Hitachi F7000 Instruction Manual
Here you can view all the pages of manual Hitachi F7000 Instruction Manual. The Hitachi manuals for Fluorescence Spectrophotometer are available online for free. You can easily download all the documents as PDF.
Page 61
1.8 Window Setting 1 - 40 1.8.7 Analysis Condition Tab For the settable range of each parameter, refer to Appendix I. Fig. 1-56 Analysis Condition Tab 1 Integration time for current data An integration time for reading the current data by the Read Data command of Tools menu can be set. 2 Change Instrumental Response Condition When this check box is turned on, the excitation slit, emission slit and photomultiplier voltage can be set on the instrument function...
Page 62
1.8.8 1 - 41 1.8.8 Report Tab Fig. 1-57 Report Tab Data List Printing Direction The direction of data list printing (with constant line spacing) can be set. In case of vertical printing: Wavelength or time and photometric values will be printed vertically in 3 rows. In case of horizontal printing: Wavelength or time and photometric values will be printed horizontally in 3 rows. That is, 3 data will be printed on each line.
Page 63
1.9 Saving a Data File 1 - 42 1.9 Saving a Data File Data can or cannot be saved depending on setting. So perform setting, referring to the table below. Setting in any of patterns ① to ⑤ is recommended. “Use sample table in analysis method” Setting ON OFF Sample name/comment input dialog “Save file automatically after data acquisition” ⎯ ON OFF Analysis Method - Monitor tab “Open data processing window after data acquisition” ON OFF ON OFF ON OFF Display × × ×...
Page 64
1.10 1 - 43 1.10 Photometer Menu Commands The Spectrophotometer menu contains the following commands. • Set Wavelength • Auto Pre-scan • Zero Adjust • Shutter Open/Close Set Wavelength This command sets the excitation/emission wavelength to a specified wavelength. Click the Set Wavelength command or button, and the dialog box shown below will come up. Fig. 1-58 Zero Adjust This command automatically closes the shutter and measures a zero point in detection....
Page 65
1.11 Utility Menu Commands 1 - 44 1.11 Utility Menu Commands The Utility menu contains the following command. • Instrument Response • Analog Output Gain • EX Power Monitor • Sensitivity Check • Lamp Usage • Lamp Off Instrument Response This command is used for spectral correction. The spectral correction accessory kit (P/N 251-0126) needs to be prepared. Analog Output Gain Use this command for setting a data value corresponding to 1 V of analog output....
Page 66
1.11 1 - 45 When the limit use time is exceeded, the following message appears. Fig. 1-60 After lamp replacement, reset this indication. For details, refer to the maintenance edition. Note: If the Model F-7000 power supply is turned on without starting up the PC, indication of the Xenon lamp turn-on time may not be updated correctly. Lamp Off Use this command for turning off the Xenon lamp. For turning on the Xenon lamp again, restarting the spectrophotometer is...
Page 67
1.12 Termination of FL Solutions Program 1 - 46 1.12 Termination of FL Solutions Program After completion of measurement, shut down the instrument in the following procedure. 1 From the File menu, select the Exit command. A window appears as shown below. Fig. 1-61 2 For termination without turning off the light source lamp, select “Close the monitor window, but keep the lamp operating?” For termination after turning off the light source lamp, select “Close (Turn...
Page 68
2.1 2 - 1 2. WAVELENGTH SCAN 2.1 Flow of Operation This section explains the condition setting and operating method for wavelength scan. Measurement proceeds along the following flow. Fig. 2-1 Operational Flow (wavelength scan) Start Method setting (See 2.2.) Sample name/ comment input (See 2.3.) Pre-scan (See 2.4) Measurement (See 2.5.) Data processing (See Section 7.) Result output in report...
Page 69
2.2 Creating an Analysis Method 2 - 2 2.2 Creating an Analysis Method Set the analytical conditions (method) prior to measurement. Select the Method command from the Edit menu or click the button. A window shown in Fig. 2-2 will appear. Fig. 2-2 Analysis Method The following tabs are provided on the Analysis Method window for the wavelength scan mode. These tabs are detailed on the subsequent pages. • General • Instrument • Monitor • Processing •...
Page 70
2.2.1 2 - 3 The parameters shown in Fig. 2-3 are provided at the General tab. For the settable range of each parameter, refer to Appendix G. Fig. 2-3 General Tab 1 Measurement (mode) A measurement mode is selectable. Here, select “Wavelength scan.” For Time scan, Photometry, 3-D Scan and 3-D Time scan, refer to Sections 3, 4, 5 and 6, respectively. 2 Operator Enter an operator name. 3 Instrument The name of the connected instrument is indicated. 4 Sampling -...