Page 91
Adjusting the Display Monitor LOGIQ 5 Basic User Manual 3-31 Direction 2300002-100 Rev. 2 Brightness and Contrast (continued) To adjust the brightness: 1. Press the Toggle button on the display monitor twice. 2. Press the Adjustment (+) button to increase brightness. Press the Adjustment (-) button to decrease brightness. The amount of brightness is shown on a slide bar on the screen. To adjust the contrast: 1. Press the Toggle button on the display monitor once. 2. Press the Adjustment (+) button...
Page 92
Preparing the System for Use 3-32 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Brightness and Contrast (continued) Record any changes to the final brightness and contrast settings and leave this information with the system. Generally speaking, do not change the controls once they have been set. Once set, the display then becomes the reference for the hard copy device(s). NOTE: After readjusting the monitors Contrast and Brightness, readjust all preset and peripheral settings.
Page 93
Adjusting the Display Monitor LOGIQ 5 Basic User Manual 3-33 Direction 2300002-100 Rev. 2 Manual Degauss Degaussing refers to the process of removing magnetic-field effects from the monitor. Operation of the monitor within a magnetic field may adversely effect color purity. Degaussing can be used to correct this problem. The monitor automatically degausses at power on. To activate the manual degauss: 1. Press and hold the monitor adjustment button for a minute. The monitor adjustment menu appears....
Page 94
Preparing the System for Use 3-34 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Probes Introduction Only use approved probes. All imaging probes can be plugged into any of the probe ports. Refer to the Probes chapter. Selecting a probe • Always start out with a probe that provides optimum focal depths and penetration for the patient size and exam. • Begin the scan session using the default Power Output setting for the probe and exam. NOTE: Selecting a new probe unfreezes the image.
Page 95
Probes LOGIQ 5 Basic User Manual 3-35 Direction 2300002-100 Rev. 2 Connecting the Probe Probes can be connected at any time, regardless of whether the console is powered on or off. To ensure that the ports are not active, place the system in the image freeze condition. To connect a probe: 1. Place the probes carrying case on a stable surface and open the case. 2. Carefully remove the probe and unwrap the probe cord. 3. DO NOT allow the probe head to hang free. Impact to the probe head could result...
Page 96
Preparing the System for Use 3-36 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Connecting the Pencil Probe (P2D and P6D) When using a pencil probe, it is necessary to attach a probe adapter to the system. To attach a probe adapter and connect a pencil probe: Turn the locking handle counterclockwise to unlock. Align the connector of the adapter with the probe port and carefully push into place. Figure 3-18. Align the connector with the probe port Turn the connector locking handle clockwise to...
Page 97
Probes LOGIQ 5 Basic User Manual 3-37 Direction 2300002-100 Rev. 2 Connecting the Pencil Probe (P2D and P6D) (continued) Insert the probe connector into the adapter until you hear the locking click. Carefully position the probe cord so it is free to move and is not resting on the floor. Figure 3-19. Insert the probe connector to the adapter WARNINGWhen connecting the probe adapter to the system, first verify the direction of the connector. The connector portion of the adapter must be properly...
Page 98
Preparing the System for Use 3-38 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Cable Handling Take the following precautions with probe cables: • Keep free from wheels • Do not bend the cable acutely • Avoid crossing cables between probes.
Page 99
Probes LOGIQ 5 Basic User Manual 3-39 Direction 2300002-100 Rev. 2 Activating the Probe To activate the probe, select the appropriate probe from the probe indicators on the Touch Panel. Figure 3-20. Probe Indicators 1. Probe Indicators The probe activates in the currently-selected operating mode. The probes default settings for the mode and selected exam are used automatically. 1
Page 100
Preparing the System for Use 3-40 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Deactivating the Probe When deactivating the probe, the probe is automatically placed in standby mode. To deactivate a probe: Disconnecting the Probe Probes can be disconnected at any time. However, the probe should not be selected as the active probe. • Move the probe locking handle counterclockwise. • Pull the probe and connector straight out of the probe port. • Carefully slide the probe and connector away...