ADS XiLog User Manual August 2011 QR 775026 A0
Have a look at the manual ADS XiLog User Manual August 2011 QR 775026 A0 online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 9 ADS manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
ADS XiLog+ System User Manual 35 Generic Analog – MilliV olt example Enable Channel Selecting this checkbox enables the channel on the logger. Not a ll channels on the logger must be enabled when the logger is in use. Di sabled c hannels do not use power. Sensor Type This drop-down includes the options for selecting the sensor type ( Current, M illiVolt , or Voltage ). Selecting the Voltage or Current also requires selection of the Electrical Range . Refer to Electrical Range on page 36. Sensor Rating (applicable only to the MilliVolt sensor type ) This drop-down list includes the options for selecting the range of the pressure transducer. The Physical Range represents this rating converted to the selected units. Units This drop- down list contains the options for selecting the units of measure corresponding to the designated data type. The options available depend on the selected data type. Data Type This drop-down list contains the options for selecting the type of data to log (e.g., pressure, depth, flow, or temperature) .
ADS XiLog+ System User Manual 36 PCAL (applicable only to the MilliVolt sensor type ) This pressure calibration value represents the number of millivolts at full range. This value is unique for each sensor, and sometimes is referred to as Full Scale Detection (FSD). Custom Offset (applicable only when measuring pressure, height, or depth) This field indi cates the offset added to all readings. For example, enter the appropriate value to add to all pressure values to compensate for ground level. To remove an amount from every reading, simply enter the value as negative. Electrical Range (applicable only t o voltage or current data types ) These fields represent the electrical range corresponding to the voltage or current, as applicable. Enter or use the up/down arrows to designate a range up to 20mA for current. Select 0 -2V, 0- 5V, and 0- 10V for voltage . Physical Range (applicable only to the current or voltage sensor type ) These fields represent the range of the sensor input corresponding to the voltage . For example, consider an input of 4- 20mA involving a temperature sensor where 4 mA represents 0 degr ees and 20 mA represents 100 degrees. This would require entering the values 0 and 100 in the Physical Range fields. Enter the values manually or use the up/down arrows. Logging Interval This drop- down list includes the options for se tting the measurement interval at which to log data on this channel. Options available range from 1 second to 24 hours. The data logged on th is channel represent s averaged data ( based on the number of samples taken). XiLog+ records the averaged data and daily minimum and maximum values for the channel.
ADS XiLog+ System User Manual 37 Send Data Reports Selecting this option ensures the logger sends data reports by SMS or GPRS based on the frequency selected for remote data reporting. T he minimum logging interval for data sent through SMS is 1 minute. Ac tivating SMS reporting for data based on a shorter logging interval requires averaging the data at a 1-minute interval before being sent . Note: Sending data logged at an interval of less than 15 minutes through SMS requires sending more than one SMS per day for each channel. For example, sending 1- minute SMS data every day requires sending 15 SMS per channel per day. Secondary Logging Interval Some XiLog+ models allow a secondary logging interval for logging the same channel input at a second rate. This data can be reported remotely, or logged internally and remotely requested (as required).
ADS XiLog+ System User Manual 38 Digital Channel This window includes the parameters for setting up digital channels. The presence of a green dot to the left of the channel name indicates the ch annel is enabled and recording data. Enable Selecting this checkbox enables the channel on the logger. Not all channels on a logger must be enabled when the logger i s in use. Disabled channels do not use power. Sensor Profile/Sensor Type The Sensor Profile drop-down list contains the options for selecting t he kind of digital sensor that will be attached to the logger. If the appropriate sensor is not available in the list, select the Custom option and then select the corresponding sensor type from the Sensor Type list. Each digital channel can receive two pulse inputs . Sensor Type options
ADS XiLog+ System User Manual 39 Several options also are included in the Sensor Types drop-down list for performing event logging using digital sensors. To implement event logging, select one of the corresponding options from the list. Pulse Direction This drop-down list contai ns the options for specifying how a bidirectional sensor indicates reverse flow. Logging Interval This drop-down list includes the options for designating the measurement interv al at which to log data on this channel. The o ptions available range from 5 seconds to 24 hours. The data logged on the channel represents averaged data (based o n the number of samples taken). XiLog+ records the ave rage d data and daily minimum and maxi mum values for the channel. The digital channel als o records daily volume. Transmitting the meter reading from the site to the logger, as be low, will ensure remote tracking of the site meter reading is accurate. Units This drop- down list contains the options for selecting the units of measure in which to log the data. Continuous Power Selecting this checkbox ensures the channel provides uninterrupted power to a pulse (sensor) unit that requires continuous power . Use this option only when the required by the sensor type. Combine Flows Selecting this checkbox adds together the flow totals from a combination meter, which represents two meters (larger and smaller) working in parallel, before the data is transmitted to the logger to provide one flow rate and meter reading . In this scenario, flows exceeding a certain rate will flow through the larger meter and flows occurring below a certain rate will flow through the smaller meter. The outputs from both meters connect to the logger on two different pulse tra ins of a single channel.
ADS XiLog+ System User Manual 40 If this option is not selected, the flow rates from the two meters will be logged and transmitted independently. Therefore, transmitting the data via SMS will result in sending double the number of SMS. Send Data Reports Selecting this checkbox ensures the logger send s data reports by SMS or GPRS , based on the interval designated for remote data reporting. The minimum logging interval for data sent through SMS reporting is 1 minute. For data logged at a shorter interval, the logg er will average the data at a 1 -minute interval before sending the report . Note: Sending data logged at an interval of less than 15 minutes via SMS requires sending more than one SMS per day for each per channel. For example, sending 1 minute SMS data every da y requires sending 15 SMS per channel per day. Calibration Factor This field represents the number of liters that correspond to a single pulse from the meter . This value should always reflect the factor set in the meter. Mo dify this factor manually or using the up/down arrows. Meter Reading This field displays the site meter reading entered into the logger. Writing this reading into the logger upon installation will ensure the meter reading in the XiLog+ track s the actual meter reading. Enter a meter reading up to 10 digits in length , with an accuracy of up to 4 decimal places. Meter Units This drop-down list includes the options for selecting the units of measure corresponding to t he meter read ing. A variety of units are available, based on the s ize of the meter and accuracy required. Secondary Logging Some XiLog+ models allow a secondary logging interval f or logging the same channel input at a second rate. This data can be reported remotely, or logged
ADS XiLog+ System User Manual 41 internally and remotely requested (as requir ed). Select the Enabled checkbox to enable the associated fields, and then enter the second interval at which to log the data and designate whether to transmit the data obtained at this interval during data reporting activities.
ADS XiLog+ System User Manual 42 Alarm Configuration Al arm Configuration sub-window Alarms may be setup independently for each channel of the logger. Three types of alarms are available: threshold, profile , and envelope alarms. Enabled Selecting this checkbox activates the alarms for this channel. Profile Mode Profile Mode menu options This drop-down contains the options for selecting the kind of alarm to generate from the XiLog+ under certain threshold conditions for a channel. The following sections provide information on the individual alarm modes and de scriptions of the corresponding parameters that must be configured for the alarms. The Common Alarm Configuration Parameters section includes the descriptions for those parameters that are applicable to all three alarm modes. Common Alarm Configuration Pa rameters This section describes the parameters that typically are common to all three alarm modes.
ADS XiLog+ System User Manual 43 Deadband (applicable only to Threshold and Profile alarms) This field displays the deadband value around the threshold. The logger will generate a high alarm when the channel reading exceeds the High Threshold plus the deadband. The alarm clears when the channel reading falls below the High Threshold and the deadband. For example, entering a high threshold of 0.456 mgd and a deadband of 0.023 mgd will trig ger a high alarm when the flow exceeds 0.479 mgd and clear when the flow falls below 0.434 mgd. This prevents the system from generating multiple alarms at locations that experience flows that may oscillate around a threshold value. If this feature is no t relevant or required, enter a deadband of 0. Note: While the deadband value always represents an amount when establishing a T hreshold alarm, the deadband can represent either a n amount or a percentage when using Profile alarms. To indicate that the val ue in the Deadband field serves as a percentage, select the Deadband is percentage checkbox after entering the value. Debounce Delay This drop-down list contains the options for selecting the debounce delay, which represents the amount of time the flow mus t remain above the designated high threshold (and deadband) or fall below the designated low threshold (and deadband) before the system generates an alarm. Delay options range from 1 to 60 minutes. For example, when selecting a debounce delay of 5 minute s, the flow must remain above the high threshold (plus the deadband) for at least 5 minutes before the system will generate a high alarm. Note: The debounce delay applies only when initiating alarms. It is not designed for use in determining whether an alarm has been cleared. Alarm Reporting Reporting method options Alarm reporting can occur through SMS or FTP (or both), and each alarm trigger can be setup to occur through a different reporting method. Delivery methods for alarms are independent from t he delivery selections for collecting logged data. For example, logged data can be sent via FTP and alarms can be sent via SMS.
ADS XiLog+ System User Manual 44 Threshold Alarms Threshold alarms generally involve designating a high and/or low threshold beyond or below which the XiLog+ wi ll generate an alarm. This section includes descriptions for the parameters that are common only to threshold alarms. Low Threshold This field represents the threshold below which the logger will trigger a low alarm. Enter the threshold manually or use t he up/down arrows. High Threshold This field represents the threshold above which the logger will trigger a high alarm. Enter the threshold manually or use the up/down arrows. Profile and Envelope Alarms Profile and Env elope alarms are more complex alarms that involve designating a different threshold value for the high and low alarms for every 15- minute increment throughout the day . For example, the h igh thresholds designated for generating flow alarm s may be much lower through the night than during the da y. Selecting either option displays a graph for creati ng and editing the profile or envelope. Importing sample data from the database The simplest way to establish a profile or envelope alarm is to import historical data from the site in the following way: 1. Select the appropriate logging function from the A vailable Data list. 2. Select the radio button corresponding to the option for the number of days for which to import data. Single day Choose this option t o import data for only one day, and then designa te the date for which to import the data using the Start Date calendar control.