Page 131

Optimizing B-Mode
LOGIQ 5 Basic User Manual 5-7
Direction 2300002-100 Rev. 2
Gain
DescriptionB-Mode Gain increases or decreases the amount of echo 
information displayed in an image. It may have the effect of 
brightening or darkening the image if sufficient echo information 
is generated. 
AdjustingGain values vary depending on the probe; they are not 
associated with a particular position of the knob. 
NOTE:  B-Mode gain is independent of M-Mode and Doppler and Color 
Flow Gain. Changing the gain while...

Page 132

Optimizing the Image
5-8 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Focus
DescriptionIncreases the number of focal zones or moves the focal zone(s) 
so that you can tighten up the beam for a specific area. A 
graphic caret corresponding to the focal zone position(s) 
appears on the right edge of the image. 
AdjustingTo increase/decrease the number of focal zones, adjust Focus 
Number. 
NOTE:  Focus Position and Focus Number share the same control. 
Press the control to toggle between Focus...

Page 133

Optimizing B-Mode
LOGIQ 5 Basic User Manual 5-9
Direction 2300002-100 Rev. 2
Auto Optimize (Auto)
DescriptionAuto lets you optimize the image based upon a specified region 
of interest or anatomy within the B-Mode image or Spectral 
Doppler in order to improve the contrast of the displayed image 
data. Auto Optimize is available in B-Mode, specifically in single 
or multi image, on live, frozen or CINE images, and while in 
zoom and in Spectral Doppler. However, it is not available in 
Display Zoom or...

Page 134

Optimizing the Image
5-10 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
B Flow 
DescriptionB Flow is intended to provide a more intuitive representation of 
non-quantitative hemodynamics in vascular structures. B Flow is 
available on the 10L probe.
All B-Mode measurements are available with B Flow active: 
depth, distance along a straight line, % stenosis, volume, trace, 
circumference, and enclosed area. 
ActivatingTo activate/deactivate B Flow, press B Flow. Doppler Mode is 
available while...

Page 135

Optimizing B-Mode
LOGIQ 5 Basic User Manual 5-11
Direction 2300002-100 Rev. 2
B Flow (continued)
BenefitsB Flow is less angle-independent, no velocity-aliasing artifacts, 
and provides better resolution when compared with Color 
Doppler Mode, and displays a full field of view. B Flow is 
therefore a more realistic (intuitive) representation of flow 
information, allowing you to view both high and low velocity flow 
at the same time. 
Affect on other 
controlsWhen you activate B Flow, the image remembers...

Page 136

Optimizing the Image
5-12 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Coded Harmonic Imaging (CHI)
DescriptionHarmonic imaging utilizes Digitally Encoded Ultrasound (DEU). 
Coded Harmonics enhances near field resolution for improved 
small parts imaging as well as far field penetration. 
AdjustingTo activate Coded Harmonic imaging, 
Press Harmonics. 
NOTE:  Multi frequency is not active when the image is frozen. 
Va l u e sOn/Off. See Chapter 17 for probe availability. `CHI appears next 
to...

Page 137

Optimizing B-Mode
LOGIQ 5 Basic User Manual 5-13
Direction 2300002-100 Rev. 2
Frequency
DescriptionMulti Frequency mode lets you downshift to the probes next 
lower frequency or shift up to a higher frequency. 
AdjustingTo select a new frequency, 
1. Select Frequency.   
2.  Select the desired frequency. Harmonic frequencies are 
currently available via Frequency. 
NOTE:  Frequency change is not active when the image is frozen. 
The selected frequency appears in the upper, right-hand portion 
of the...

Page 138

Optimizing the Image
5-14 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Virtual Convex
DescriptionOn Linear and Sector probes, Virtual Convex provides a larger 
field of view in the far field.
NOTE:  Color Flow can be steered when in straight fire, but not in Virtual 
Convex. 
AdjustingTo activate/deactivate Virtual Convex, select Virtual Convex. 
Va l u e sOn/Off. 
BenefitsVirtual Convex can be especially helpful in Obstetrics. Available 
in B-Mode, Color Flow Mode, and Doppler Mode....

Page 139

Optimizing B-Mode
LOGIQ 5 Basic User Manual 5-15
Direction 2300002-100 Rev. 2
Scan Area
DescriptionYou can widen or narrow the size of the sector angle to 
maximize the images region of interest (ROI). 
AdjustingTo narrow/widen the angle, press Scan Area and move the 
Trackball. 
Va l u e sVaries, depending upon the probe (not applicable to linear 
probes) and application. 
BenefitsIncrease the sector angle to see a wide field of view; decrease 
the sector angle when you need to have a faster frame rate,...

Page 140

Optimizing the Image
5-16 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Reverse
DescriptionFlips the image 180 degrees left/right. 
AdjustingTo flip the image 180 degrees, press Rotate. Linear probes do 
not rotate. 
Va l u e sThe image rotates in 180 degrees left/right. Rotation settings 
vary by probe and application. 
Rotation settings are returned to the factory or user preset value 
when you change the following: Probe, Exam Category, Exam 
Calcs, or New Patient. 
BenefitsUsed for...