GE Logiq P5 User Manual
Here you can view all the pages of manual GE Logiq P5 User Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.
Page 321
Mode Measurements LOGIQ 5 Basic User Manual 7-71 Direction 2300002-100 Rev. 2 To view a worksheet (continued) The system displays the worksheet for the current study. Figure 7-33. OB Worksheet To return to scanning, do one of the following: • Select the Worksheet Display key. • Press Esc. • Select the Exit button.
Page 322
General Measurements and Calculations 7-72 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 To view a worksheet (continued) To view a different worksheet, select the worksheet key for the desired worksheet. Figure 7-34. Worksheet Display Touch Panel: OB Study To view worksheet data for a particular mode, select the key for that mode. To view a worksheet with data for more than one mode, select Expand Worksheet. If a worksheet has more data on a second page, to view the next page, adjust the...
Page 323
Mode Measurements LOGIQ 5 Basic User Manual 7-73 Direction 2300002-100 Rev. 2 To edit a worksheet To change data on a worksheet: 1. To position the cursor at the field you want to change, move the Trackball. The field is highlighted. 2. Press Set. 3. Type the new data in the field. The new data is displayed in blue to indicate that it was manually entered. To delete or exclude data on a worksheet: 1. To position the cursor at the field you want to delete or exclude, move the Trackball. The field is...
Page 324
General Measurements and Calculations 7-74 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Generic Measurements Overview Each exam category has a Generic study. The Generic studies provide you quick access to measurements such as volume, angle, A/B ratio, and % stenosis. The particular measurements available in each Generic study vary, depending on the exam category and the mode. This section describes generic measurements, organized by mode. To access Generic studies: 1. On the Control...
Page 325
Generic Measurements LOGIQ 5 Basic User Manual 7-75 Direction 2300002-100 Rev. 2 B-Mode Measurements In B-Mode, the Generic study includes the following measurements: • % Stenosis • Volume • Angle • A/B Ratio Figure 7-35. B-Mode Generic Study NOTE: The following instructions assume that you first scan the patient and then press Freeze.
Page 326
General Measurements and Calculations 7-76 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 % Stenosis You can calculate % Stenosis by diameter or by area. Diameter NOTE: When you use diameter to calculate the %stenosis, always take the measurement from a cross-sectional view of the vessel. To calculate percent stenosis by diameter: 1. From the Generic Touch Panel, select % Stenosis. 2. Select %sten(Diam). The system displays an active caliper. 3. Make a distance measurement of the inner area...
Page 327
Generic Measurements LOGIQ 5 Basic User Manual 7-77 Direction 2300002-100 Rev. 2 % Stenosis (continued) AreaTo calculate percent stenosis by area: 1. From the Generic Touch Panel, select % Stenosis. 2. Select %sten(Area). The system displays a trace caliper. 3. Make a trace measurement of the inner area of the blood vessel. NOTE: To erase an open trace, move the Trackball. The system displays a second trace caliper. 4. Make a trace measurement of the outer area of the blood vessel. The system...
Page 328
General Measurements and Calculations 7-78 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Vo l u m e The volume calculation can be made from any of the following measurements: • One distance • Two distances • Three distances • One ellipse • One distance and one ellipse For details on how to make a distance measurement, See ‘Distance measurement’ for more information.. For details on how to make an ellipse measurement, See ‘Circumference and area (ellipse) measurement’ for more...
Page 329
Generic Measurements LOGIQ 5 Basic User Manual 7-79 Direction 2300002-100 Rev. 2 Volume (continued) To make a volume calculation using one ellipse: 1. Select Volume. 2. Make one ellipse measurement. 3. Select Volume. The system displays the ellipse measurement and the volume in the Results Window. To make a volume calculation using one ellipse and one distance: 1. Select Volume. 2. Make one distance measurement and one ellipse measurement. 3. Select Volume. The system displays the distance and...
Page 330
General Measurements and Calculations 7-80 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Volume (continued) Figure 7-36. Volume Calculation Examples 1. One distance 2. Two distances 3. Three distances 4. One ellipse 5. One distance and one ellipseTable 7-3: Volume Calculations Calc NameInput Measurements Volume (spherical) One distance Volume (prolate spheroidal) Two distances, d1>d2 Volume (spheroidal) Three distances Volume (prolate spheroidal) One ellipse: (d1 major axis, d2 minor axis)...