Page 81

Powering the System
LOGIQ 5 Basic User Manual 3-21
Direction 2300002-100 Rev. 2
Connecting and Using the System (continued)
 Figure 3-10. Example Plug and Outlet Configurations
1.   100-120 VAC, 1200 VA
 Plug and Outlet Configuration
2.   220-240 VAC, 1200 VA
 Plug and Outlet Configuration
WARNINGTo avoid risk of fire, the system power must be supplied from a 
separate, properly rated outlet. See ‘Before the system arrives’ 
for more information.
Under no circumstances should the AC power plug be...

Page 82

Preparing the System for Use
3-22 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Acclimation Time
After being transported, the unit requires one hour for each 2.5° 
increment its temperature is below 10° C or above 40° C.
Table 3-3: System Acclimation Time Chart
° C60 55 50 45 40 35 30 25 20 15 10
° F140 131 122 113 104 95 86 77 68 59 50
hours86420000000
° C5 0 -5 -10 -15 -20 -25 -30 -35 -40
° F41 32 23 14 5 -4 -13 -22 -31 -40
hours2 4 6 8 101214161820 

Page 83

Powering the System
LOGIQ 5 Basic User Manual 3-23
Direction 2300002-100 Rev. 2
Power On/Standby 
 Figure 3-11. Power On/Standby Switch Location
CAUTIONPress the Power On/Standby switch to turn the power on. The 
circuit breaker must also be in the on position. For circuit 
breaker location, see Figure 3-16.
Composite
B/WB/W PrinterPower220-240V  500VA MaxIncluding rear panel 

Page 84

Preparing the System for Use
3-24 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Power Up Sequence 
The system is initialized. During this time:
•  System diagnostics run. Its status is reflected on the monitor 
by the graphic in Figure 3-12.
 Figure 3-12. Power Up Graphic Sequence
•  Probes are initialized for immediate operation.
NOTE:  If no probe is connected, the system goes into standby mode.
•  Peripheral devices are activated on power up.
After initialization is complete, all lighted...

Page 85

Powering the System
LOGIQ 5 Basic User Manual 3-25
Direction 2300002-100 Rev. 2
Password Protection
LoginPersonal IDs and associated passwords can be preset in the 
LOGIQ 7. 
If IDs and passwords have been entered and the Use Auto 
Logon is off, the Operator Login Window appears, requesting for 
an ID and password when Power up sequence is completed or 
when its needed.
 Figure 3-13. Operator Login Window
1.  Operator: Select the Operator.
2.  Password: Enter Operator’s password.
3.  Dataflow: Select...

Page 86

Preparing the System for Use
3-26 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Password Protection (continued)
LogoffWhen the user enters the system by login, at the end of the 
exam the system displays the SYSTEM-EXIT window. Select 
“Logoff” to logoff the system.
 Figure 3-14. System Exit Window 

Page 87

Powering the System
LOGIQ 5 Basic User Manual 3-27
Direction 2300002-100 Rev. 2
Standby Mode
To place the system in standby mode:
1. Press the Power On/Standby switch at the front of the 
system once.
2.  The System-Exit window is displayed.
3. Using the Trackball or Tab key, select Shutdown.
The shutdown process takes a few seconds and is 
completed when the control panel illumination is turned off.
The shutdown process takes a few seconds to complete.
 Figure 3-15. System-Exit Window
4.  Disconnect the...

Page 88

Preparing the System for Use
3-28 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Power Off
To power off the system:
1. Press the Power On/Standby switch at the front of the 
system once.
2.  The System-Exit window is displayed.
3. Using the Trackball or Tab key, select Shutdown.
The shutdown process takes a few seconds and is 
completed when the control panel illumination is turned off.
4.  Disconnect the probes.
Clean or disinfect all probes as necessary. Store them in 
their shipping cases to...

Page 89

Powering the System
LOGIQ 5 Basic User Manual 3-29
Direction 2300002-100 Rev. 2
Circuit breaker 
The Circuit Breaker is located at the rear panel of the system. 
On supplies main power to all internal systems. Off removes 
main power from all internal systems. The circuit breaker 
automatically shuts off power to the system in case of a power 
overload.
If a power overload occurs:
1.  Turn off all peripheral devices.
2.  Reactivate the Circuit Breaker switch.
The Circuit Breaker switch should stay in the...

Page 90

Preparing the System for Use
3-30 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Adjusting the Display Monitor
Rotate, tilt, raise and lower the monitor
The monitor position can be adjusted for easy viewing.
•  The monitor can be rotated around its central pivot point.
•   The monitor can be tilted for the optimum viewing angle.
•  The monitor and console panel can be raised or lowered for 
the best viewing height.
The monitor position can be adjusted for easy viewing.
NOTE:  When moving the...