GE Logiq P5 User Manual
Here you can view all the pages of manual GE Logiq P5 User Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.
Page 291
Measurement and Calculation Setup LOGIQ 5 Basic User Manual 7-41 Direction 2300002-100 Rev. 2 Changing or adding measurement parameters (continued) Adding measurement parametersTo add a measurement parameter: 1. In the Selection menu, select the measurement. 2. To change the tool used to make a measurement: In the Measurement section of the Measurement & Analysis screen, select the desired tool from the Tool list. Select the arrow to display the drop-down list. NOTE: If the Tool field is gray, it...
Page 292
General Measurements and Calculations 7-42 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Changing or adding measurement parameters (continued) Adding measurement parameters (continued)4. In the Measurement section, move the Trackball to an empty line at the bottom of the Parameter list. Press Set. The system adds a parameter with a name of (Name). Figure 7-25. Adding a Parameter
Page 293
Measurement and Calculation Setup LOGIQ 5 Basic User Manual 7-43 Direction 2300002-100 Rev. 2 Changing or adding measurement parameters (continued) Adding measurement parameters (continued)5. To change the name of the Parameter, move the Trackball to the (Name) and press Set twice. Type a name for the parameter. 6. Move the Trackball to the Tool result field, and press Set once. The Edit Formula window is displayed. Figure 7-26. Edit Formula 7. To create a formula: a. In the Value Type field,...
Page 294
General Measurements and Calculations 7-44 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Changing or adding measurement parameters (continued) Formula Unit ConversionWhen you create a formula, the system changes the calculation result into an output unit as defined in the following table. Table 7-2: Formula Unit Conversion Unit Conversion (coefficient value) Time sx1 ms x1,000 min x0.0167 h x0.00027778 Ratio % x100 Frequency bpm or BPM x1.0 Angle rad x1.0 deg x57.2958 grad x63.6620 Distance cm...
Page 295
Measurement and Calculation Setup LOGIQ 5 Basic User Manual 7-45 Direction 2300002-100 Rev. 2 m/s x1 dm/s x10 cm/s x100 mm/s x1,000 inch/s x39.37 Acceleration m/s2 x1 dm/s2 x10 cm/s2 x100 mm/s2 x1,000 inch/s2 x39.37 Area m2 or m^2 x1 dm2 x100 cm2 or cm^2 x10,000 mm2 or mm^2 x1,000,000 inch2 x1550 Vo lu m e m3 x1 dm3 x1,000 cm3 x1,000,000 l x1,000 dl x10,000 cl x100,000 ml x1,000,000 gallon x264,178 quart x1056.71 Vo lu m e Fl o w Table 7-2: Formula Unit Conversion Unit Conversion (coefficient value)
Page 296
General Measurements and Calculations 7-46 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 m3/s x1 dm3/s x1,000 cm3/s x1,000,000 l/s x1,000 dl/s x10,000 cl/s x100,000 ml/s x1,000,000 m3/min x60 dm3/min x60,000 cm3/min x60,000,000 l/min or L/min x60,000 dl/min x600,000 cl/min 6,000,000 ml/min x60,000,000 ml/m2 x1,000,000 Pressure mmHg x1 Pa x133.322 kPa x0.133322 bar x0.00133322 Pressure/Time mmHg/s x1 Mass kg x1 g x1,000 ounce x35.273962 pound x2.2046226 Others mmHG x100 Table 7-2: Formula Unit...
Page 297
Measurement and Calculation Setup LOGIQ 5 Basic User Manual 7-47 Direction 2300002-100 Rev. 2 Changing or adding measurement parameters (continued) Formula Unit Conversion (continued)For example, when a Volume formula is created: Vol [ml or cm3] = 0.523598*{D1}*{D2}*{D3} (D1, D2, and D3 indicate a measurement result.) In this case, the measurement (D1, D2, and D3) is a distance measurement, so the measured data is a meter [m] unit according to the above table. To change into a milliliter, the system...
Page 298
General Measurements and Calculations 7-48 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Changing or adding measurement parameters (continued) Formula Unit Conversion (continued)The system defines the standard value of each measurement as a meter [m]. If the unit of each measurement value of this formula is defined as centimeter [cm], you must define the formula as follows: efw[g] = 10^(1.5662-0.0108*{P1}*100 + 0.0468*{P2}*100 + 0.171*{D1}*100 + 0.00034*{P1}*{P1}*100*100 -...
Page 299
Measurement and Calculation Setup LOGIQ 5 Basic User Manual 7-49 Direction 2300002-100 Rev. 2 Editing Calculations To modify user-defined calculations: In the Measurement menu section, selection Edit Calc. Figure 7-27. Edit Calc
Page 300
General Measurements and Calculations 7-50 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Editing Calculations (continued) The Modify User CALC window is displayed. Figure 7-28. Modify User CALC window In the User Defined list, select the calculation that you want modified, then select OK.