Page 731

Configuring Connectivity
LOGIQ 5 Basic User Manual 16-71
Direction 2300002-100 Rev. 2
Removing a destination device 
1.  In the Destination Device section, select the destination 
device. 
2.  In the Destination Device section, select the Remove 
button.
The system displays a warning message to confirm that you 
want to remove the device and associated services.
NOTE:  When you remove a destination device, all associated 
services are automatically removed.
3. Select OK
 Figure 16-48. Services
Table...

Page 732

Customizing Your System
16-72 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Adding a service to a destination device
1.  In the Destination Device section, in the Name field, select 
the destination device. 
2.  In the Services section, select the service from the pull-
down menu to the right.
3.  Select the Add button below the list.
The system adds the service to the list of services.
Removing a service
1.  In the Destination Device section, in the Name field, select 
the destination device....

Page 733

Configuring Connectivity
LOGIQ 5 Basic User Manual 16-73
Direction 2300002-100 Rev. 2
Changing parameters for a service
 Figure 16-49. Service Parameters
1.  In the Destination Device section, in the Name field, select 
the destination device. 
2.  In the Services section, move the Trackball to highlight a 
service, and press Set.
The system displays parameters for the service in the 
bottom section of the screen.
3.  Define the appropriate service configuration parameters.
Table 16-44: Service...

Page 734

Customizing Your System
16-74 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Changing parameters for a service (continued)
Many service parameters are specific to each type of service. 
The parameters are described on the following pages:
• DICOM Verification
• DICOM Image Storage
•  DICOM Performed Procedure
• DICOM Print
•  DICOM Media Storage
• DICOM Worklist
•  Echo Database V1.1
•  Echo Database V2.0
• EchoPac Media
• Standard Print 

Page 735

Configuring Connectivity
LOGIQ 5 Basic User Manual 16-75
Direction 2300002-100 Rev. 2
DICOM Verification
To verify a service:
1.  In the Destination Device section, in the Name field, select 
the destination device. 
2.  In the Services section, select the service.
3.  In the bottom section of the screen with the service 
parameters, select DICOM Verify. 
If the device is communicating with the network, the yellow 
face is displayed with a smile. See Figure 16-50. 
If the device is not connected, the...

Page 736

Customizing Your System
16-76 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
DICOM Image Storage
DICOM Image Storage allows the system to send or receive 
ultrasound images in a format that can be interpreted by PACS.
 Figure 16-51. DICOM Image Storage Service Type
Table 16-45: DICOM Image Storage
Preset ParameterDescription
Reopen pr. image Reopen per image 
Allow raw data Select to save data in both TruAccess (raw data) 
and DICOM format.Clear to save in DICOM format 
only.
Only Black/white...

Page 737

Configuring Connectivity
LOGIQ 5 Basic User Manual 16-77
Direction 2300002-100 Rev. 2
DICOM Performed Procedure
DICOM Performed Procedure provides an acknowledgement 
that a study has been performed.
 Figure 16-52. DICOM Performed Procedure Service Type 

Page 738

Customizing Your System
16-78 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
DICOM Print
DICOM Print provides the ability to send or receive ultrasound 
image data to DICOM printers.
 Figure 16-53. DICOM Print Service Type: Settings 
Table 16-46: Settings
Preset ParameterDescription
Format Indicates how many prints to print per page, for 
example, 1,1, 1,2, 1,3, up to 5,5. Partial prints are 
displayed as one print job.
Magnification Type Specify how the printer magnifies the image to fit it...

Page 739

Configuring Connectivity
LOGIQ 5 Basic User Manual 16-79
Direction 2300002-100 Rev. 2
DICOM Print (continued)
 Figure 16-54. DICOM Print Service Type: Densities 
Table 16-47: Densities
Preset ParameterDescription
Min Enter a number indicating the minimum density 
level of the film.
Max Enter a number indicating the maximum density 
level of the film.
Border Select to have the border area surrounding and 
between the images of the film: Black or White.
Empty Image Select to have a Black or White empty...

Page 740

Customizing Your System
16-80 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
DICOM Print (continued)
 Figure 16-55. DICOM Print Service Type: Job 
Table 16-48: Job
Preset ParameterDescription
Film Session Label Type a name for the group of film labels associated 
with the print job.
Film Destination Specify the film destination for the exposed 
filmMagazine – Store in a film magazineProcessor 
– Develop in a film processor
Color Select whether to have the image Color or Grey.
Priority Specify the...