1.8  Window Setting 
1 - 40 
1.8.7  Analysis Condition Tab  
 
For the settable range of each parameter, refer to Appendix I.   
 
 
Fig. 1-56    Analysis Condition Tab   
 
1 Integration time for current data   
An integration time for reading the 
current data by the Read Data 
command of Tools menu can be set.   
2 Change Instrumental Response 
Condition  
When this check box is turned on, the 
excitation slit, emission slit and 
photomultiplier voltage can be set on 
the instrument function...

Page 62

1.8.8 
1 - 41 
1.8.8  Report Tab  
 
 
Fig. 1-57  Report Tab  
 Data List Printing Direction   
The direction of data list printing (with 
constant line spacing) can be set.   
In case of vertical printing:   
Wavelength or time and photometric 
values will be printed vertically in 3 
rows.  
 
 
 
 
 
In case of horizontal printing:   
Wavelength or time and photometric 
values will be printed horizontally in 
3 rows.    That is, 3 data will be 
printed on each line.   
 
 
 
 
 
  

Page 63

1.9    Saving a Data File 
1 - 42 
1.9    Saving a Data File   
 
Data can or cannot be saved depending on setting.    So perform setting, referring to the table below.   
Setting in any of patterns ① to ⑤ is recommended.  
 
“Use sample table in analysis method” Setting 
ON OFF 
Sample name/comment input 
dialog 
  “Save file automatically 
after data acquisition” ⎯ ON OFF 
Analysis Method - Monitor tab 
  “Open data processing 
window after data 
acquisition” ON OFF ON OFF ON OFF 
Display  ×  ×  ×...

Page 64

1.10 
1 - 43 
1.10  Photometer Menu Commands  
 
The Spectrophotometer menu contains the following commands.   
• Set Wavelength  
• Auto Pre-scan  
• Zero Adjust  
• Shutter Open/Close  
 
Set Wavelength   
This command sets the excitation/emission wavelength to a specified wavelength.   
Click the Set Wavelength command or button, and the dialog box shown below will come up.   
 
 
 
Fig. 1-58   
 
 
Zero Adjust   
This command automatically closes the shutter and measures a zero point in detection....

Page 65

1.11  Utility Menu Commands 
1 - 44 
1.11  Utility Menu Commands  
 
The Utility menu contains the following command.   
• Instrument Response  
•  Analog Output Gain   
•  EX Power Monitor   
• Sensitivity Check  
• Lamp Usage  
• Lamp Off  
 
Instrument Response   
 
This command is used for spectral correction.    The spectral correction accessory kit (P/N 251-0126) 
needs to be prepared.   
 
Analog Output Gain   
 
Use this command for setting a data value corresponding to 1 V of analog output....

Page 66

1.11 
1 - 45 
When the limit use time is exceeded, the following message appears.   
 
 
 
Fig. 1-60   
 
 
After lamp replacement, reset this indication.    For details, refer to the maintenance edition.   
 
Note:  If the Model F-7000 power supply is turned on without starting up the PC, indication of the 
Xenon lamp turn-on time may not be updated correctly.     
 
Lamp Off   
 
Use this command for turning off the Xenon lamp.   
For turning on the Xenon lamp again, restarting the spectrophotometer is...

Page 67

1.12    Termination of FL Solutions Program 
1 - 46 
1.12    Termination of FL Solutions Program   
 
After completion of measurement, shut down the instrument in the following procedure.   
 
1 From the File menu, select the Exit command.   
A window appears as shown below.   
 
 
 
Fig. 1-61   
 
 
2 For termination without turning off the light source lamp, select “Close the monitor window, but 
keep the lamp operating?”   
For termination after turning off the light source lamp, select “Close (Turn...

Page 68

2.1 
2 - 1 
2.  WAVELENGTH SCAN  
 
2.1  Flow of Operation  
 
This section explains the condition setting and operating method for 
wavelength scan.   
Measurement proceeds along the following flow.   
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Fig. 2-1    Operational Flow (wavelength scan)   
Start 
Method setting 
(See 2.2.) 
Sample name/ 
comment input 
(See 2.3.) 
Pre-scan 
(See 2.4) 
Measurement 
(See 2.5.) 
Data processing 
(See Section 7.) 
Result output in report...

Page 69

2.2    Creating an Analysis Method 
2 - 2 
2.2    Creating an Analysis Method   
 
Set the analytical conditions (method) prior to measurement.     
Select the Method command from the Edit menu or click the 
 
button.    A window shown in Fig. 2-2 will appear.   
 
 
 
Fig. 2-2  Analysis Method  
 
 
The following tabs are provided on the Analysis Method window for the 
wavelength scan mode.    These tabs are detailed on the subsequent 
pages.  
• General  
• Instrument  
• Monitor  
• Processing  
•...

Page 70

2.2.1 
2 - 3 
The parameters shown in Fig. 2-3 are provided at the General tab.     
For the settable range of each parameter, refer to Appendix G.   
 
 
 
Fig. 2-3  General Tab  
 
 
1 Measurement (mode)   
A measurement mode is selectable.    Here, select “Wavelength 
scan.”  
For Time scan, Photometry, 3-D Scan and 3-D Time scan, refer to 
Sections 3, 4, 5 and 6, respectively.   
2 Operator  
Enter an operator name.   
3 Instrument  
The name of the connected instrument is indicated.   
4 Sampling 
-...