Page 141
Optimizing B-Mode LOGIQ 5 Basic User Manual 5-17 Direction 2300002-100 Rev. 2 Line Density DescriptionOptimizes B-Mode frame rate or spatial resolution for the best possible image. AdjustingPress Line Density right to increase resolution or left to increase frame rate. Va l u e sVaries by probe. NOTE: Not available in timeline. Values vary by probe and application. Line Density values are returned to the factory or user preset value when you change the following: Probe, Exam Category, Exam...
Page 142
Optimizing the Image 5-18 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Map DescriptionThe system supplies B, M, and Doppler Mode system maps. AdjustingTo select a map, press Gray Map. A map window displays adjacent to the image. Move the Trackball to select the map. The image reflects the map as you scroll through the selections. Press Set to select. Va l u e sGray maps gradually change from least contrasty (Map A) to most contrasty (Map R). Map values vary by probe, application, and...
Page 143
Optimizing B-Mode LOGIQ 5 Basic User Manual 5-19 Direction 2300002-100 Rev. 2 M Assigns a lesser amount of gray scale pixel values to weak reflections in the image than Map C. Useful when imaging abdomen and liver. N S-shaped map. This map highlights tissue differentiation for a certain band of signals. Useful when imaging structures where you want more contrast, e.g., renal, tendon, vein, carotid, thyroid, breast, Cardiology, etc. O S-shaped map. This map highlights tissue differentiation for a...
Page 144
Optimizing the Image 5-20 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Frame Average DescriptionTemporal filter that averages frames together, thereby using more pixels to make up one image. This has the effect of presenting a smoother, softer image. AdjustingTo adjust frame averaging, adjust Frame Average on the Touch Panel. Va l u e s0-5, with 5=high. The current value displays on the Touch Panel. Frame Average values vary by probe, application, and multi frequency setting. Frame...
Page 145
Optimizing B-Mode LOGIQ 5 Basic User Manual 5-21 Direction 2300002-100 Rev. 2 Colorize DescriptionColorize is the colorization of a conventional B-Mode image or Doppler Spectrum to enhance the users ability to discern B, M, and Doppler Mode intensity valuations. Colorize is NOT a Doppler Mode. NOTE: You can colorize realtime or CINE images or Timeline CINE, but not VCR images. Colorizes the gray scale image to enhance the eyes discrimination capability. Spectrum Colorize colorizes the spectrum...
Page 146
Optimizing the Image 5-22 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Edge Enhance DescriptionEdge Enhance brings out subtle tissue differences and boundaries by enhancing the gray scale differences corresponding to the edges of structures. Adjustments to M- Modes edge enhancement affects the M-Mode only. AdjustingTo cycle through settings, adjust Edge Enhance on the Touch Panel. Va l u e s0-5, with 0 = Off and 5 = High. The current value displays on the Touch Panel. Values vary by...
Page 147
Optimizing B-Mode LOGIQ 5 Basic User Manual 5-23 Direction 2300002-100 Rev. 2 Rejection DescriptionSelects a level below which echoes will not be amplified (an echo must have a certain minimum amplitude before it will be processed). AdjustingAdjust to the right to increase and left to decrease. Va l u e s0-5 BenefitsAllows for the elimination from the display of low level echoes caused by noise.
Page 148
Optimizing the Image 5-24 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 LOGIQView DescriptionLOGIQView provides the ability to construct and view a static 2D image which is wider than the field of view of a given transducer. This feature allows viewing and measurements of anatomy that is larger than what would fit in a single image. Examples include scanning of vascular structures and connective tissues in the arms and legs. LOGIQView constructs the extended image from individual image...
Page 149
Optimizing B-Mode LOGIQ 5 Basic User Manual 5-25 Direction 2300002-100 Rev. 2 LOGIQView (continued) 4. To restart the scan, press ‘L’ again. 5. To complete the scan, press ‘R’ or Freeze (or allow the scan to auto complete. The LOGIQView is then displayed, scaled to fit entirely on the screen. 6. Perform measurements and record images. NOTE: Measurement error is within 5% of the distance you measured for all linear probes. Uniform MotionThe quality and usefulness of LOGIQView images is affected...
Page 150
Optimizing the Image 5-26 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Optimizing M-Mode Intended Use M-Mode is intended to provide a display format and measurement capability that represents tissue displacement (motion) occurring over time along a single vector. Introduction M-Mode is used to determine patterns of motion for objects within the ultrasound beam. The most common use is for viewing motion patterns of the heart. Typical exam protocol A typical examination using M-Mode might...