Optimizing Spectral Doppler LOGIQ 5 Basic User Manual 5-57 Direction 2300002-100 Rev. 2 Figure 5-16. CW Display (Actual Image Not Yet Available) HPRF HPRF mode is used when detected velocities exceed the processing capabilities of the currently selected PW Doppler scale or when the selected anatomical site is too deep for the selected PW Doppler scale. Time Scale Eight speeds: 12, 8, 6, 5, 4, 3, 2, and 1 second sweeps. Angle Correct Indicates flow direction. Sample Volume GateIndicates sample...
Page 182
Optimizing the Image 5-58 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Doppler Mode Scanning Hints Control overview Dynamic Range. Affects the amount of Doppler amplitude data displayed. Wall Filter. Removes the noise caused by vessel or heart wall motion at the expense of low flow sensitivity. Sweep Speed. Controls speed of spectral update. HINTSThe best Doppler data is collected when parallel to flow, with orientation also parallel to the anatomic target; whereas, the best B-Mode image...
Page 183
Optimizing Spectral Doppler LOGIQ 5 Basic User Manual 5-59 Direction 2300002-100 Rev. 2 Doppler Mode Touch Panel Figure 5-17. PW Mode Touch Panel, Page 1 Figure 5-18. PW Mode Touch Panel, Page 2
Page 184
Optimizing the Image 5-60 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Doppler sample volume gate position (Trackball) DescriptionMoves the sample volume gate on the B-Modes Doppler Mode cursor. The gate is positioned over a specific position within the vessel. AdjustingTo move Doppler Mode cursor position, move Trackball left or right until positioned over the vessel. To move sample volume gate position, move Trackball up or down until positioned inside the vessel. Va l u e sDefaults...
Page 185
Optimizing Spectral Doppler LOGIQ 5 Basic User Manual 5-61 Direction 2300002-100 Rev. 2 PRF DescriptionAdjusts the velocity scale to accommodate faster/slower blood flow velocities. Velocity scale determines pulse repetition frequency. If the sample volume gate range exceeds single gate PRF capability, the system automatically switches to high PRF mode. Multiple gates appear, and HPRF is indicated on the display. High PRFHigh Pulse Repetition Frequency (HPRF) is a special operating mode of PW...
Page 186
Optimizing the Image 5-62 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Angle Correct DescriptionEstimates the flow velocity in a direction at an angle to the Doppler vector by computing the angle between the Doppler vector and the flow to be measured. NOTE: When the Doppler Mode Cursor and angle correct indicator are aligned (the angle is O), you cannot see the angle correct indicator. AdjustingFlow toward the probe is mapped above the baseline and right speaker (as you are facing the...
Page 187
Optimizing Spectral Doppler LOGIQ 5 Basic User Manual 5-63 Direction 2300002-100 Rev. 2 Wall Filter DescriptionInsulates the Doppler signal from excessive noise caused from vessel movement. AdjustingTo increase/decrease, select Wall Filter, then adjust Wall Filter on the Touch Panel (PRF and Wall Filter share the same control. Press the control to toggle between PRF and Wall Filter). Each adjustment cycles you to the next setting. Va l u e sValues vary, depending upon the probe and application....
Page 188
Optimizing the Image 5-64 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 M/D Cursor DescriptionDisplays the Doppler Mode cursor on the B-Mode image. AdjustingTo activate/deactivate the Doppler Mode cursor, press M/D Cursor. Trackball to position sample volume graphic. Adjust SV Length on the Touch Panel to adjust sample volume gate size. The M/D Cursor key backlights. Va l u e sOn/Off. If blood flow velocities cannot be presented, the message PRF unavailable. Change Vel Scale appears....
Page 189
Optimizing Spectral Doppler LOGIQ 5 Basic User Manual 5-65 Direction 2300002-100 Rev. 2 Audio Volume DescriptionControls audio output. AdjustingTo adjust the audio, adjust Vo l u m e left/right. The volume defaults to the last Doppler audio volume setting. Va l u e sUsually, a one-third adjustment of the dial gives you the best volume. The volume decreases/increases logarithmically, with positive flow toward the probe through the right speaker; negative flow away from the probe through the left...
Page 190
Optimizing the Image 5-66 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Dynamic Range DescriptionDynamic range controls how echo intensities are converted to shades of gray, thereby increasing the range of contrast you can adjust. AdjustingYou access Doppler dynamic range from the Doppler Mode Touch Panel. Each click cycles you to the next setting. To increase/decrease, press Dynamic Range. Va l u e sThe settings cycle in 4 dB steps from 20 dB to 40 dB. The current value displays on the...