    Event reports, and images received from remote Storage SCP’s .  In addition, TLS security for incoming DICOM 
    communication may be e nabled, optional encryption enabled/disabled, and certificates imported, selected, or removed 
    from this interface.  
    4.4.1.2  Remote AE Title/Presentation Address Mapping  
    Remote AE’s may be configured on the PSC –  Network/DICOM  – DICOM Settings screen.   The screen is  organized 
    into DICOM Devices, each specified by  
    •   The Device Type which indicates the services for which the device is to be used  
    •   A Device Name given by the user in order to reference the device elsewhere in the EPIQ/Affiniti  user interface 
    •   AE Title for the  remote AE 
    •   Port for the remote AE  
    •   Network address, specified either by IPv4 IP address or a Hostname that is resolved by DNS  
    •   TLS security for outgoing DICOM communication may be enabled, optional encryption enabled/disabled, and 
    certificates imported, selected, or removed  
    •   Additional settings specific to each device type  
    The following Device Types may be configured:  
    Table  4.64  
    CONFIGURABLE DICOM DEVICES  
    Device Type   Supported SOPs 
    DICOM Storage  Server  Storage SOP Classes defined in Section 4.1.2.1
     
    Study Root Query Retrieve -FIND  
    Study Root Query Retrieve -MOVE  
    DICOM Commit Server   Storage Commitment Push Model  
    DICOM MPPS Server   Modality Performed Procedure Step  
    DICOM Worklist Server  Modality Worklist FIND  
    DICOM Structured Report Server  Comprehensive Structured Report  Storage 
    DICOM BW Printer   Basic Grayscale Print Meta 
    DICOM Color Printer  Basic Color Print Meta 
     
    To configure a single server that supports  multiple services, such as  image store, storage  commitment, and MPPS,  a 
    separate “Device” entry must be configured for each service using the correct device type,  an appropriate AE Title, 
    Port, and network address .    
    To configure Storage Commitment, define a Storage Server to which objects to be committed will be stored and a 
    Storage Commitment server (which may or may not be the same physical computer).  Then, on the DICOM Selection 
    screen, associate this Storage Server with the associated Storage Commitment server.  
    Configuration specific to pa rticular Device Types is described below.  
    4.4.1.2.1  Image Storage  Configuration  
    Multiple remote Storage SCPs can be defined and up to five  may be selected simultaneously for export.  For automatic 
    export of acquired images during an exam if set to send  “After Each Print/Acquire” or at the end of an exam i f set to 
    send  “At End of Exam”, a destination device must  be enabled in Setups  – Acquisition/Capture  – Archive/Print screen.  
    This screen allows each capture control to be separately  configured for export destination.  If a destination  device is 
    not enabled for a capture control, any images acquired through that capture control will not  be automatically sent to 
    that  destination device.   
    E7 and E5/A70 and A50 1.5.x.x    DICOM Conformance Statement 000490000000111   Rev  A   Page 61   
    						
    							 
    Each Storage Server includes a set of Advanced Settings allowing for device-specific control of the data to be stored to 
    that device.  These  settings include:  
    •   Compression to be used for single- frame images 
    •   Compression to be used for multi -frame images  
    •   Whether to send images in Monochrome or Color photom etric interpretations 
    •   Whether to limit export to Implicit VR Little Endian transfer syntax.  Generally used only for the few devices 
    that don’t understand any other transfer syntax.  
    •   Whether to send Structured Reports to this destination  
    •   Whether to send Nat ive Data of various imaging modes to this device.  Native Data  should not be selected 
    unless  the data will be displayed by  a workstation using Philips’ QLAB  analysis software  which is able to 
    understand this data, as the datasets are large  and could possibly affecting data transfer performance.  
    •   Whether to send Volume data and Philips Proprietary 3D Presentation State to this device.  If not enabled, the 
    device would still receive 2D representations of acquired volumes.  
    •   Association Timeout settings  (see 4.4.1.2.3)  
    •   Whether to send Pixel Spacing (0028,0030) with single- mode images to this device.  This is in addition to any 
    Region Calibration that may otherwise be sent to this device.  Use for allowing measurements to be made on 
    DICOM viewers that don’t understand Ultrasound Region Calibration.  
    4.4.1.2.2  Structured Report Storage  Configuration  
    A  Structured Report server  is a Storage SCP that only  supports and receives reports.   To enable this method, create a 
    Structured Report server in PSC –  Network/DICOM  – DICOM  Settings and then select it as the active Structured 
    Report server in PSC  – Network/DICOM  – DICOM Selection.  
    For Structured Reports to be sent automatically at end exam, the storage destination must be selected for either the 
    Acquire 1 or Acquire 2 keys in  Setups – Acquisition/Capture  – Archive/Print.  
    If SR support is confirmed using verify and measurements are made during the exam, one or more  Structured Reports 
    will be sent to the device . In Support > Network/DICOM > DICOM Settings > Structured Report Global Settings, you 
    may select which Structured Reports to send.  Additionally, Link Measurements to Images allows for inclusion of 
    coordinate data with the SR.   OB -GYN study types generate OB -GYN Ultrasound Procedure Reports, Vascular or 
    Abdominal study measurements generate a Vascular  Ultrasound Report, and the Adult Echo Study generates an 
    Echocardiography Procedure Report.  Two types of Pediatric Echo Reports are available  depending on configuration:  
    DICOM -standard Pediatric, Fetal, and Congenital Cardiac Ultrasound Report or Philips proprietary Pediatric Echo 
    Report. Note that there may be more than one report instance per exam, so long as they are from different study 
    types.  If no separate  Structured Report server is configured and the SOP Class fails  negotiation on the configured 
    Storage server , then no SR objects will be created, however, a failed job will remain in the queue.  This will need to be 
    manually removed.  
    User Defined measurements will now be sent in a Structured report.  No Structured Report i s sent for General or 
    Breast measurements.  
    4.4.1.2.3  Advanced  DICOM Device Association  Configuration  
    Configuration Settings for all DICOM devices.  
    Max PDU (Protocol Data Unit)  Size 
    o   The m aximum  PDU size in bytes  
    ARTIM (Association Request/Reject/Release) Ti meout 
    o    as defined by DICOM PS 3.8  
    Network Reply Timeout  
    E7 and E5/A70 and A50 1.5.x.x    DICOM Conformance Statement 000490000000111   Rev  A   Page 62   
    						
    							 o  If there is no response from the SCP within this time period, the Network/Reply timer expires.  The system will 
    send a DICOM A -RELEASE  Request . 
    Idle  Association Timeout  
    o   Association inactivity timeout.   
    Additional Configuration Settings for Storage Commit Devices . 
    There is an option provided for user if automatic retry of Commit job is required or not.  
    Number of retry attempts before abandoning storage commitment  
    o   Maximum  number of retries the  s ystem  would perform before setting the job to failed state.  Factory default 
    value is provided.  
    Retry delay after failed storage commit (minutes)  
    o   Time interval before a retry attempt when  storage commit has failed. Factory default value is provided.   
    Retry delay  after no response from storage commit server (hours)  
    o   Time interval before a retry attempt when no response has been received from the  storage commit server. 
    Factory default value is provided.  
     
    4.4.1.2.4  Serial Structured Report Storage Configuration  
    Serial  Report data  is a non-DICOM representation of Structured Reports which may additionally be exported via a USB 
    –  RS -232 adapter cable to a null -modem RS -232 cable to a serial port configured for 115200 baud, 8 bits, No Parity, 1 
    Stop Bit and Xon- Xoff Flow Control.  O utput will be an XML representation of the DICOM Structured Report  object . 
    4.4.1.2.5  Workflow  Configuration  
    Multiple  Worklist servers  may be defined, but only a single  W orklist server can be selected at a time  on the PSC – 
    Network/DICOM  – DICOM Selection page .  See  4.2.2.3.1 for the specification  of query criteria for worklist queries.  
    Multiple MPPS  servers may be defined, but only a single MPPS server  can be selected at a time  on the PSC – 
    Network/DICOM  – DICOM Selection page .  No other configuration is required for  MPPS servers.  
    4.4.1.2.6  Hardcopy 
    Multiple remote Print SCPs can be defined, but  no more than one Grayscale and one Color Print SCP may be selected 
    at a time.  
    Automatic sending of color images to the color printer and BW  images to the BW printer is selectable in  Setups  – 
    Acquisition/Capture  – Acquisition/Capture  – Send Images/Clips –  Automatic Printer Choice.  
    4.4.2  Parameters  
    The following parameters apply to all Application Entities.  The system must be power -cycled  after changing any of 
    these values : 
    Table A.4.4-2 
    Config uration Parameters table  
    E7 and E5/A70 and A50 1.5.x.x    DICOM Conformance Statement 000490000000111   Rev  A   Page 63   
    						
    							 Parameter  Configurable 
    (Yes/No) 
    Default Value  
    General Parameters  
    AE Title for all AE’s  Yes  AETitle 
    Listening Port   Yes  104 
    Station Name  Yes    
    Station Location   Yes    
    Hostname  Yes    
    MAX PDU Size  Yes  32768 
    Max Simultaneous Associations   Yes  5 
    ARTIM Timeout   Yes  60 
     
     
    5  MEDIA STORAGE  
    5.1  IMPLEMENTATION MODEL  
    5.1.1  Application Data Flow  
    Media
    Application
    Entity
    Export to
    Media
    Media
    Storage
    Medium
    Import from
    Media
    FSCFSU
    FSR*
     
    Figure 9  
    APPLICATION DATA FLO W DIAGRAM FOR MEDIA  STORAGE 
    The Media Application Entity exports Images, 3D Presentation States and structured Reports to a disk Storage 
    medium.  It is associated with the local real -world activity “Export to Media”.  “Export to Media” is performed upon user 
    request for selected studies,  performed procedures, or instances (images, 3D Volumes, 3D Presentation States,  and 
    Structured Reports) . 
    Throughout this section, the term “Media” refers to any of the media listed below which is in use.  
    EPIQ/Affiniti  will support the use of most writ able media including CD-R, CD -RW , DVD -R, DVD+R, DVD- RW, 
    DVD+RW  and removable USB Storage Device.  DICOM structure will be the same regardless of media used.   
    Note that although  –R or +R media may be erased multiple times using “erase”, the space may not b e 
    recovered.  If a – R or +R media is “Erased”, the previously written data is no longer available, and only the 
    E7 and E5/A70 and A50 1.5.x.x    DICOM Conformance Statement 000490000000111   Rev  A   Page 64   
    						
    							 remaining unwritten space on the media is available for use.  This restriction does not apply to + / -  RW media.  
    Erasing + / -  RW media allows  the entire disk’s space to be used.  
    If a – R or +R is erased, it is not usable again for DICOM exports due to reliability concerns associated with DVD 
    media. Only DVD+RW/ -RW can be erased and used again for DICOM export.  
    5.1.2  Functional Definition of AEs  
    5.1.2.1  Functio nal Definition of Media Application Entity  
    Using “ Export to… Media” will pass the currently selected studies,  exams, or individually selected images to the Media 
    Application Entity.  The SOP Instances associated with the selection will be collected into one or more export jobs.  
    The contents of each export job will be written to the installed media.  If the capacity of a disk is exceeded, the user is 
    provided a dialog, stating capacity exceeded and to insert another disk.  
    5.1.3  Sequencing of Real -World Activities  
    At least one image must exist and be selected before the Media Application Entity can be invoked.  The operator can 
    insert new media at any time.  If no writable media is available, the Media queue management Icon will be Yellow.   
    5.1.4  File Meta Information Op tions 
    The implementation information written to the File Meta Header in each file is:  
    Table  5.1  
    DICOM IMPLEMENTATION CLASS AND VERSION FOR MEDIA STORAGE  
    Implementation Class UID  1.3.46.670589.14.10000.105 
    Implementation Version Name  EPIQ-AFFIN_1.05  
     
    5.2  AE SPECIFICATIONS  
    5.2.1  Media Application Entity Specification  
    The Media Application Entity  provides standard conformance to the DICOM Interchange Option of the Media Storage 
    Service Class.  The Application Profiles and roles are listed in  DICOM PS 3.10 and PS 3.11. 
    Table  5.2  
    MEDIA  APPLICATION PRO FILES, ACTIVITIES, AND ROLES  
    Application Profiles Supported Real World Activity Roles SC Option 
    STD-US -SC -SF&MF -CDR  
    Export to….Media  FSC
     
    Interchange  
    STD-US -SC -SF&MF -DVD  
    STD -GEN -USB- JPEG   FSC
    , FSU  
    STD-US -SC -SF&MF -CDR  
    STD -US -SC -SF&MF -DVD  
    STD -GEN -USB- JPEG   Send to Hard Disk
     
    (import)   FSR
     
     
    5.2.1.1  File Meta Information for the Application Entity  
    The File -Set  ID (0004,1130)  included in the File Meta Header  of the DICOMDIR is  the same as the volume label of the 
    media.  
    E7 and E5/A70 and A50 1.5.x.x    DICOM Conformance Statement 000490000000111   Rev  A   Page 65   
    						
    							 
    5.2.1.2 Real -World Activities  
    5.2.1.2.1  Activity  – Export to  Media 
    The Media Application Entity acts as an FSC using the interchange option when requested to export SOP Instances 
    from the local database to media.  
    The contents of the export job will be written together with a corresponding DICOMDIR to media.  The user can cancel 
    an export job in the job queue. Writing in multi -se ssion format to CDs and DVDs is not  supported; a ny CD or DVD with 
    DICOM data on it will  be viewed as read- only and additional information will not be written to  the media. USB media 
    can be written m any times. 
    5.2.1.2.2  Activity  – Import from Media  
    The Media Application Entity acts as an FSR using the interchange option when requested to import SOP Instances 
    from media to the local database.   
    The Patient Directory user interface  displays the contents of removable media.  Exams selected by the user  are 
    transferred from the media to the system for review.  Objects transferred to the system retain their original SOP 
    Instance UIDs.   Patient Directory must be set to display the contents of the Hard Disk before the im ported exam 
    information may be displayed.  
    Note: Structured Repor ts may not be read back into EPIQ/Affiniti.  
    5.2.1.2.3  Activity  – Update to Media  
    The Media Application Entity acts as an FSU using the interchange option when requested to export SOP Instances 
    from the local database to media upon which DICOM data already resides.   
    The system user selects exams from the system’s directory for transfer to media that already contains data.  The 
    DICOMDIR is updated allowing access to original and new data.  
    DVD +RW media may be erased at any time, removing all previously recorded data.  
    5.2.1.2.3.1  Media Storage Application Profiles  
    See  Table  5.2  for supported Application Profiles.   
    5.2.1.2.3.2  Options  
    The Media Application Entity supports the SOP Classes and Transfer Syntaxes listed in  Table 5.3 . 
    Table  5.3  
    IODS, SOP CLASSES  AND TRANSFER SYNTAXES FOR REMOVABLE ¬MEDIA  
    Information Object Definition  SOP Class UID  Transfer Syntax  Transfer Syntax 
    UID 
    Media Storage Directory Storage  1.2.840.10008.1.3.10  Explicit VR Little Endian  1.2.840.10008.1.2.1  
    US Image Storage  1.2.840.10008.5.1.4.1.1.6.1 Explicit VR Little Endian  
    JPEG Lossy Baseline  
    JPEG Lossless Baseline  1.2.840.10008.1.2.1
     
    1.2.840.10008.1.2.4.50  
    1.2.840.10008.1.2.4.70  
    US Image Storage (Retired)  1.2.840.10008.5.1.4.1.1.6  Explicit VR Little Endian  
    JPEG Lossy Baseline  
    JPEG Lossless Baseline  1.2.840.10008.1.2.1
     
    1.2.840.10008.1.2.4.50  
    1.2.840.10008.1.2.4.70  
    US Multiframe Image Storage  1.2.840.10008.5.1.4.1.1.3.1  Explicit VR Little Endian  1.2.840.10008.1.2.1  
    E7 and E5/A70 and A50 1.5.x.x    DICOM Conformance Statement 000490000000111   Rev  A   Page 66   
    						
    							 
    JPEG Baseline Lossy 
    Compression  1.2.840.10008.1.2.4.50
     
    US Multiframe Image Storage (Retired)*  1.2.840.10008.5.1.4.1.1.3  Explicit VR Little Endian  1.2.840.10008.1.2.1  
    JPEG Baseline Lossy 
    Compression  1.2.840.10008.1.2.4.50
     
    Secondary Capture Image Storage  1.2.840.10008.5.1.4.1.1.7  Explicit  VR Little Endian  
    JPEG Lossy Baseline  
    JPEG Lossless Baseline  1.2.840.10008.1.2.1
     
    1.2.840.10008.1.2.4.50  
    1.2.840.10008.1.2.4.70  
    Multi-frame True Color Secondary Capture 
    Image Storage   1.2.840.10008.5.1.4.1.1.7.4
     Explicit VR Little Endian  
    JPEG Lossy Baseline  
    JPEG Lossless Baseline  1.2.840.10008.1.2.1
     
    1.2.840.10008.1.2.4.50  
    1.2.840.10008.1.2.4.70  
    CT Image Storage  1.2.840.10008.5.1.4.1.1.2 Explicit VR Little Endian  
    JPEG Lossy Baseline  
    JPEG Extended 12-bit Lossy 
    JPEG Lossless Baseline  1.2.840.10008.1.2.1
     
    1.2.840.10008.1.2.4.50  
    1.2.840.10008.1.2.4.51  
    1.2.840.10008.1.2.4.70  
    Enhanced CT Image Storage  1.2.840.10008.5.1.4.1.1.2.1  Explicit VR Little Endian  
    JPEG Lossy Baseline  
    JPEG Extended 12-bit Lossy 
    JPEG Lossless Baseline  1.2.840.10008.1.2.1
     
    1.2.840.10008.1.2.4.50  
    1.2.840.10008.1.2.4.51  
    1.2.840.10008.1.2.4.70  
    Digital Mammography X -Ray Image Storage 
    –  For Presentation   1.2.840.10008.5.1.4.1.1.1.2
     Explicit VR Little Endian  
    JPEG Lossy Baseline  
    JPEG Extended 12-bit Lossy 
    JPEG Lossless Baseline  1.2.840.10008.1.2.1
     
    1.2.840.10008.1.2.4.50  
    1.2.840.10008.1.2.4.51  
    1.2.840.10008.1.2.4.70  
    Digital Mammography X -Ray Image Storage 
    –  For Processing  1.2.840.10008.5.1.4.1.1.1.2.1
     Explicit VR Little Endian  
    JPEG Lossy Baseline  
    JPEG Extended 12-bit Lossy  
    JPEG Lossless Baseline  1.2.840.10008.1.2.1
     
    1.2.840.10008.1.2.4.50  
    1.2.840.10008.1.2.4.51  
    1.2.840.10008.1.2.4.70  
    MR Image Storage  1.2.840.10008.5.1.4.1.1.4 Explicit VR Little Endian  
    JPEG Lossy Baseline  
    JPEG Extended 12-bit Lossy 
    JPEG Lossless Baseline  1.2.840.10008.1.2.1
     
    1.2.840.10008.1.2.4.50  
    1.2.840.10008.1.2.4.51  
    1.2.840.10008.1.2.4.70  
    Enhanced MR Image Storage  1.2.840.10008.5.1.4.1.1.4.1  Explicit VR Little Endian  
    JPEG Lossy Baseline  
    JPEG Extended 12-bit Lossy 
    JPEG Lossless Baseline  1.2.840.10008.1.2.1
     
    1.2.840.10008.1.2.4.50  
    1.2.840.10008.1.2.4.51  
    1.2.840.10008.1.2.4.70  
    MR Spectroscopy Image Storage  1.2.840.10008.5.1.4.1.1.4.2  Explicit VR Little Endian  
    JPEG Lossy Baseline  
    JPEG Extended 12-bit Lossy 
    JPEG Lossless Baseline  1.2.840.10008.1.2.1
     
    1.2.840.10008.1.2.4.50  
    1.2.840.10008.1.2.4.51  
    1.2.840.10008.1.2.4.70  
    Multi-frame Single Bit Secondary Capture 
    Image Storage   1.2.840.10008.5.1.4.1.1.7.1
     Explicit VR Little Endian  
      1.2.840.10008.1.2.1
     
     
    Multi -frame  Grayscale Byte Secondary 
    Capture Image Storage   1.2.840.10008.5.1.4.1.1.7.2
     Explicit VR Little Endian  
    JPEG Lossy Baseline  
    JPEG Lossless Baseline  1.2.840.10008.1.2.1
     
    1.2.840.10008.1.2.4.50  
    1.2.840.10008.1.2.4.70  
    Multi-frame  Grayscale Word Secondary 
    Capture Image Storage   1.2.840.10008.5.1.4.1.1.7.3
     Explicit VR Little Endian  
    JPEG Lossy Baseline  
    JPEG Lossless Baseline  1.2.840.10008.1.2.1
     
    1.2.840.10008.1.2.4.50  
    1.2.840.10008.1.2.4.70  
    E7 and E5/A70 and A50 1.5.x.x    DICOM Conformance Statement 000490000000111   Rev  A   Page 67   
    						
    							 
    Positron Emission Tomography Image 
    Storage   1.2.840.10008.5.1.4.1.1.128
     Explicit VR Little Endian  
    JPEG Lossy Baseline  
    JPEG Lossless Baseline  1.2.840.10008.1.2.1
     
    1.2.840.10008.1.2.4.50  
    1.2.840.10008.1.2.4.70  
    X-Ray Angiographic Image Storage  1.2.840.10008.5.1.4.1.1.12.1  Explicit VR Little Endian  
    JPEG Lossy Baseline  
    JPEG Lossless Baseline  1.2.840.10008.1.2.1
     
    1.2.840.10008.1.2.4.50  
    1.2.840.10008.1.2.4.70  
    Comprehensive Structured Report Storage  1.2.840.10008.5.1.4.1.1.88.33  Explicit VR Little Endian  1.2.840.10008.1.2.1  
    Private 3D Presentation State**  1.3.46.670589.2.5.1.1  Explicit VR Little Endian  1.2.840.10008.1.2.1  
    ** For import to Philips QLAB or Xcelera workstations only.  
     
    6   TRANSFORMATION OF DICOM TO CDA  
    EPIQ/Affiniti exports a CDA file format which complies with  the HL7 CDA -R2 schema.  
    7  SUPPORT OF CHARACTER SETS 
    All EPIQ/Affiniti  DICOM applications support the  following values of Specific Character Set (0008,0005):  
    ISO_IR 100 (ISO 8859- 1:1987 Latin Alphabet No. 1 supplementary set)  
    GB 18030  (Chinese)  
    8  SECURITY  
    8.1  GENERAL SECURITY 
    EPIQ/ Affiniti  incorporates  an internal firewall that only accepts incoming traffic on the designated listening port, as 
    configured in PSC  – Network/DICOM  – DICOM Settings.   The port is only opened if one or more servers are defined.   
    8.2   SUPPORTED  DICOM  SECURITY PROFILES  
    8.2.1  TLS Secure Transport Connection Profile s 
    Basic  TLS Secure Transport Connection Profile  and the AES TLS Secure Transport Connection Profile are  supported 
    using Transport Layer Security Version 1.2 protocol  with the following features:  
    Supported TLS Feature Mechanism 
    Entity Authentication RSA based certificates 
    Exchange of Master Secrets RSA 
    Data Integrity SHA 
    Privacy (Cyphersuite Options) TLS_RSA_WITH_AES_128_CBC_SHA  (preferred)  
    TLS_RSA_WITH_3DES_EDE_CBC_SHA
     
    E7 and E5/A70 and A50 1.5.x.x    DICOM Conformance Statement 000490000000111   Rev  A   Page 68   
    						
    							 
     
    TLS authentication may be used with or without TLS encryption – default is with  TLS encryption.  
    For outgoing TLS requests where EPIQ/Affiniti acts as a TLS Client, the received server RSA certificate is validated by 
    verifying its digital signature against  a certificate in the local Trusted Certificate Authorities store whose Subject 
    matches the Issuer of the received certificate.  If requested by the TLS Server, EPIQ/Affiniti will then send the 
    certificate configured for “This System” from the local Persona l store. 
    For incoming TLS connection requests where EPIQ/Affiniti acts as a TLS Server, the certificate configured for “This 
    System” is sent from the Personal store to the client as specified in the TLS protocol. EPIQ/Affiniti always requests a 
    client cert ificate from the TLS Client.  
    Certificates are locally managed. The site administrator may import certificates from media to the Trusted Certificate 
    Authority store for verifying incoming certificates; these may be either CA certificates or self -signed end- entity 
    certificates.  The site administrator may also import certificates from media to the Personal store along with the 
    certificates private key for certificates representing ‘This System’.  A wide variety of certificate file formats are 
    supported, inclu ding 
    •   Base 64 Text:  PEM format (.pem, .crt), usually unencrypted but may be encrypted. If encrypted, the 
    encryption password must be provided when importing.  
    •   ASN.1 Binary:  BER, DER, CER formats (.der, .cer, .crt)  
    •   PKCS#12: Encrypted Container for certificates with private keys (.pfx, .p12) . This format is used primarily for 
    the “This System” certificate with corresponding private key. The encryption password must be provided when 
    importing.  
    •   PKCS#7:  Container for multiple certificates. If there is only one certificate in the container, then the certificate  
    is saved to the user -specified store name. If there are multiple certificates in the container, then the Issuer and 
    Subject names of each certificate are compared; if the names are the same, the certificat e is stored in the 
    Trusted Certificate Authority store, else in the Intermediate Certificate Authority store.  
    The site administrator may also remove previously -imported certificates.  
    The incoming TCP port defaults to port 2762 when TLS security is enabled;  this port may be changed by the user if 
    desired.   
    E7 and E5/A70 and A50 1.5.x.x    DICOM Conformance Statement 000490000000111   Rev  A   Page 69   
    						
    							 
    9  ANNEXES 
    9.1  IOD CONTENTS  
    9.1.1  Created SOP Instances  
    Table  9.1  specifies the attributes of an Ultrasound Image transmitted b y the EPIQ/Affiniti storage application.  
    Table  9.2  specifies the attributes of a Comprehensive Structured Reports transmitted by the  EPIQ/Affiniti storage 
    application .  Please note that there are differences between which Structured Report Templates are used in each 
    product.  
    The following tables use a number of abbreviations.  The abbreviations used in the “Presence of …” column are:  
    VNAP    Value Not Always Present  
    ANAP    Attribute Not Always Present  
    ALWAYS    Always Present  
    ALWAYSUA   Always Present, unless anonymized  
    EMPTY   Attribute is sent without a value  
     
    The abbreviations used in the “Source” column:  
    MW L   the attribute value source Modality Worklist  
    USER  the attribute v alue source is from User input  
    AUTO  the attribute value is generated automatically   
    MPPS  the attribute value is the same as the Modality Performed Procedure Step service  
    CONFIG  the attribute value source is a configurable parameter  
     
    9.1.1.1  US or US Multiframe Ima ge IOD 
    Table 9.1  
    IOD OF CREATED US OR US MULTIFRAME SOP INSTANCES  
    IE Module Reference Presence of Module 
    Patient Patient   Table 9.5   ALWAYS 
    Study  General Study
      Table 9.6   ALWAYS Patient Study   Table 9.7   ALWAYS 
    Series General Series   Table 9.9   ALWAYS 
    Equipment General Equipment   Table 9.10  ALWAYS 
    Image  General Image 
    Table 9.11  ALWAYS 
    Image Pixel   Table 9.12  ALWAYS 
    Cine  Table 9.13  Only if Multi- frame 
    Multi-frame   Table 9.14  Only if Multi- frame 
    US Region Calibration   Table 9.15  ANAP 
    US Image   Table 9.16  ALWAYS 
    VOI LUT  Table 9.17  Only if Single frame  
    E7 and E5/A70 and A50 1.5.x.x    DICOM Conformance Statement 000490000000111   Rev  A   Page 70   
    						