GE Logiq P5 User Manual
Here you can view all the pages of manual GE Logiq P5 User Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.
Page 331
Generic Measurements LOGIQ 5 Basic User Manual 7-81 Direction 2300002-100 Rev. 2 Angle This function measures the angle between two intersecting planes. 1. From the Generic Touch Panel, select Angle. The system displays an active caliper. 2. To position the caliper, move the Trackball. 3. To fix the position of the first caliper, press Set. The system displays a second active caliper. 4. To position the second caliper at the apex of the angle, move the Trackball. 5. To fix the position of the...
Page 332
General Measurements and Calculations 7-82 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 A/B Ratio In B-Mode and M-Mode, you can calculate A/B ratio by diameter or by area. DiameterTo calculate A/B ratio by diameter: 1. From the Generic Touch Panel, select A/B Ratio. 2. Select ratio(Diam). The system displays an active caliper. 3. Make a distance measurement of the first velocity. The system displays an active caliper for the second distance measurement. 4. Make a distance measurement of...
Page 333
Generic Measurements LOGIQ 5 Basic User Manual 7-83 Direction 2300002-100 Rev. 2 M-Mode Measurements Figure 7-37. M-Mode Generic Study % Stenosis See ‘% Stenosis’ for more information. A/B Ratio See ‘A/B Ratio’ for more information.
Page 334
General Measurements and Calculations 7-84 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Heart Rate To calculate the heart rate: 1. From the Generic Touch Panel, select HR. The system displays an active caliper. 2. To position the caliper at a recognizable point in the first cycle, move the Trackball. 3. To fix the first caliper, press Set. The system displays a second active caliper. 4. To position the caliper at the identical point in the second or third cycle, move the Trackball....
Page 335
Generic Measurements LOGIQ 5 Basic User Manual 7-85 Direction 2300002-100 Rev. 2 Doppler Mode Measurements In Doppler Mode, the Generic study includes the following measurements: • PI (Pulsatility Index) • RI (Resistive Index) • S/D Ratio • D/S Ratio • A/B Ratio • Max PG (Pressure Gradient) • Mean PG (Pressure Gradient) • SV (Stroke Volume) • HR (Heart Rate) Figure 7-39. Doppler Mode Generic Study NOTE: The following instructions assume that you do the following: 1. In the B-Mode part of the...
Page 336
General Measurements and Calculations 7-86 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Pulsatility Index (PI) 1. From the Doppler Generic Touch Panel, select PI. The system displays a trace caliper. 2. To position the caliper at the beginning of the waveform (V max), move the Trackball. 3. To fix the start point, press Set. The system displays a second active caliper. 4. To trace the waveform from V max to Vd move the Trackball. 5. To complete the measurement, press Set. The system...
Page 337
Generic Measurements LOGIQ 5 Basic User Manual 7-87 Direction 2300002-100 Rev. 2 Resistive Index (RI) 1. From the Doppler Generic Touch Panel, select RI. The system displays an active caliper with vertical and horizontal dotted lines. 2. To position the caliper at the peak systolic velocity, move the Trackball. 3. To fix the measure point, press Set. The system displays a second active caliper. 4. To position the second caliper at the end diastolic velocity, move the Trackball. 5. To complete...
Page 338
General Measurements and Calculations 7-88 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 S/D or D/S Ratio To calculate the Peak Systole/End Diastole ratio or End Diastole/Peak Systole ratio: 1. Select S/D or D/S. The system displays an active caliper with vertical and horizontal dotted lines. 2. To position the caliper at peak systole (PS) or end diastole (ED), move the Trackball. 3. To fix the measure point, press Set. The system displays a second active caliper. 4. To position the...
Page 339
Generic Measurements LOGIQ 5 Basic User Manual 7-89 Direction 2300002-100 Rev. 2 A/B Ratio In Doppler Mode you can measure A/B ratio by velocity, time, or acceleration. Ve l o c i tyTo measure AB ratio by velocity: 1. Select A/B. 2. Select ratio(Velocity). The system displays an active caliper with vertical and horizontal dotted lines. 3. To position the caliper at the A velocity, move the Trackball. 4. To fix the measure point, press Set. The system displays a second active caliper. 5. To...
Page 340
General Measurements and Calculations 7-90 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 A/B Ratio (continued) AccelerationTo measure A/B ratio by acceleration: 1. Select A/B. 2. Select ratio(Acc). The system displays an active caliper. 3. Make a distance measurement of the A acceleration point. a. To position the active caliper at the start point, move the Trackball. b. To fix the start point, press Set. The system fixes the first caliper and displays a second active caliper. c. To...