GE Logiq P5 User Manual
Here you can view all the pages of manual GE Logiq P5 User Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.
Page 131
Optimizing B-Mode LOGIQ 5 Basic User Manual 5-7 Direction 2300002-100 Rev. 2 Gain DescriptionB-Mode Gain increases or decreases the amount of echo information displayed in an image. It may have the effect of brightening or darkening the image if sufficient echo information is generated. AdjustingGain values vary depending on the probe; they are not associated with a particular position of the knob. NOTE: B-Mode gain is independent of M-Mode and Doppler and Color Flow Gain. Changing the gain while...
Page 132
Optimizing the Image 5-8 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Focus DescriptionIncreases the number of focal zones or moves the focal zone(s) so that you can tighten up the beam for a specific area. A graphic caret corresponding to the focal zone position(s) appears on the right edge of the image. AdjustingTo increase/decrease the number of focal zones, adjust Focus Number. NOTE: Focus Position and Focus Number share the same control. Press the control to toggle between Focus...
Page 133
Optimizing B-Mode LOGIQ 5 Basic User Manual 5-9 Direction 2300002-100 Rev. 2 Auto Optimize (Auto) DescriptionAuto lets you optimize the image based upon a specified region of interest or anatomy within the B-Mode image or Spectral Doppler in order to improve the contrast of the displayed image data. Auto Optimize is available in B-Mode, specifically in single or multi image, on live, frozen or CINE images, and while in zoom and in Spectral Doppler. However, it is not available in Display Zoom or...
Page 134
Optimizing the Image 5-10 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 B Flow DescriptionB Flow is intended to provide a more intuitive representation of non-quantitative hemodynamics in vascular structures. B Flow is available on the 10L probe. All B-Mode measurements are available with B Flow active: depth, distance along a straight line, % stenosis, volume, trace, circumference, and enclosed area. ActivatingTo activate/deactivate B Flow, press B Flow. Doppler Mode is available while...
Page 135
Optimizing B-Mode LOGIQ 5 Basic User Manual 5-11 Direction 2300002-100 Rev. 2 B Flow (continued) BenefitsB Flow is less angle-independent, no velocity-aliasing artifacts, and provides better resolution when compared with Color Doppler Mode, and displays a full field of view. B Flow is therefore a more realistic (intuitive) representation of flow information, allowing you to view both high and low velocity flow at the same time. Affect on other controlsWhen you activate B Flow, the image remembers...
Page 136
Optimizing the Image 5-12 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Coded Harmonic Imaging (CHI) DescriptionHarmonic imaging utilizes Digitally Encoded Ultrasound (DEU). Coded Harmonics enhances near field resolution for improved small parts imaging as well as far field penetration. AdjustingTo activate Coded Harmonic imaging, Press Harmonics. NOTE: Multi frequency is not active when the image is frozen. Va l u e sOn/Off. See Chapter 17 for probe availability. `CHI appears next to...
Page 137
Optimizing B-Mode LOGIQ 5 Basic User Manual 5-13 Direction 2300002-100 Rev. 2 Frequency DescriptionMulti Frequency mode lets you downshift to the probes next lower frequency or shift up to a higher frequency. AdjustingTo select a new frequency, 1. Select Frequency. 2. Select the desired frequency. Harmonic frequencies are currently available via Frequency. NOTE: Frequency change is not active when the image is frozen. The selected frequency appears in the upper, right-hand portion of the...
Page 138
Optimizing the Image 5-14 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Virtual Convex DescriptionOn Linear and Sector probes, Virtual Convex provides a larger field of view in the far field. NOTE: Color Flow can be steered when in straight fire, but not in Virtual Convex. AdjustingTo activate/deactivate Virtual Convex, select Virtual Convex. Va l u e sOn/Off. BenefitsVirtual Convex can be especially helpful in Obstetrics. Available in B-Mode, Color Flow Mode, and Doppler Mode....
Page 139
Optimizing B-Mode LOGIQ 5 Basic User Manual 5-15 Direction 2300002-100 Rev. 2 Scan Area DescriptionYou can widen or narrow the size of the sector angle to maximize the images region of interest (ROI). AdjustingTo narrow/widen the angle, press Scan Area and move the Trackball. Va l u e sVaries, depending upon the probe (not applicable to linear probes) and application. BenefitsIncrease the sector angle to see a wide field of view; decrease the sector angle when you need to have a faster frame rate,...
Page 140
Optimizing the Image 5-16 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Reverse DescriptionFlips the image 180 degrees left/right. AdjustingTo flip the image 180 degrees, press Rotate. Linear probes do not rotate. Va l u e sThe image rotates in 180 degrees left/right. Rotation settings vary by probe and application. Rotation settings are returned to the factory or user preset value when you change the following: Probe, Exam Category, Exam Calcs, or New Patient. BenefitsUsed for...