Page 91

Adjusting the Display Monitor
LOGIQ 5 Basic User Manual 3-31
Direction 2300002-100 Rev. 2
Brightness and Contrast (continued)
To adjust the brightness:
1.  Press the Toggle button on the display monitor twice.
2.  Press the Adjustment (+) button to increase brightness.  
Press the Adjustment (-) button to decrease brightness.
The amount of brightness is shown on a slide bar on the 
screen.
To adjust the contrast:
1.  Press the Toggle button on the display monitor once.
2.  Press the Adjustment (+) button...

Page 92

Preparing the System for Use
3-32 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Brightness and Contrast (continued)
Record any changes to the final brightness and contrast settings 
and leave this information with the system. Generally speaking, 
do not change the controls once they have been set. Once set, 
the display then becomes the reference for the hard copy 
device(s).
NOTE:  After readjusting the monitors Contrast and Brightness, readjust 
all preset and peripheral settings. 

Page 93

Adjusting the Display Monitor
LOGIQ 5 Basic User Manual 3-33
Direction 2300002-100 Rev. 2
Manual Degauss
Degaussing refers to the process of removing magnetic-field 
effects from the monitor. Operation of the monitor within a 
magnetic field may adversely effect color purity. Degaussing can 
be used to correct this problem. The monitor automatically 
degausses at power on.
To activate the manual degauss:
1.  Press and hold the monitor adjustment button for a minute. 
The monitor adjustment menu appears....

Page 94

Preparing the System for Use
3-34 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Probes
Introduction
Only use approved probes. 
All imaging probes can be plugged into any of the probe ports.
Refer to the Probes chapter.
Selecting a probe
•  Always start out with a probe that provides optimum focal 
depths and penetration for the patient size and exam.
•  Begin the scan session using the default Power Output 
setting for the probe and exam.
NOTE:  Selecting a new probe unfreezes the image. 

Page 95

Probes
LOGIQ 5 Basic User Manual 3-35
Direction 2300002-100 Rev. 2
Connecting the Probe
Probes can be connected at any time, regardless of whether the 
console is powered on or off. To ensure that the ports are not 
active, place the system in the image freeze condition.
To connect a probe:
1.  Place the probes carrying case on a stable surface and 
open the case.
2.  Carefully remove the probe and unwrap the probe cord.
3.  DO NOT allow the probe head to hang free. Impact to the 
probe head could result...

Page 96

Preparing the System for Use
3-36 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Connecting the Pencil Probe (P2D and P6D)
When using a pencil probe, it is necessary to attach a probe 
adapter to the system.
To attach a probe adapter and connect a pencil probe:
Turn the locking handle counterclockwise to unlock.
Align the connector of the adapter with the probe port and 
carefully push into place.
 Figure 3-18. Align the connector with the probe port
Turn the connector locking handle clockwise to...

Page 97

Probes
LOGIQ 5 Basic User Manual 3-37
Direction 2300002-100 Rev. 2
Connecting the Pencil Probe (P2D and P6D) (continued)
Insert the probe connector into the adapter until you hear the 
locking click. Carefully position the probe cord so it is free to 
move and is not resting on the floor.
 Figure 3-19. Insert the probe connector to the adapter
WARNINGWhen connecting the probe adapter to the system, first verify 
the direction of the connector. The connector portion of the 
adapter must be properly...

Page 98

Preparing the System for Use
3-38 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Cable Handling 
Take the following precautions with probe cables:
•  Keep free from wheels
•  Do not bend the cable acutely
•  Avoid crossing cables between probes. 

Page 99

Probes
LOGIQ 5 Basic User Manual 3-39
Direction 2300002-100 Rev. 2
Activating the Probe
To activate the probe, select the appropriate probe from the 
probe indicators on the Touch Panel.
 Figure 3-20. Probe Indicators
1. Probe Indicators
The probe activates in the currently-selected operating mode. 
The probes default settings for the mode and selected exam are 
used automatically.
1 

Page 100

Preparing the System for Use
3-40 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Deactivating the Probe
When deactivating the probe, the probe is automatically placed 
in standby mode.
To deactivate a probe:
Disconnecting the Probe
Probes can be disconnected at any time. However, the probe 
should not be selected as the active probe.
•  Move the probe locking handle counterclockwise.
•  Pull the probe and connector straight out of the probe port.
•  Carefully slide the probe and connector away...