    							ADS XiLog+ System  User Manual 
     55 
     
    7.  Feed the battery cable f rom the new battery through the hole in the 
    battery holder. 
    8.  Place and properly position the battery into the holder. 
    9.  Insert the battery holder back  into the XiLog+, and secure it with the 
    two screws.  
    10.  Connect the battery cable to the circuit board . 
    11.  Slide  the circuit board into the rails inside the housing . 
    12.  Place the yellow cover back onto the XiLog+ and into position. 
    13.  Turn the yellow cover clockwise 1/8 of a turn.   
    						
    							ADS XiLog+ System User Manual 
     56 
      
    						
    							ADS XiLog+ System  User Manual 
     57 
    C H A P T E R  5  
    9 -Channel Connections  
    Two options are available for wiring the outputs on the XiLog+ 9 : 
    •  NXG- 302 Expansion Box  
    •  NXG- 303 Bare Wire Cable 
    NXG-302 Expansion Box  
    ADS offers an expansion box for the wiring.  Make connections based on the 
    figures below.  Seal any unused connectors on the box with a blank to prevent 
    damage to the connection box. 
      
    						
    							ADS XiLog+ System User Manual 
     58 
     
    Note:   Each voltage input has a 2- , 5-, or 10- volt range,  based on the 
    position of the jumper next to the appropriate input.  Verify the correct 
    position.  Selecting the wrong position could r esult in damage to the 
    logger. 
    NXG-303 Bare Wire Cable  
      
    						
    							ADS XiLog+ System  User Manual 
     59 
     
    R  represents the following:  
    V= 0 -2 volts; R=0Ω  
    V= 0 -5 volts; R=69k8Ω , ±0.1%, 25ppm 
    V= 0 -10 volts; R=187k Ω, ±0.1%, 25ppm   
    						
    							ADS XiLog+ System User Manual 
     60 
     
      
    						
    							ADS XiLog+ System  User Manual 
     61 
    C H A P T E R  6  
    Appendices 
    T his chapter contain s import ant instructional and referenc e information 
    regarding the XiLog+ FastTrack  software, meter scaling, battery life, and 
    airfreight restrictions.   
    						
    							ADS XiLog+ System User Manual 
     62 
     Appendix  A:  XiLog+  FastTrack Software 
    The XiLog+ FastTrack  software primarily  is designed for expediting the 
    process of programming the logger , viewing the current logger configuration, 
    and testing the signal strength.   Although the PrimeWorks configuration 
    process is much more comprehensive, FastTrack  represents an efficient and 
    effective alternative for deploy ing and/or testing the logger quic kl y in the 
    field.  
    Note:   The XiLog+ FastTrack  software does not support XiLog+ 
    models 2 or 9. 
    To perform any or all of these tasks, first launch the FastTrack  software 
    (provided it has been previously installed on the computer) by clicking on the 
    icon on t he Windows  desktop or selecting All Programs > Primayer > 
    XiLog+ FastTrack > XiLog+ FastTrack  from the Windows Start menu.  
    Then, connect the XiLog+ to the computer using the yellow communication 
    cable (ADS part number NXG -881- 001).  The FastTrack  software will 
    display the current serial number and phone number  (when applicable).  It also 
    will display other settings based on the last time the software was run.  
     
    XiLog+ FastTrack software   
    						
    							ADS XiLog+ System  User Manual 
     63 
    Viewing the Current Configuration 
    To view the current configuration programmed into the logger, click on the 
    View  button at the bottom of the XiLog+ FastTrack  application window.  
    Programming the Logger 
    1. Edit the name of the logger in the Site Name  field, when necessary. 
    2.  Edit the phone number (when applicable) in the Phone N umber field. 
    3.  Designate the parameters under the Transmission  header in the 
    following way:  
      Enter the time at which the logger should transmit the data to the 
    host system in the Transmit Time  field.  The logger will send the 
    data at this time every day.  
      Select the checkbox corresponding to Auto GSM Clock 
    Correction  to synchronize  the logger clock  to the time in  the GSM 
    network every 30 days. 
      Select the radio button ( SMS Transmission or FTP Transmission ) 
    corresponding to the method of GSM communication that wi ll be in 
    use.  For SMS , enter the number for the GSM modem connected to 
    the host system to which the data will be sent in the Host Number  
    field.  For FTP, enter the IP address for the server to which the data 
    will be sent in the Server Address  field.  Enter the user name and 
    the password necessary to access the server in the associated fields.  
    Note:  When using the Primayer XMS  System for FTP data delivery, 
    enter www.waterlogger.co.uk for the server address.  Whe n using 
    another server, enter the IP address without the initial ftp:// . 
    4. Designate the parameters under the Flow Channel 1  header in the 
    following way:  
      Select the type of sensor connected to Channel 1 from the Sensor 
    Profile  drop-down list. 
      Enter the amount of flow that must occur to generate a pulse in the 
    FCAL  field.  Select the unit of measure for the value from the 
    associated drop -down list.  
      Enter the unit of measure in which the flow data will be reported 
    from the Units  drop-down list. 
      Enter the current reading from the meter  in the Meter Reading 
    field, and select the appro priate unit of measure from the associated 
    drop- down list.  
    						
    							ADS XiLog+ System User Manual 
     64 
     Select the interval at  which the data will be logged and averaged  (as 
    necessary) from the Interval  drop-down list.  Sending  5- or 10-
    minute data will require sending more than one SMS per channel 
    per day.   Choosing Off  will prevent the channel from recording 
    measurements.  
    5.  Designate the parameters under the Pressure Channel 1  header in the 
    following wa y: 
      Select the unit of meas ure in which to log the data from the Units  
    drop- down list. 
      Select the frequency at which to record the data on this channel 
    from the Interval  drop-down list.   Choose Off  to prevent the 
    channel from recor ding data. 
    6.  Initiate logger programming based on the  designated parameters in one 
    of the following ways:  
      Program Logger, Delete Data   To delete all old data from the 
    logger during programming, click on the Start  button. 
      Program Logger, Retain Data   To program the logger without 
    deleting the data, click on t he down arrow to the right of the Start  
    button and select the Changes Only  option. 
    XiLog+ FastTrack programs the logger based on the selected method.  
    Pressing start also synchronizes the logger clock to the PC clock. 
      
    						