Page 141

Optimizing B-Mode
LOGIQ 5 Basic User Manual 5-17
Direction 2300002-100 Rev. 2
Line Density
DescriptionOptimizes B-Mode frame rate or spatial resolution for the best 
possible image. 
AdjustingPress Line Density right to increase resolution or left to 
increase frame rate. 
Va l u e sVaries by probe.  
NOTE:  Not available in timeline. 
Values vary by probe and application. Line Density values are 
returned to the factory or user preset value when you change 
the following: Probe, Exam Category, Exam...

Page 142

Optimizing the Image
5-18 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Map
DescriptionThe system supplies B, M, and Doppler Mode system maps.
AdjustingTo select a map, press Gray Map. A map window displays 
adjacent to the image. Move the Trackball to select the map. 
The image reflects the map as you scroll through the selections. 
Press Set to select. 
Va l u e sGray maps gradually change from least contrasty (Map A) to 
most contrasty (Map R). 
Map values vary by probe, application, and...

Page 143

Optimizing B-Mode
LOGIQ 5 Basic User Manual 5-19
Direction 2300002-100 Rev. 2
M Assigns a lesser amount of gray scale pixel values to weak reflections in the image than Map 
C. Useful when imaging abdomen and liver. 
N S-shaped map. This map highlights tissue differentiation for a certain band of signals. Useful 
when imaging structures where you want more contrast, e.g., renal, tendon, vein, carotid, 
thyroid, breast, Cardiology, etc. 
O S-shaped map. This map highlights tissue differentiation for a...

Page 144

Optimizing the Image
5-20 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Frame Average
DescriptionTemporal filter that averages frames together, thereby using 
more pixels to make up one image. This has the effect of 
presenting a smoother, softer image. 
AdjustingTo adjust frame averaging, adjust Frame Average on the Touch 
Panel. 
Va l u e s0-5, with 5=high.  The current value displays on the Touch 
Panel. Frame Average values vary by probe, application, and 
multi frequency setting. 
Frame...

Page 145

Optimizing B-Mode
LOGIQ 5 Basic User Manual 5-21
Direction 2300002-100 Rev. 2
Colorize
DescriptionColorize is the colorization of a conventional B-Mode image or 
Doppler Spectrum to enhance the users ability to discern B, M, 
and Doppler Mode intensity valuations. Colorize is NOT a 
Doppler Mode. 
NOTE:  You can colorize realtime or CINE images or Timeline CINE, but 
not VCR images. 
Colorizes the gray scale image to enhance the eyes 
discrimination capability. 
Spectrum Colorize colorizes the spectrum...

Page 146

Optimizing the Image
5-22 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Edge Enhance
DescriptionEdge Enhance brings out subtle tissue differences and 
boundaries by enhancing the gray scale differences 
corresponding to the edges of structures. Adjustments to M-
Modes edge enhancement affects the M-Mode only. 
AdjustingTo cycle through settings, adjust Edge Enhance on the Touch 
Panel. 
Va l u e s0-5, with 0 = Off and 5 = High. The current value displays on the 
Touch Panel. Values vary by...

Page 147

Optimizing B-Mode
LOGIQ 5 Basic User Manual 5-23
Direction 2300002-100 Rev. 2
Rejection
DescriptionSelects a level below which echoes will not be amplified (an 
echo must have a certain minimum amplitude before it will be 
processed).
AdjustingAdjust to the right to increase and left to decrease.  
Va l u e s0-5
BenefitsAllows for the elimination from the display of low level echoes 
caused by noise. 

Page 148

Optimizing the Image
5-24 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
LOGIQView
DescriptionLOGIQView provides the ability to construct and view a static 2D 
image which is wider than the field of view of a given transducer. 
This feature allows viewing and measurements of anatomy that 
is larger than what would fit in a single image. Examples include 
scanning of vascular structures and connective tissues in the 
arms and legs. 
LOGIQView constructs the extended image from individual 
image...

Page 149

Optimizing B-Mode
LOGIQ 5 Basic User Manual 5-25
Direction 2300002-100 Rev. 2
LOGIQView (continued)
4.  To restart the scan, press ‘L’ again. 
5.  To complete the scan, press ‘R’ or Freeze (or allow the scan 
to auto complete. The LOGIQView is then displayed, scaled 
to fit entirely on the screen. 
6.  Perform measurements and record images. 
NOTE:  Measurement error is within 5% of the distance you 
measured for all linear probes. 
Uniform MotionThe quality and usefulness of LOGIQView images is affected...

Page 150

Optimizing the Image
5-26 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Optimizing M-Mode
Intended Use
M-Mode is intended to provide a display format and 
measurement capability that represents tissue displacement 
(motion) occurring over time along a single vector. 
Introduction
M-Mode is used to determine patterns of motion for objects 
within the ultrasound beam. The most common use is for 
viewing motion patterns of the heart. 
Typical exam protocol
A typical examination using M-Mode might...