Using 3D
LOGIQ 5 Basic User Manual 5-77
Direction 2300002-100 Rev. 2Moving Through
the 3D VOIYou can move through the 3D VOI using the red hand. Press 
Set when the red pointer finger is positioned on the red box. 
Move the closed red hand to move through the 3D VOI. 
 Figure 5-24. Moving through a 3D Volume of Interest (Red Hand)
NOTE:  Any plane in the volume can be made active (highlighted with 
red box) by clicking on it.  

Page 202

Optimizing the Image
5-78 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2 Viewing Specific
Portions of the
AnatomyYou can pull back tissue to view specific portions of anatomy 
using the yellow hand. Press Set when the yellow pointer finger 
is positioned on the yellow box. Move the closed yellow hand to 
manipulate the 3D VOI. 
NOTE:  This actually moves an edge. A yellow hand appears only when 
the pointer is on an edge of the VOI. 
 Figure 5-25. Manipulating the Edge of a 3D Volume of Interest...

Page 203

Using 3D
LOGIQ 5 Basic User Manual 5-79
Direction 2300002-100 Rev. 2Pulling Back a
Corner of the VOI
to View Specific
AnatomyYou can pull back a corner to view specific portions of anatomy 
using the green hand. Press Set when the green pointer finger is 
positioned on the green box. Move the closed green hand to 
manipulate the 3D VOI.
 Figure 5-26. Manipulating a Corner of the 3D Volume of Interest (Green Hand) 

Page 204

Optimizing the Image
5-80 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
3D Acquisition Parameter Description
 Figure 5-27. 3D Acquisition 
Table 5-5: 3D Acquisition Description and Instructions for Use
3D ParameterDescription
Application 
PresetsSelections: None, OB - Baby Face, Vascular
None. No application preset applied. 
OB - Baby Face. After having scanned in this mode, certain rendering parameters 
are set automatically. The gray surface mode is activated and the texture mode is 
switched...

Page 205

Using 3D
LOGIQ 5 Basic User Manual 5-81
Direction 2300002-100 Rev. 2
Acquisition Mode Selections: Sensorless Parallel, Sensorless Sweep, With Sensor
Sensorless Parallel. In this mode the probe must be moved during 3D data 
acquisition without angling it. You should scan the object you want to render in 2-4 
seconds. The speed at which you scan should be constant. No sensor is mounted on 
the probe. 
• Since the time for post-processing depends on the acquired number of frames, it is 
recommended that you...

Page 206

Optimizing the Image
5-82 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Easy 3D
 Figure 5-28. Easy 3D 
Descriptions and instructions for using Easy 3D follow: 
Table 5-6: Easy 3D Description and Instructions for Use
3D ParameterDescription
Reset Resets the 3D volume of interest back to its original orientation. 
Undo Undoes any manipulation you have done to your 3D dataset. 
Auto Movie Initializes the calculation and display of a 3D movie. A rotation of 30 degrees left and 
right around the...

Page 207

Using 3D
LOGIQ 5 Basic User Manual 5-83
Direction 2300002-100 Rev. 2
Texture Activates the texture or photorealistic rendering mode. It creates a photorealistic 
appearance of the object. The shading depends on the orientation of the surface of 
the object. 
If both Texture and Gray Surface mode are switched on, the mixture percentage of 
both modes can be defined. 
Render Changes between the rendered image view and the view of a volume of interest. The 
volume of interest shows the acquired ultrasound...

Page 208

Optimizing the Image
5-84 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Advanced 3D
 Figure 5-29. Advanced 3D 
Descriptions and instructions for using Advanced 3D follow: 
Table 5-7: Advanced 3D Description and Instructions for Use
3D ParameterDescription
Tile The display can be divided into 1, 2, 4, or 6 windows. Switching to a lower number of 
windows keeps the images from left to right. 
3D Landscape Shows a combination of 2D slices and a 3D rendered image. After a color acquisition 
you can...

Page 209

Using 3D
LOGIQ 5 Basic User Manual 5-85
Direction 2300002-100 Rev. 2
Movie 3D
 Figure 5-30. Movie 3D 
Descriptions and instructions for using Movie 3D follow
Table 5-8: Movie 3D Descriptions and Instructions for Use
3D ParameterDescription
Manual An animated rotation of the rendered image can be calculated and displayed by this 
function. Using this function, you first need to define the start and end position of the 
rotation. To define this, move the VOI to the start position, the press Define Start....

Page 210

Optimizing the Image
5-86 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2 