Page 151

Optimizing M-Mode
LOGIQ 5 Basic User Manual 5-27
Direction 2300002-100 Rev. 2
M-Mode Display
 Figure 5-4. M-Mode Display
M-Mode Display
Description, Format, Values
(For Future Vadliation)
M-Mode Gain
Depth Scale  
IndicatorDisplays Overall M-Mode Gain as ##dB. 
Image depth, in cm.
Time Scale Eight speeds: 16, 12, 8, 6, 4, 3, 2, 1. Major marker = 1 second. 
M-Mode Cursor Displays where the M-Mode Cursor is positioned.  

Page 152

Optimizing the Image
5-28 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
M-Mode Touch Panel
 Figure 5-5. M-Mode Touch Panel
Scanning Hints
HINTSThese M-Mode controls produce the following results: 
Edge Enhance. Affects border sharpness. 
Sweep Speed. Controls speed of M-Mode update. 
Dynamic Range. Affects the amount of gray scale information 
displayed.  

Page 153

Optimizing M-Mode
LOGIQ 5 Basic User Manual 5-29
Direction 2300002-100 Rev. 2
M/D Cursor
DescriptionDisplays the M-Mode cursor on the B-Mode image. 
AdjustingTo activate/deactivate the M-Mode cursor, press M/D Cursor. 
Trackball to position M-Mode cursor. 
Va l u e sOn/Off.
BenefitsLets you position the cursor before you go into M-Mode so that 
you can make optimum use of the larger B-Mode image. 
B Pause
DescriptionToggles between simultaneous and update presentation while 
viewing Spectral Doppler....

Page 154

Optimizing the Image
5-30 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Sweep Speed
DescriptionChanges the speed at which the timeline is swept. Available in 
M-Mode, M Color Flow Mode, and Doppler Mode. 
AdjustingTo increase/decrease, select Sweep Speed. 
Va l u e sM-Mode: 16, 12, 8, 6, 4, 3, 2, and 1 second  sweeps across the 
display. Doppler Mode: 12, 8, 6, 5, 4, 3, 2, and 1 second 
sweeps. Values vary by probe and application. 
Sweep Speed values are returned to the factory or user preset...

Page 155

Optimizing M-Mode
LOGIQ 5 Basic User Manual 5-31
Direction 2300002-100 Rev. 2
Anatomical M-Mode
DescriptionAnatomical M-Mode gives you the ability to manipulate the 
cursor at different angles and positions. The M-Mode display 
changes according to a motion of the M cursor.
 Figure 5-6. Anatomical M Mode
ActivatingTo activate Anatomical M-Mode, press the Anatomical M Touch 
Panel control.   

Page 156

Optimizing the Image
5-32 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Anatomical M-Mode (continued)
AdjustingUse the Touch Panel control and Trackball to position the 
M cursor over the required area of the B-Mode image.
•  Turn the Touch Panel control left or right to move the M 
cursor to the desired area. 
•  Move the Trackball left/right to move the cursor line left/right.
•  Move the Trackball up/down to move the M cursor up/down.
Va l u e sOn/Off. 
BenefitsColor Flow Mode and Color...

Page 157

Optimizing M Color Flow
LOGIQ 5 Basic User Manual 5-33
Direction 2300002-100 Rev. 2
Optimizing M Color Flow
M Color Flow Mode
DescriptionM Color Flow is used for fetal cardiac applications. Color Flow 
overlays color on the M-Mode image using velocity and variance 
color maps. The Color Flow wedge overlays the B-Mode image 
and M-Mode timeline.
The Color Flow maps available in M-Mode are the same as in 
Color Flow Mode. The size and position of the Color Flow 
window in B-Mode determines the size and...

Page 158

Optimizing the Image
5-34 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
M Color Flow Mode (continued)
BioeffectsChanging the Sweep Speed, Packet Size, Frame Rate/
Resolution, Zoom, PRF, and ROI size may change the TI and/or 
MI. Observe the output display for possible effects. 
 Figure 5-7. M Color Flow, Page 1
 Figure 5-8. M Color Flow, Page 2 

Page 159

Optimizing Color Flow
LOGIQ 5 Basic User Manual 5-35
Direction 2300002-100 Rev. 2
Optimizing Color Flow
Intended Use
Color Flow Mode is a Doppler Mode intended to add color-
coded qualitative information concerning the relative velocity 
and direction of fluid motion within the B-Mode image.
Introduction
A typical examination using Color Flow Mode, 
1.  Follow the same procedure as described under B-Mode to 
locate the anatomical area of interest. 
2.  After optimizing the B-Mode image, add Color Flow....

Page 160

Optimizing the Image
5-36 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Activating Color Flow
To activate Color Flow Mode, 
1. Press CF (Color Flow). The CF window appears over the B-
Mode image. Move the Trackball to move the CF window.
 Figure 5-9. Color Flow Mode Display
Uses
Color Flow is useful to see flow in a broad area. Color Flow 
allows visualization of flow in the whole image, whereas Doppler 
Mode provides spectral information in a smaller area. 
Color Flow is also sometimes used as...