Mode Measurements
LOGIQ 5 Basic User Manual 7-61
Direction 2300002-100 Rev. 2
Circumference and area (trace) measurement
To trace the circumference of a portion of the anatomy and 
calculate its area: 
Measure• Press Measure twice; a trace caliper displays.
• To position the trace caliper at the start point, move the Trackball.
Set• To fix the trace start point, press Set. The trace caliper changes to an active 
caliper. 
• To trace the measurement area, move the Trackball around the anatomy. A 
dotted...

Page 312

General Measurements and Calculations
7-62 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Doppler Mode Measurements
Four basic measurements can be made in Doppler Mode.
• Velocity
•  TAMAX, TAMIN, and TAMEAN (Manual or Auto Trace)
•  Two Velocities with the Slope and Time Interval between 
them
• Time Interval
NOTE:  The following instructions assume that you do the following:
1.  In the B-Mode part of the display, scan the anatomy you 
want to measure.
2.  Go to the Doppler Mode part of the...

Page 313

Mode Measurements
LOGIQ 5 Basic User Manual 7-63
Direction 2300002-100 Rev. 2
TAMAX and TAMEAN (manual)  [TAMIN is not available at this time.] 
The value measured depends upon a preset selection in the... 
The three selections available are: Peak (TAMAX), Floor 
(TAMIN), and Mean (TAMEAN).
To do a manual trace of TAMAX, TAMIN or TAMEAN: 
Measure• Press Measure twice; a trace caliper displays. 
• To position the caliper at the trace start point, move the Trackball.
Set• To fix the start point, press...

Page 314

General Measurements and Calculations
7-64 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
TAMAX and TAMEAN (auto trace)  [TAMIN is not available at this time.] 
The value measured depends upon a preset selection in the... 
The three selections available are: Peak (TAMAX), Floor 
(TAMIN), and Mean (TAMEAN).
To auto trace TAMAX, TAMIN, or TAMEAN: 
Measure• Press Measure twice; an active caliper with a vertical dotted line displays.
• To position the caliper at the trace start point in the Doppler...

Page 315

Mode Measurements
LOGIQ 5 Basic User Manual 7-65
Direction 2300002-100 Rev. 2
Slope and time interval
To measure two velocity values, the time interval (sec), and 
slope (m/s2):
Measure• Press Measure three times; an active caliper with vertical and horizontal 
dotted lines displays.
• To position the caliper at the start point, move the Trackball.
Set• To fix the start point, press Set. The system fixes the first caliper and displays 
a second active caliper.
• To position the second caliper at the end...

Page 316

General Measurements and Calculations
7-66 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Time interval
To measure a horizontal time interval: 
Measure• Press Measure four times; an active caliper with a vertical dotted line 
displays.
• To position the active caliper at the start point, move the Trackball.
Set• To fix the start point, press Set. The system fixes the first caliper and displays 
a second active caliper.
• To position the second caliper at the end point, move the Trackball.
Set• To...

Page 317

Mode Measurements
LOGIQ 5 Basic User Manual 7-67
Direction 2300002-100 Rev. 2
M-Mode Measurements
Basic measurements that can be taken in the M-Mode portion of 
the display are:
• Tissue Depth (Distance)
• Time Interval
•  Time Interval and Slope
NOTE:  The following instructions assume that you do the following:
1.  In the B-Mode part of the display, scan the anatomy you 
want to measure.
2.  Go to the M-Mode part of the display.
3. Press Freeze.
Tissue depth
Tissue depth measurement in M-Mode functions...

Page 318

General Measurements and Calculations
7-68 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Time interval
To measure a horizontal time interval: 
Measure• Press Measure twice; an active caliper with a vertical dotted line displays.
• To position the caliper at the start point, move the Trackball.
Set• To fix the first caliper, press Set. The system fixes the first caliper and 
displays a second active caliper.
• To position the second caliper at the end point, move the Trackball.
Set• To complete...

Page 319

Mode Measurements
LOGIQ 5 Basic User Manual 7-69
Direction 2300002-100 Rev. 2
Time and slope
To measure time and slope between two points: 
MeasurePress Measure three times; an active caliper with a vertical and horizontal 
dotted line displays.
To position the active caliper at the start point, move the Trackball.
SetTo fix the start point, press Set.
The system fixes the first caliper and displays a second active caliper.
To position the second caliper at the end point, move the Trackball.
SetTo...

Page 320

General Measurements and Calculations
7-70 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Viewing and Editing Worksheets
As you complete measurements, the system puts measurement 
data in the appropriate worksheets. 
To view a worksheet
To view a worksheet, select the Worksheet Display key on the 
Touch Panel.
 Figure 7-32. OB-General Study 