Home > ADS > Monitoring system > ADS XiLog User Manual August 2011 QR 775026 A0

ADS XiLog User Manual August 2011 QR 775026 A0

    Download as PDF Print this page Share this page

    Have a look at the manual ADS XiLog User Manual August 2011 QR 775026 A0 online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 9 ADS manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							ADS XiLog+ System  User Manual 
     35 
     
    Generic Analog –  MilliV olt example 
    Enable Channel  
    Selecting this checkbox enables  the channel on the logger.  Not a ll channels 
    on the  logger must  be enabled when the logger is in use.  Di sabled c hannels do 
    not use power. 
    Sensor Type  
    This drop-down includes the options for selecting the sensor type ( Current, 
    M illiVolt , or Voltage ).  Selecting the Voltage  or Current also requires 
    selection  of the Electrical Range .   Refer to Electrical Range  on page 36. 
    Sensor Rating  
    (applicable only to the MilliVolt sensor type )  This drop-down list includes the 
    options for  selecting the range of the pressure transducer.  The Physical 
    Range  represents this rating  converted to the selected units. 
    Units 
    This drop- down list contains the options for selecting the units of measure 
    corresponding to the designated data type.  The options available depend on 
    the selected data type.  
    Data Type 
    This drop-down list contains the options for selecting the type of data to log 
    (e.g., pressure, depth, flow, or temperature) .  
    						
    							ADS XiLog+ System User Manual 
     36 
    PCAL  
    (applicable only to the MilliVolt sensor type )  This pressure calibration value 
    represents  the number of millivolts at full range.  This value is unique for each 
    sensor, and sometimes is referred to as Full Scale Detection  (FSD). 
    Custom Offset 
    (applicable only when measuring pressure, height, or depth)  This field 
    indi cates the offset added to all readings.   For example, enter the appropriate 
    value to add to all pressure values to compensate for ground level.  To remove 
    an amount from every reading, simply enter the value as negative.  
    Electrical Range 
    (applicable only t o voltage or current data types )  These fields represent the  
    electrical range corresponding to the voltage or current, as applicable.   Enter 
    or use the up/down arrows to designate  a range up to 20mA for current.  
    Select 0 -2V, 0- 5V, and 0- 10V for voltage .  
    Physical Range  
    (applicable only to the  current or voltage sensor type )  These fields represent 
    the range of the sensor input  corresponding to the voltage .  For example, 
    consider an input of 4- 20mA involving a temperature sensor where 4 mA 
    represents 0 degr ees and 20 mA represents 100 degrees.  This would require 
    entering the values 0 and 100 in  the Physical Range  fields.  Enter the values 
    manually or use the up/down arrows. 
    Logging Interval  
    This drop- down list includes the options for se tting  the measurement interval 
    at which  to log data on this channel.  Options available range from 1 second to 
    24 hours.  The  data logged on th is channel represent s averaged data ( based on 
    the number of samples taken).  XiLog+ records the averaged data and daily 
    minimum and  maximum values for the channel.  
    						
    							ADS XiLog+ System  User Manual 
     37 
    Send Data Reports  
    Selecting this option ensures the logger sends data reports by SMS or GPRS  
    based on the frequency selected for remote data reporting.   
    T he minimum logging interval for data sent through SMS is 1 minute.  
    Ac tivating SMS reporting for data based on a shorter logging interval  requires 
    averaging the  data at a 1-minute interval before being sent . 
    Note:  Sending data logged at an interval of less than 15 minutes 
    through SMS  requires sending more than one SMS per day for each 
    channel.  For example, sending 1- minute SMS data every day requires 
    sending 15 SMS per channel per day. 
    Secondary Logging Interval 
    Some XiLog+ models allow a secondary logging interval for logging the same 
    channel input at a second rate.  This data can be reported remotely, or logged 
    internally and remotely requested (as required).   
    						
    							ADS XiLog+ System User Manual 
     38 
    Digital Channel  
     
    This window includes the parameters for setting up digital channels.  The 
    presence of a green  dot to the left of the channel name indicates the ch annel is 
    enabled and recording data.  
    Enable 
    Selecting this checkbox enables  the channel on the logger.   Not all channels 
    on a logger must  be enabled when the logger i s in use.  Disabled channels do 
    not use power. 
    Sensor  Profile/Sensor  Type 
    The Sensor Profile  drop-down list contains the options for selecting t he kind 
    of digital sensor that  will be attached to the logger.   If the appropriate sensor is 
    not available in the list,  select the Custom option and then select the 
    corresponding sensor type  from the  Sensor Type  list.  Each  digital channel 
    can receive two pulse inputs . 
     
    Sensor Type options   
    						
    							ADS XiLog+ System  User Manual 
     39 
    Several options also are included in the Sensor Types  drop-down list for 
    performing event logging using digital sensors.  To  implement event logging, 
    select one of the corresponding options from the list.  
    Pulse Direction 
    This drop-down list contai ns the options for specifying how a bidirectional 
    sensor  indicates reverse flow.  
    Logging Interval 
    This drop-down list includes the options for designating the measurement 
    interv al at which  to log data on this channel.   The o ptions available range 
    from 5  seconds to 24 hours.  The  data logged on the channel represents 
    averaged data (based o n the number of samples taken).  XiLog+ records the 
    ave rage d data and daily minimum and maxi mum values for the channel.  The 
    digital channel als o records daily volume.  Transmitting the meter reading 
    from the site to the logger, as be low, will ensure remote tracking of the site 
    meter reading  is accurate.  
    Units 
    This drop- down list contains the options for selecting the units of measure in 
    which to log  the data. 
    Continuous Power  
    Selecting this checkbox ensures the channel provides  uninterrupted power to a 
    pulse (sensor) unit that requires continuous power .  Use this option only when 
    the required by  the sensor type. 
    Combine Flows  
    Selecting this checkbox adds together the flow totals from  a combination 
    meter, which represents two meters (larger and smaller) working in parallel, 
    before the data is transmitted to the logger  to provide one flow rate and meter 
    reading .  In this scenario, flows exceeding  a certain rate will flow through the 
    larger  meter and flows occurring below  a certain rate will flow through the 
    smaller meter.   The outputs from both meters connect to the logger on two 
    different pulse tra ins of a single channel.  
    						
    							ADS XiLog+ System User Manual 
     40 
    If this option is not selected, the flow rates from the two meters will be logged 
    and transmitted independently.  Therefore, transmitting the data via SMS will 
    result in sending  double the number of SMS. 
    Send Data Reports  
    Selecting this checkbox ensures the logger send s data reports by SMS or 
    GPRS , based on the interval  designated for remote data reporting.  
    The minimum logging interval for data sent through SMS reporting is 1 
    minute.   For data logged at a shorter interval, the logg er will average the data 
    at a 1 -minute interval before sending the report . 
    Note:  Sending data logged at an interval of less than 15 minutes via 
    SMS requires sending more than one SMS per day for each  per 
    channel.  For example, sending 1 minute SMS data every da y requires 
    sending 15 SMS per channel per day. 
    Calibration Factor  
    This field represents the number of liters that correspond to a single pulse 
    from  the meter .  This value should always reflect the factor set in the meter.  
    Mo dify this factor manually  or using the up/down arrows. 
    Meter Reading 
    This field displays the site meter reading  entered into the logger.  Writing this 
    reading  into the  logger upon installation  will ensure the meter reading in the 
    XiLog+ track s the actual meter reading.   Enter a meter reading up to 10 digits 
    in length , with an  accuracy of up to 4 decimal places.  
    Meter Units 
    This drop-down list includes the options for selecting the units of measure 
    corresponding to t he meter read ing.  A variety of units  are available, based on 
    the s ize of the meter and accuracy required.  
    Secondary Logging 
    Some XiLog+ models allow a secondary logging interval f or logging the same 
    channel input at a second rate.  This data can be reported remotely, or logged  
    						
    							ADS XiLog+ System  User Manual 
     41 
    internally and remotely requested (as requir ed).  Select the Enabled checkbox 
    to enable the associated fields, and then enter the second interval at which to 
    log the data and designate whether to transmit the data obtained at this interval 
    during data reporting activities.    
    						
    							ADS XiLog+ System User Manual 
     42 
    Alarm Configuration  
     
    Al arm Configuration  sub-window  
    Alarms may be setup  independently for each channel of the logger.   Three 
    types of alarms  are available:  threshold, profile , and envelope alarms. 
    Enabled 
    Selecting this checkbox activates the alarms for this channel.  
    Profile Mode 
     
    Profile Mode  menu options  
    This drop-down contains the options for selecting the kind of alarm to 
    generate from the XiLog+  under certain threshold conditions  for a channel.  
    The following sections provide information on the individual alarm modes 
    and de scriptions of  the corresponding parameters that must be configured for 
    the alarms.  The Common Alarm Configuration Parameters  section includes 
    the descriptions for those parameters that are applicable to all three alarm 
    modes. 
    Common Alarm Configuration Pa rameters 
    This section describes the parameters that typically are common to all three 
    alarm modes.  
    						
    							ADS XiLog+ System  User Manual 
     43 
       Deadband  (applicable only to Threshold and Profile alarms)  
    This field displays the deadband value around the threshold.  The logger will 
    generate a high alarm when the channel reading exceeds  the High Threshold 
    plus the deadband.  The alarm clears when the channel reading falls below the 
    High Threshold  and the deadband.  For example, entering a high threshold of 
    0.456 mgd and a deadband of 0.023 mgd will trig ger a high alarm when the 
    flow exceeds 0.479 mgd and clear when the flow falls below 0.434 mgd.  This 
    prevents the system from generating multiple alarms at locations that 
    experience flows that may oscillate around a threshold value.  
    If this feature is no t relevant or required, enter a deadband of 0. 
    Note:  While the deadband value always represents an amount  when 
    establishing a T hreshold alarm,  the deadband can represent either a n 
    amount or a percentage when using  Profile alarms.  To indicate that 
    the val ue in the Deadband field serves as a percentage, select the 
    Deadband is percentage  checkbox after entering the value.  
     Debounce Delay 
    This drop-down list contains the options for selecting the debounce delay, 
    which represents the amount of time the flow mus t remain above the 
    designated high threshold (and deadband) or fall below the designated low 
    threshold (and deadband) before the system generates an alarm.  Delay options 
    range from 1 to 60 minutes.  For example, when selecting a debounce delay of 
    5 minute s, the flow must remain above the high threshold (plus the deadband) 
    for at least 5 minutes before the system will generate a high alarm.  
    Note:  The debounce delay applies only when initiating alarms.   It is 
    not designed for use in determining whether an alarm has been cleared.  
     Alarm Reporting 
     
    Reporting method options  
    Alarm reporting can occur through SMS or FTP (or both), and each alarm 
    trigger can be setup to occur through a different reporting method.  Delivery 
    methods for alarms are independent from t he delivery selections for collecting 
    logged data.  For example, logged data can be sent via FTP and alarms can be 
    sent via SMS.   
    						
    							ADS XiLog+ System User Manual 
     44 
    Threshold Alarms  
    Threshold alarms generally involve designating a high and/or low threshold 
    beyond or below which the XiLog+ wi ll generate an alarm.  This section 
    includes descriptions for the parameters that are common only to threshold 
    alarms.  
      Low Threshold 
    This field represents the threshold below which the logger will trigger a low  
    alarm.  Enter the threshold manually or use t he up/down arrows. 
      High Threshold 
    This field represents the threshold above which the logger will trigger a high 
    alarm.  Enter the threshold manually or use the up/down arrows. 
    Profile  and Envelope Alarms  
    Profile and Env elope alarms  are more complex alarms  that involve  
    designating a different threshold value for the high and low alarms for every 
    15- minute increment throughout the day .  For example, the h igh thresholds 
    designated  for generating flow alarm s may be much lower through the night 
    than during the da y.  
    Selecting  either option displays a graph  for creati ng and editing the profile or 
    envelope. 
     
    Importing sample data from the database 
    The simplest way to  establish a profile or envelope alarm is to import 
    historical data from the site  in the following way: 
    1.  Select the appropriate logging function from the A vailable Data list.  
    2.  Select the radio button corresponding to the option for the number of 
    days for which to import data. 
      Single day   Choose this option t o import data for only one day, 
    and then designa te the date for which to import the data using the 
    Start Date  calendar control.   
    						
    All ADS manuals Comments (0)

    Related Manuals for ADS XiLog User Manual August 2011 QR 775026 A0