2 conformance, are used in the display of these 
    images, and are exported if the imported images are exported from EPIQ/Affiniti . 
    The usage of attributes received  via Modality W orklist C-FIND is described in  Section 4.2.2.3.1.3 .   
    9.1.3  Attribute Mapping  
    Table  9.28  summarizes the relationships between attributes received via MWL, stored in acquired images and 
    communicated via MPPS.  The format and conventions used in Table  9.28  are the same as the corresponding table in 
    I HE Technical Framework, Rev. 11.0 07- 24-2012, vol. II: Transactions.  
       
    E7 and E5/A70 and A50 1.5.x.x    DICOM Conformance Statement 000490000000111   Rev  A   Page 87   
    						
    							 Table 9.28  
    ATTRIBUTE MAPPING  BETWEEN MODALITY WORKLIST, IMAGE AND MPPS  
    Modality Worklist Image IOD MPPS IOD 
    Patient’s Name Patient’s Name Patient’s Name 
    Patient ID Patient ID Patient ID 
    Patient’s Birth Date Patient’s Birth Date Patient’s Birth Date 
    Patient’s Sex Patient’s Sex Patient’s Sex 
    Patient’s W eight Patient’s W eight  
    Referring Physician’s Name Referring Physician’s Name  
    ----  ---- Scheduled Step Attributes 
    Sequence 
    Study Instance UID Study Instance UID >Study Instance UID 
    Referenced Study Sequence Referenced Study Sequence >Referenced Study Sequence 
    Accession Number Accession Number >Accession Number 
    ---- Request Attributes Sequence ---- 
    Requested Procedure ID >Requested Procedure ID >Requested Procedure ID 
    Requested Procedure 
    Description 
    >Requested Procedure 
    Description 
    >Requested Procedure 
    Description 
    Scheduled Procedure Step ID >Scheduled Procedure Step ID >Scheduled Procedure Step ID 
    Scheduled Procedure Step 
    Description  
    >Scheduled Procedure Step 
    Description 
    > Series Description  
    > Performed Procedure Step 
    Description 
    >Scheduled Procedure Step 
    Description  
    Scheduled Protocol Code 
    Sequence 
    >Scheduled Protocol Code 
    Sequence ---- 
    ---- Performed Protocol Code 
    Sequence 
    Performed Protocol Code 
    Sequence 
    ---- Study ID – Requested Procedure 
    ID from MWL, else generated 
    Study ID – Requested Procedure 
    ID from MWL, else generated 
    ----  Performed Procedure Step ID Performed Procedure Step ID 
    ---- Performed Procedure Step Start 
    Date 
    Performed Procedure Step Start 
    Date 
    ---- Performed Procedure Step Start 
    Time 
    Performed Procedure Step Start 
    Time 
    ---- Performed Procedure Step 
    Description 
    Performed Procedure Step 
    Description 
    ---- ---- Performed Series Sequence 
    Requested Procedure Code 
    Sequence Procedure Code Sequence  Procedure Code Sequence  
    ---- Referenced Performed 
    Procedure Step Sequence ---- 
    ---- >Referenced SOP Class UID SOP Class UID 
    ---- >Referenced SOP Instance UID SOP Instance UID 
    ---- Protocol Name Protocol Name 
     
    9.1.4  Coerced/Modified Fields  
    The MWL AE will truncate attribute values received in the response to a MW L Query if the value length is longer than 
    the maximum length permitted by the attribute’s VR.  
    E7 and E5/A70 and A50 1.5.x.x   DICOM Conformance Statement 000490000000111   Rev  A   Page 88   
    						
    							 
    9.1.5 Attribute  Anonymization  
    The  EPIQ/Affiniti  supports basic anonymization in that a set of identifying DICOM  attributes that are stored to media 
    wil l be either removed or replaced with non- identifying data, according to the HIPAA guidelines for anonymization.  
    The option to de- identify is available at the time of export for the manual send to media workflows.  
    Table  9.29  contains a list of attributes containing patient identification and how  EPIQ/Affiniti processing the attribute 
    value when DICOM de- identification is selected on export.  This table i s derived from DICOM 2007 PS3.15 Annex E, 
    Table E.1- 1.  N ote most attributes  are labeled ‘Keep’ because they are not classified as identifying by  the United 
    States Health Insurance Portability and Accountability Act of 1996 (HIPAA) Safe- Harbor De-Identification Guidelines . 
    Table  9.29  
    DICOM ATTRIBUTES CONTAINING PATIENT IDENTIFYING INFORMATION 
    Attribute Name   Tag  Delete / 
    Replace?  
    Instance Creation Date   (0008,0012) Delete 
    Instance Creation Time  (0008,0013) Delete 
    SOP Instance UID   (0008,0018) Replace  
    Study Date   (0008,0020) Replace  
    Series Date   (0008,0021) Delete 
    Content Date   (0008,0023) Delete 
    Acquisition DateTime   (0008,002A) Delete 
    Study Time   (0080,0030) Replace  
    Series Time  (0008,0031) Delete 
    Content  Time  (0008,0033) Delete 
    Accession Number   (0008,0050) Delete 
    Institution Name   (0008,0080) Delete 
    Referring Physician’s Name  (0008,0090) Delete 
    Code Meaning   (0008,0104) Delete 
    Station Name  (0008,1010) Delete 
    Study Description   (0008,1030) Delete 
    Series Description   (0008,103E) Delete 
    Physicians of Record   (0008,1048) Delete 
    Operators’ Name  (0008,1070) Delete 
    Admitting Diagnosis Description  (0008,1080) Delete 
    Referenced SOP Instance UID  (0008,1155) Replace  
    Derivation Description   (0008,2111) Delete 
    Patient’s Name  (0010,0010) Replace  
    Patient ID   (0010,0020) Replace  
    Patient’s Birth Date   (0010,0030) Delete 
    Patient’s Sex   (0010,0040) Delete 
    Other Patient IDs   (0010,1000) Delete 
    Patient’s Size   (0010,1020) Replace  
    E7 and E5/A70 and A50 1.5.x.x    DICOM Conformance Statement 000490000000111   Rev  A   Page 89   
    						
    							 Attribute Name  Tag 
    Delete / 
    Replace? 
    Patient’s W eight   (0010,1030) Replace  
    Pregnancy Status   (0010,10C0) Replace  
    Medical Alerts   (0010,2000) Delete 
    Alergies   (0010,2110) Delete 
    Ethnic Group   (0010,2160) Delete 
    Additional Patient History  (0010,21B0) Delete 
    Patient Comments   (0010,4000) Delete 
    Device Serial Number   (0018,1000) Delete 
    Protocol Name  (0018,1030) Delete 
    Study Instance UID   (0020,000D) Replace  
    Series Instance UID   (0020,000E) Replace  
    Study ID   (0020,0010) Delete 
    Requesting Physician   (0032,1032) Delete 
    Scheduled Procedure Step Description  (0040,0007)  Delete 
    Performed Procedure Step Start Date  (0040,0244)  Delete 
    Performed Procedure Step Start Time  (0040,0245)  Delete 
    Performed Procedure Step ID  (0040,0253) Delete 
    Performed Procedure Step Description  (0040,0254)  Delete 
    Comments on Performed Procedure Step  (0040,0280)  Delete 
    Names of Intended Recipients of Results  (0040,1010)  Delete 
     
    o  We will replace the Patient Name with an unidentifiable string composed of a prefix ‘DeId- ‘ and followed by 
    a  unique  numeric sequence,   
    o   We  will replace the Patient ID with a new generated unique ID.  
    o   All noted Instance UIDs are modified because the Voyager  generated instance UID’s are generated at the 
    time of the start of the procedure and suffixed with a date/time string noting the current date and time. 
    Because all manner of dates are considered identifying, the date and time portion of these strings is 
    replaced with a new date and time at the time of export. In this way we can be sure that they cannot  be 
    linked back to the original date and time  of the study. Referenced SOP instance UID links are  maintained. 
    o   The  Study ID  and  Performed Procedure Step ID on  Voyager include date strings, and thus must be 
    removed.  
    9.2   DATA DICTIONARY OF P RIVATE ATTRIBUTES 
    Philips private attributes in groups 0x2001 and 0x200 D may be included in objects created by  EPIQ/Affiniti. 
    9.3   CODED TERMINOLOGY AN D TEMPLATES 
    The Workflow AE is capable of supporting arbitrary coding schemes for Procedure and Protocol Codes.  The contents 
    of Requested Procedure Code Sequence (0032,1064) and Scheduled Protocol Code Sequence (0040,0008) supplied 
    in Worklist Items will be mapped to  Image IOD and MPPS attributes as described in Table  9.28.    
    E7 and E5/A70 and A50 1.5.x.x    DICOM Conformance Statement 000490000000111   Rev  A   Page 90   
    						