Page 111
3.3 Defining Your Samples 3 - 16 When “Use sample table” is selected in Analysis Method, samples must be defined in the following procedure before measurement. Open the monitor window and click the button. When the Sample Table window in Fig. 3-11 appears, enter the sample name and comment for each sample. 1 For creating a sample table newly, make entry for Number of Samples and click the Update button. For instance, if there are five samples, enter “5” for Number of Samples and...
Page 112
3.3.2 3 - 17 Fig. 3-12 4 Make entries for Sample and Comments of each sample. The sample names entered here appear on the monitor window. Maximum number of characters to be entered (half-size) Sample : 40 characters Comment : 255 characters Next, you must enter a file name for automatic saving after measurement. When you click a File Name column, the [...] button appears as shown in Fig. 3-13. Click this button. Upon opening of the window shown in Fig. 3-14,...
Page 113
3.3 Defining Your Samples 3 - 18 Fig. 3-14 5 After entries, check them and then click the OK button. 6 Load button Used for reading out the existing sample table. Clicking the Load button displays the file opening dialog. Fig. 3-15 File Opening Dialog Select a sample table and click the Open button. The specified sample table will be read out.
Page 114
3.3.2 3 - 19 7 Save button The contents of change can be saved by overwriting. In case of a new generation, the Save As dialog will open. 8 Save As button Used for saving the sample table with a name newly given. Click this button, and the Save As dialog will open. Fig. 3-16 Save As Dialog Select a saving location and enter a file name. Then, click the Save button.
Page 115
3.4 Conducting Measurement 3 - 20 3.4 Conducting Measurement (1) When “Use sample table” is not selected in Analysis Method, you should click the button and make sure that a sample has already been defined. (2) Set a sample and click (measurement) button. Measurement starts and a spectrum is displayed. For halting the measurement midway, click the (stop) button. (3) Upon completion of the measurement, the measured data will be saved automatically at the specified...
Page 116
3.4.2 3 - 21 (1) Click the button and make sure a sample has already been defined. (2) Set a sample and click the button to start measurement. (3) Upon start of measurement, the window shown in Fig. 3-18 appears. Click Yes. Fig. 3-18 (4) A dialog box opens as shown below. Fig. 3-19 Click OK. Measurement starts and a spectrum is displayed. For halting the measurement midway, click the (stop) button. Measure all the samples set in the sample table...
Page 117
3.4 Conducting Measurement 3 - 22 (5) Upon completion of the measurement, the measured data will be saved automatically at the specified saving location using the specified file name. Opening the saved data file will display the data processing window as shown in Fig. 3-20. This figure shows a spectrum and the result of rate calculation. Peak table can be displayed, but both the result of rate calculation and the peak table cannot be displayed simultaneously. Simultaneous display...
Page 118
3.4.3 3 - 23 (1) Put a check mark for ‘Stopped Flow’ at the Instrument tab in the Analysis Method window. At the General tab in this window, remove a check mark from “Use sample table.” (2) Click the button to make sure a sample has already been defined. (3) Click the (measurement) button to start measurement. (4) When measurement starts, the window shown in 3-21 appears to await an external start signal. Fig. 3-21 Stopped Flow Measurement Window (external...
Page 119
3.4 Conducting Measurement 3 - 24 (5) On arrival of the external signal, measurement starts and a spectrum is drawn in the window. Fig. 3-22 Stopped Flow Measurement Window For stopping the measurement halfway, click the (stop) button. (6) Upon completion of the measurement, the measured data will be saved automatically at the specified saving location using the specified file name when auto saving is selected. If auto saving is not selected, select the “Save As”...
Page 120
3.4.3 3 - 25 NOTICE: 1. The data measured under the following conditions will be saved automatically in Program Files\FL Solutions\Data. (Example: C:Program Files\FL Solutions\Data) (1) Measurement was carried out using a sample table with no file name specified in the File Name column. (2) Measurement was carried out with “Auto File” specified and a file name unspecified, avoiding use of a sample table. 2. The data measured under the following conditions will be saved...