Page 681
System Presets LOGIQ 5 Basic User Manual 16-21 Direction 2300002-100 Rev. 2 System/System Measure Preset Menu (continued) Table 16-10: Measurement Preset ParameterDescription Repeat Measurement Select Off, Repeat, DefaultMeas OB Type Select which OB measurements and calculations studies to use: USA, Europe, Tokyo, Osaka, or ASUM. EFW GP Select the source used to calculate EFW-GP Estimated Fetal Weight-Growth Percentile) CUA/AUA for HadlockSelect to use CUA (Composite Ultrasound Age) or AUA (Average...
Page 682
Customizing Your System 16-22 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Table 16-12: Results Window Preset ParameterDescription Result Window Location-2DSelect the Result Window location on the Monitor Display: Left-Bottom, Left-Top, Right-Bottom, or Right-Top. Result Window Location-TimelineSelect the Result Window location: Left-Bottom, Left-Top, Right-Bottom, or Right-Top. Result Window FormatSelect Wide or Narrow. Font Color Select White, Off White, Yellow, Bright Red or Orange...
Page 683
System Presets LOGIQ 5 Basic User Manual 16-23 Direction 2300002-100 Rev. 2 System/Backup and Restore Preset Menu The backup and restore procedures described in this section are divided into two parts. The first part describes procedures to backup and restore patient data. The second part describes procedures to backup and restore system and user-defined configurations. Figure 16-16. System/Backup/Restore Preset Menu WARNINGGE Medical Systems is not responsible for lost data if the suggested...
Page 684
Customizing Your System 16-24 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 System/Backup and Restore Preset Menu (continued) Table 16-13: Backup Preset ParameterDescription Patient Archive Select to back up patient data. User Defined ConfigurationSelect to back up the user-defined configuration settings. Backup Select to begin the backup. Table 16-14: Media Preset ParameterDescription Media Select MO or CD. Table 16-15: Restore Preset ParameterDescription Patient Archive Select to restore...
Page 685
System Presets LOGIQ 5 Basic User Manual 16-25 Direction 2300002-100 Rev. 2 System/Backup and Restore Preset Menu (continued) The detailed section of this menu allows you to restore one area at a time from the user defined configuration. This allows you to selectively restore what you want to restore across multiple machines. Check the box(es) you want to restore, insert the MOD/CD-ROM, and press Restore. Table 16-16: Detailed Restore Preset ParameterDescription Imaging Presets Select to restore...
Page 686
Customizing Your System 16-26 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Backup and restore strategy: patient data To minimize accidental loss of data, perform backup of the patient archives stored on the local hard drive DAILY as described in this section. Use a formatted Magneto-Optical Disk (Backup/Restore disk) to back up patient archives from the hard drive, using the backup procedure described in this section. Data from the Backup/Restore disk may be restored to the local hard...
Page 687
System Presets LOGIQ 5 Basic User Manual 16-27 Direction 2300002-100 Rev. 2 Backup procedure: patient data 1. Insert a formatted MOD (Magneto-Optical Disk) into the MOD drive. NOTE: For information about formatting an MOD, See ‘Formatting removable media’ for more information. 2. On the Touch Panel, press Utility. 3. On the Utility Touch Panel, press System. 4. On the monitor display, select Backup/Restore. The Backup/Restore screen is displayed. See Figure 16-16. NOTE: If you are not logged in...
Page 688
Customizing Your System 16-28 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Restore procedure: patient data 1. Insert the Backup/Restore MOD (Magneto-Optical Disk) into the MOD drive. 2. On the Touch Panel, press Utility. 3. On the Utility Touch Panel, press System. 4. On the monitor display, select Backup/Restore. The Backup/Restore screen is displayed. See Figure 16-16. NOTE: If you are not logged in with administrator privileges, the Operator Login window is displayed. Log on with...
Page 689
System Presets LOGIQ 5 Basic User Manual 16-29 Direction 2300002-100 Rev. 2 Backup and restore strategy: user-defined configurations In addition to generating a safety copy, the backup/restore function of the user-defined configuration (presets) can be used to configure several LOGIQ 5 systems with identical presets (preset synchronization). Preset synchronization The procedure for preset synchronization of several scanners is as follow: 1. Make a backup of the user-defined configurations on a...
Page 690
Customizing Your System 16-30 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Backup procedure: user-defined configurations 1. Insert a formatted MOD (Magneto-Optical Disk) into the MOD drive. NOTE: For information about formatting an MOD, See ‘Formatting removable media’ for more information. 2. On the Touch Panel, press Utility. 3. On the Utility Touch Panel, press System. 4. On the monitor display, select Backup/Restore. The Backup/Restore screen is displayed. See Figure 16-16. NOTE: If...