GE Logiq P5 User Manual
Here you can view all the pages of manual GE Logiq P5 User Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.
Page 691
System Presets LOGIQ 5 Basic User Manual 16-31 Direction 2300002-100 Rev. 2 Restore procedure: user-defined configurations 1. Insert the MOD (Magneto-Optical Disk) with the user- defined configurations into the MOD drive 2. On the Touch Panel, press Utility. 3. On the Utility Touch Panel, press System. 4. On the monitor display, select Backup/Restore. The Backup/Restore screen is displayed. See Figure 16-16. NOTE: If you are not logged in with administrator privileges, the Operator Login window is...
Page 692
Customizing Your System 16-32 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 System/Peripherals Preset Menu The System/Peripherals screen allows you to specify video and system setup parameters. Figure 16-17. System/Peripherals Preset Menu Table 16-17: Video Settings Preset ParameterDescription Format Select the video format, PAL or NTSC. Mixed Mode Allows you to select the PAL format, but still operate the color printer in NTSC mode. This setting has no effect on NTSC mode. Table 16-18:...
Page 693
System Presets LOGIQ 5 Basic User Manual 16-33 Direction 2300002-100 Rev. 2 System/Peripherals Preset Menu (continued) Table 16-19: Print and Store Options Preset ParameterDescription Set Preview Size to Video AreaSelect to set capture image area as the entire video area -- desired format for DICOM prints. Dicom Store with Title BarCheck box to select. Store Dual as Dicom OnlySelect to always store dual images as a DICOM (secondary capture) store, rather than Raw DICOM. Store Multiframe for Sec...
Page 694
Customizing Your System 16-34 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 System/About Preset Menu The System/About screen lists information about the system software. Figure 16-18. System About Preset Menu Table 16-22: Software Preset ParameterDescription Software Version The current software version on this system. Software Part NumberThe software part number. Build View The software build view. Build Date The software build date. Table 16-23: System Image Preset ParameterDescription...
Page 695
Imaging Presets LOGIQ 5 Basic User Manual 16-35 Direction 2300002-100 Rev. 2 Imaging Presets Overview Imaging screens allow you to specify parameters for the following: • B-Mode (B) • Color Flow Mode (CF) • Power Doppler Imaging (PDI) • M-Mode (M) • Pulse Wave Mode (PW) • Continuous Wave Mode (CW) [LOGIQ 7 Only] • Harmonics (HAR) • General
Page 696
Customizing Your System 16-36 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Changing imaging presets To change imaging presets: 1. On the Touch Panel, select Utility. 2. On the Touch Panel, select Imaging. The system displays the Imaging screens. See the Imaging screens on the following pages. 3. In the Preset list, select the exam. 4. In the Probe list, select the probe. 5. In the row across the top of the screen, select the mode. The system displays two sets of parameters and settings....
Page 697
Imaging Presets LOGIQ 5 Basic User Manual 16-37 Direction 2300002-100 Rev. 2 LOGIQ 5 Imaging Presets B-Mode Figure 16-19. B-Mode parameters Color Flow Mode Figure 16-20. Color Flow parameters
Page 698
Customizing Your System 16-38 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Power Doppler Imaging (PDI) Figure 16-21. PDI parameters M-Mode Figure 16-22. M-Mode parameters
Page 699
Imaging Presets LOGIQ 5 Basic User Manual 16-39 Direction 2300002-100 Rev. 2 Pulse Wave (PW) Figure 16-23. PW parameters Continuous Wave (CW) Figure 16-24. CW parameters
Page 700
Customizing Your System 16-40 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Harmonics Figure 16-25. Harmonics parameters General Figure 16-26. General parameters