GE Logiq P5 User Manual
Here you can view all the pages of manual GE Logiq P5 User Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.
Page 161
Optimizing Color Flow LOGIQ 5 Basic User Manual 5-37 Direction 2300002-100 Rev. 2 Activating Power Doppler Imaging (PDI) To activate PDI, 1. Press PDI. The color flow window appears over the B-Mode image. Move the Trackball to move the CF window. To exit, press PDI or select a new mode. Uses Power Doppler Imaging is a color flow mapping technique used to map the strength of the Doppler signal coming from the flow rather than the frequency shift of the signal. Using this technique, the ultrasound...
Page 162
Optimizing the Image 5-38 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Color Flow and Power Doppler Displays Figure 5-10. Power Doppler Imaging Display ItemDescription 1 Color Flow Region of Interest (ROI) 2 Color Flow Parameters: Frequency, Gain, Power Output, PRF, Wall Filter, and Packet Size 3 Color Flow Bar ItemDescription 1 Color Flow Region of Interest (ROI) 2 Color Flow Parameters: Frequency, Gain, Power Output, PRF, Wall Filter, and Packet Size 3 Color Flow Bar
Page 163
Optimizing Color Flow LOGIQ 5 Basic User Manual 5-39 Direction 2300002-100 Rev. 2 Color Flow Scanning Hints HINTSColor Flow Mode controls produce the following results: Line Density. Trades frame rate for sensitivity and spatial resolution. If the frame rate is too slow, reduce the size of the region of interest, select a different frame rate setting, or reduce the packet size. Wall Filter. Affects low flow sensitivity versus motion artifact. Color Threshold. Percentage of gray scale level where...
Page 164
Optimizing the Image 5-40 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Color Flow Mode Touch Panel Figure 5-11. Color Flow Mode Touch Panel, Page 1 Figure 5-12. Color Flow Mode Touch Panel, Page 2
Page 165
Optimizing Color Flow LOGIQ 5 Basic User Manual 5-41 Direction 2300002-100 Rev. 2 PDI (Power Doppler Imaging) Mode Touch Panel Figure 5-13. PDI Mode Touch Panel, Page 1 Figure 5-14. PDI Mode Touch Panel, Page 2
Page 166
Optimizing the Image 5-42 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Gain DescriptionGain amplifies the overall strength of echoes processed in the Color Flow window or spectral Doppler timeline. AdjustingGain values change depending on the probe and application; they are not associated with a particular position of the button. To decrease/increase Gain, adjust the dial to the left/right. Va l u e sValues vary by probe, application, and multi frequency setting. Gain displays as dB. Gain...
Page 167
Optimizing Color Flow LOGIQ 5 Basic User Manual 5-43 Direction 2300002-100 Rev. 2 Wall Filter DescriptionFilters out low flow velocity signals in order to hide unusable motion. It helps get rid of motion artifacts caused from breathing and other patient motion. AdjustingTo raise/lower the wall filter, press PRF/Wall Filter until you reach Wall Filter, then adjust Wall Filter left/right. Va l u e sValues vary, depending upon probe, application, and packet size. The wall filter is shown graphically...
Page 168
Optimizing the Image 5-44 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Invert (Color Invert) DescriptionLets you view blood flow from a different perspective, e.g., red away (negative velocities) and blue toward (positive velocities). You can invert a real-time or frozen image. NOTE: Invert reverses the color map, NOT the color PRF. AdjustingTo reverse the color flow, press Invert (Color Invert) . In Triplex, both Color Flow and Doppler Mode velocity scales are inverted. Va l u e...
Page 169
Optimizing Color Flow LOGIQ 5 Basic User Manual 5-45 Direction 2300002-100 Rev. 2 Angle Steer and Fine Angle Steer DescriptionYou can slant the B-Mode or Color Flow linear image left or right to get more information without moving the probe. AdjustingTo slant the linear image to the left/right, adjust Angle Steer. Press Angle Steer to access Fine Angle Steer. Va l u e sLinear probes can be steered left, center, or right up to a maximum of 20 degrees, depending on the probe. Angle Steer values vary...
Page 170
Optimizing the Image 5-46 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Map DescriptionBrings out the tissue behind the color map. AdjustingPress Duplex to adjust. Va l u e sOn/Off. Variance Maps (VV). Provides a measure of turbulence (stenosis). Adds green to velocity maps. Symmetric Velocity Imaging Map. Map V3 shows both positive and negative velocities using the same colors. BenefitsShows the direction of the flow and highlights the higher velocity flows.