    							2.8.21 E 21, Program-Dependent Activation of 10th Wash Block NS1 (Final Rinse 1)
    For programming mode access, see Programming Mode Summary, 070 4.1.
    x
    Program Level E 21
    Program Selector
    Switch SettingDisplay Function (Depending on Cleaning Program)
    1 o'clock10Flashing display, wash block 10 NS1 (final rinse 1) not
    active
    10
    –Constant display, dash at top: wash block 10 NS1 (final
    rinse 1) activated with KW (cold water)
    10–Constant display, dash in middle: wash block 10 NS1 (final
    rinse 1) activated with WW (hot water)
    10
    –Constant display, dash at bottom: wash block 10 NS1
    (final rinse 1) activated with AD (demineralized water)
    2 o'clockAs described for the 1 o'clock setting, but only applicable for the cleaning
    program that can be started at the 2 o'clock setting
    3 o'clockAs described for the 1 o'clock setting, but only applicable for the cleaning
    program that can be started at the 3 o'clock setting
    etc. . . . . . .
    Table 30:Program LevelE 21
    2.8.22 E 22, Program-Dependent Temperature Setting T5 in Wash Block NS1 (Final
    Rinse 1)
    For programming mode access, see Programming Mode Summary, 070 4.1.
    x
    Program Level E 22
    Program Selector
    Switch SettingDisplay Function (Depending on Cleaning Program)
    1 o'clock e.g., 55Temperature setting in 1°C steps between 30°C and
    93°C (some programs 70°C)
    2 o'clockAs described for the 1 o'clock setting, but only applicable for the cleaning
    program that can be started at the 2 o'clock setting
    3 o'clockAs described for the 1 o'clock setting, but only applicable for the cleaning
    program that can be started at the 3 o'clock setting
    etc. . . . . . .
    Table 31:Program LevelE 22
    2.8.23 E 23, Program-Dependent Setting of Temperature Holding Time t5 in Wash
    Block NS1 (Final Rinse 1)
    For programming mode access, see Programming Mode Summary, 070 4.1.
    x
    Program Level E 23
    Program Selector
    Switch SettingDisplay Function (Depending on Cleaning Program)
    Descriptive Technical Documentation
    070-36 DTD no. 62-7881
    This information should not be duplicated or passed on without Miele approval. All rights reserved. 18.09.2013, US_am 
    						
    							Program Level E 23
    Program Selector
    Switch SettingDisplay Function (Depending on Cleaning Program)
    1 o'clock e.g., 10Setting the holding time t5 in temperature step T5 in 1-min
    steps, between 1 min and 15 min
    2 o'clockAs described for the 1 o'clock setting, but only applicable for the cleaning
    program that can be started at the 2 o'clock setting
    3 o'clockAs described for the 1 o'clock setting, but only applicable for the cleaning
    program that can be started at the 3 o'clock setting
    etc. . . . . . .
    Table 32:Program LevelE 23
    2.8.24 E 24, Program-Dependent Activation of 11th Wash Block NS2 (Final Rinse 2)
    For programming mode access, see Programming Mode Summary, 070 4.1.
    x
    Program Level E 24
    Program Selector
    Switch SettingDisplay Function (Depending on Cleaning Program)
    1 o'clock11Flashing display, wash block 11 NS2 (final rinse 2) not
    active
    11
    –Constant display, dash at top: wash block 11 NS2 (final
    rinse 2) activated with KW (cold water)
    11–Constant display, dash in middle: wash block 11 NS2 (final
    rinse 2) activated with WW (hot water)
    11
    –Constant display, dash at bottom: wash block 11 NS2 (final
    rinse 2) activated with AD (demineralized water)
    2 o'clockAs described for the 1 o'clock setting, but only applicable for the cleaning
    program that can be started at the 2 o'clock setting
    3 o'clockAs described for the 1 o'clock setting, but only applicable for the cleaning
    program that can be started at the 3 o'clock setting
    etc. . . . . . .
    Table 33:Program LevelE 24
    2.8.25 E 25, Program-Dependent Temperature Setting T2 in Wash Block NS2 (Final
    Rinse 2)
    For programming mode access, see Programming Mode Summary, 070 4.1.
    x
    Program Level E 25
    Program Selector
    Switch SettingDisplay Function (Depending on Cleaning Program)
    1 o'clock e.g., 55Temperature setting in 1°C steps between 30°C and
    93°C (some programs 70°C)
    2 o'clockAs described for the 1 o'clock setting, but only applicable for the cleaning
    program that can be started at the 2 o'clock setting
    Descriptive Technical Documentation
    DTD no. 62-7881 070-37
    18.09.2013, US_am This information should not be duplicated or passed on without Miele approval. All rights reserved. 
    						
    							Program Level E 25
    Program Selector
    Switch SettingDisplay Function (Depending on Cleaning Program)
    3 o'clockAs described for the 1 o'clock setting, but only applicable for the cleaning
    program that can be started at the 3 o'clock setting
    etc. . . . . . .
    Table 34:Program LevelE 25
    2.8.26 E 26, Program-Dependent Setting of Temperature Holding Time t2 in Wash
    Block NS2 (Final Rinse 2)
    For programming mode access, see Programming Mode Summary, 070 4.1.
    x
    Program Level E 26
    Program Selector
    Switch SettingDisplay Function (Depending on Cleaning Program)
    1 o'clock e.g., 10Setting the exposure time t2 in temperature step T2 in 1-
    min steps, between 1 min and 15 min
    2 o'clockAs described for the 1 o'clock setting, but only applicable for the cleaning
    program that can be started at the 2 o'clock setting
    3 o'clockAs described for the 1 o'clock setting, but only applicable for the cleaning
    program that can be started at the 3 o'clock setting
    etc. . . . . . .
    Table 35:Program LevelE 26
    2.8.27 E 27, Program-Dependent Activation of 12th Wash Block (Drying) TA1
    G 7882 CD, G 7883 CD
    For programming mode access, see Programming Mode Summary, 070 4.1.
    x
    Program Level E 27
    Program Selector
    Switch SettingDisplay Function (Depending on Cleaning Program)
    1 o'clock12Flashing display, wash block 12 TA1 (drying unit) not active
    12Constant display: wash block 12 TA1 (drying unit only)
    activated
    2 o'clockAs described for the 1 o'clock setting, but only applicable for the cleaning
    program that can be started at the 2 o'clock setting
    3 o'clockAs described for the 1 o'clock setting, but only applicable for the cleaning
    program that can be started at the 3 o'clock setting
    etc. . . . . . .
    Table 36:Program LevelE 27
    2.8.28 E 28, Program-Dependent Temperature Setting for Drying TA1
    G 7882 CD, G 7883 CD
    Descriptive Technical Documentation
    070-38 DTD no. 62-7881
    This information should not be duplicated or passed on without Miele approval. All rights reserved. 18.09.2013, US_am 
    						
    							For programming mode access, see Programming Mode Summary, 070 4.1.
    x
    Program Level E 28
    Program Selector
    Switch SettingDisplay Function (Depending on Cleaning Program)
    1 o'clock e.g., 55Temperature setting in 1°C steps between 30°C and
    99°C
    2 o'clockAs described for the 1 o'clock setting, but only applicable for the cleaning
    program that can be started at the 2 o'clock setting
    3 o'clockAs described for the 1 o'clock setting, but only applicable for the cleaning
    program that can be started at the 3 o'clock setting
    etc. . . . . . .
    Table 37:Program LevelE 28
    2.8.29 E 29, Program-Dependent Setting of Temperature Holding Time tA1 in
    Drying Block TA1
    G 7882 CD, G 7883 CD
    For programming mode access, see Programming Mode Summary, 070 4.1.
    x
    Program Level E 29
    Program Selector
    Switch SettingDisplay Function (Depending on Cleaning Program)
    1 o'clock e.g., 10Setting the holding time tA1 in temperature step TA1 in 1-
    min steps, between 1 min and 99 min
    2 o'clockAs described for the 1 o'clock setting, but only applicable for the cleaning
    program that can be started at the 2 o'clock setting
    3 o'clockAs described for the 1 o'clock setting, but only applicable for the cleaning
    program that can be started at the 3 o'clock setting
    etc. . . . . . .
    Table 38:Program LevelE 29
    2.8.30 E 30, Program-Dependent Activation of 13th Wash Block (Drying) TR / TA2
    For programming mode access, see Programming Mode Summary, 070 4.1.
    x
    Program Level E 30
    Program Selector
    Switch SettingDisplay Function (Depending on Cleaning Program)
    1 o'clock13Flashing display, wash block 13 TR / TA2 (drying) not
    active
    13Constant display: wash block 13 TR / TA2 (drying)
    activated (drying block TA2 only with TA (drying unit) and
    TR only without TA)
    2 o'clockAs described for the 1 o'clock setting, but only applicable for the cleaning
    program that can be started at the 2 o'clock setting
    Descriptive Technical Documentation
    DTD no. 62-7881 070-39
    18.09.2013, US_am This information should not be duplicated or passed on without Miele approval. All rights reserved. 
    						
    							Program Level E 30
    Program Selector
    Switch SettingDisplay Function (Depending on Cleaning Program)
    3 o'clockAs described for the 1 o'clock setting, but only applicable for the cleaning
    program that can be started at the 3 o'clock setting
    etc. . . . . . .
    Table 39:Program LevelE 30
    2.8.31 E 31, Program-Dependent Temperature Setting for Drying TA2
    G 7882 CD, G 7883 CD
    For programming mode access, see Programming Mode Summary, 070 4.1.
    x
    Program Level E 31
    Program Selector
    Switch SettingDisplay Function (Depending on Cleaning Program)
    1 o'clock e.g., 55Temperature setting in 1°C steps between 30°C and
    99°C
    2 o'clockAs described for the 1 o'clock setting, but only applicable for the cleaning
    program that can be started at the 2 o'clock setting
    3 o'clockAs described for the 1 o'clock setting, but only applicable for the cleaning
    program that can be started at the 3 o'clock setting
    etc. . . . . . .
    Table 40:Program LevelE 31
    2.8.32 E 32, Program-Dependent Setting of Temperature Holding Time tA2 in
    Drying Block TA2
    G 7882 CD, G 7883 CD
    For programming mode access, see Programming Mode Summary, 070 4.1.
    x
    Program Level E 32
    Program Selector
    Switch SettingDisplay Function (Depending on Cleaning Program)
    1 o'clock e.g., 10Setting the holding time tA2 in temperature step TA2 in 1-
    min steps, between 1 min and 95 min
    2 o'clockAs described for the 1 o'clock setting, but only applicable for the cleaning
    program that can be started at the 2 o'clock setting
    3 o'clockAs described for the 1 o'clock setting, but only applicable for the cleaning
    program that can be started at the 3 o'clock setting
    etc. . . . . . .
    Table 41:Program LevelE 32
    Descriptive Technical Documentation
    070-40 DTD no. 62-7881
    This information should not be duplicated or passed on without Miele approval. All rights reserved. 18.09.2013, US_am 
    						
    							2.8.33 E 33, Resetting Electronic
    For programming mode access, see Programming Mode Summary, 070 4.1.
    x
    Program Level E 33
    Program Selector
    Switch SettingDisplay Function (Depending on Cleaning Program)
    Immaterial00Unprogrammed electronic - Cannot be modified
    PPWith programmed electronic - Can be reset to00or back to
    PP
    Table 42:Program LevelE 33
    2.8.34 E 34, Clock Setting (From P 06)
    For programming mode access, see Programming Mode Summary, 070 4.1.
    x
    Program Level E 34
    Program Selector Switch
    SettingDisplay
    1 o'clock e.g., 04 Day setting (01 - 31)
    2 o'clock
    e.g., 09 Month setting (01 - 12)
    3 o'clock e.g., 01 Year setting (00 - 99) = 2000 - 2099
    4 o'clock e.g., 22 Hour setting (00 - 23)
    5 o'clock
    e.g., 43 Minute setting (00 - 55)
    Table 43:Program LevelE 34
    2.9 Service Mode 2 - Service Levels S 21 to S 27
    2.9.1 S 21, Checking Operating Hours / Filter Operating Hours and Printer
    Interface Activation (RS-232)
    For service mode 2 access, see Service Mode 2 Summary, 070 4.3.
    x
    Service Level S 21
    Program Selector
    Switch SettingDisplay Function
    1 o'clockThe last 3 fault codes (e.g.,
    F00)If service mode 2 is quit with saving, then all
    entries in the fault memory are deleted, see
    Service Mode 2 Summary, 070 4.3
    2 o'clockOperating hours
    1)in 1000 h
    (e.g., 001)—
    3 o'clockOperating hours
    1)in h (e.g.,
    234)—
    Descriptive Technical Documentation
    DTD no. 62-7881 070-41
    18.09.2013, US_am This information should not be duplicated or passed on without Miele approval. All rights reserved. 
    						
    							Service Level S 21
    Program Selector
    Switch SettingDisplay Function
    4 o'clock
    4)
    40Serial printer interface not available
    41Serial printer interface activated
    42Serial printer interface activated and test
    printout. In order to receive a printout
    immediately after entry, press theStart
    button twice (save).
    5 o'clockCoarse filter operating
    hours
    2)in h (e.g., 250)—
    6 o'clockFine filter operating hours
    2)in
    h (e.g., 500)—
    7 o'clockCoarse filter operating hours
    with reset
    3)(e.g., 010)—
    8 o'clockFine filter operating hours
    with reset
    3)(e.g., 210)—
    Table 44:Service LevelS 21
    1)Only periods of cleaning program operation are registered and added together.2)The filter life figures given by the factory can only be displayed or modified if a drying unit is
    installed (CD models).
    3)The display of current filter life figures and resetting of figures to those given by the factory is
    only possible if a drying unit is installed (CD models).
    4)Interface data can only be set if an activated interface exists.
    2.9.2 S 22, Setting Printer Interface (RS-232) and Dispenser Pumps
    For service mode 2 access, see Service Mode 2 Summary, 070 4.3.
    x
    Service Level S 22
    Program Selector
    Switch SettingDisplay Function
    1 o'clock10Selected language - German
    11Selected language - English
    12Selected language - Italian
    13Selected language - French
    14Selected language - Spanish
    2 o'clock20Paper length 2 inches - New label after 2 inches
    21Paper length 4 inches - New label after 4 inches
    22Paper length 6 inches - New label after 6 inches
    23DIN A4 page printout
    24Endless printout with 4 blank lines at end of report
    3 o'clock e.g., 04Day setting (01 - 31)
    4 o'clock e.g., 02Month setting (01 - 12)
    5 o'clock e.g., 01Year setting (01 - 99)
    Descriptive Technical Documentation
    070-42 DTD no. 62-7881
    This information should not be duplicated or passed on without Miele approval. All rights reserved. 18.09.2013, US_am 
    						
    							Service Level S 22
    Program Selector
    Switch SettingDisplay Function
    6 o'clock e.g., 11Hour setting (01 - 23)
    7 o'clock e.g., 45Minute setting (01 - 59)
    8 o'clock e.g., 108Capacity - DOS dispenser pump 1, range 10 - 150 mL/min
    9 o'clock—DOS dispenser pump 2 is rinse aid door dispenser, no
    capacity setting
    10 o'clock e.g., 18Capacity - DOS dispenser pump 3, range 10 - 150 mL/min
    11 o'clock e.g., 110Capacity - DOS dispenser pump 4, range 10 - 150 mL/min
    Table 45:Service LevelS 22
    2.9.3 S 23, Checking Outputs 1 to 11
    For service mode 2 access, see Service Mode 2 Summary, 070 4.3.
    x
    Note
    If themin°Cbutton is pressed and held during testing of outputs 1 to 13, then if
    thecbutton is pressed, the circulation pump is also activated.
    Service Level S 23
    Program
    Selector Switch
    SettingDisplay Function
    1 o'clock 1Output 1: Circulation pump
    2 o'clock 2Output 2: Drain pump
    3 o'clock 3Output 3: Water inlet valve (WPS), cold water
    4 o'clock 4Output 4: Water inlet valve (WPS), hot water
    5 o'clock 5Output 5: Water inlet valve (WPS), demineralized water or
    non-pressurized demineralized water (G 7883 and G 7883
    CD only)
    6 o'clock 6Output 6: Reactivation valve
    7 o'clock 7Output 7: DOS 1 dispenser, detergent
    8 o'clock 8Output 8: DOS 2 dispenser, rinse aid
    9 o'clock 9Output 9: Door opener
    1)
    10 o'clock 10Output 10: Heating, water and drying2)
    11 o'clock 11Output 11: Heating, water
    Table 46:Service LevelS 23
    1)If the door opener is checked and the door opens, service mode 2 is then quit because there is
    a break in the power supply (door contact switch).
    2)Drying with pulsed power via output 10 is only for washer-disinfectors without a drying unit.
    Descriptive Technical Documentation
    DTD no. 62-7881 070-43
    18.09.2013, US_am This information should not be duplicated or passed on without Miele approval. All rights reserved. 
    						
    							2.9.4 S 24, Checking Outputs 12 - 19
    For service mode 2 access, see Service Mode 2 Summary, 070 4.3.
    x
    Service Level S 24
    Program
    Selector Switch
    SettingDisplay Function
    1 o'clock—Vacant
    2 o'clock 12
    1)Output 12: DOS 4 dispenser, disinfectant
    3 o'clock 13
    1)Output 13: DOS 3 dispenser, neutralizing agent
    4 o'clock 14
    2)Output 14: Drying unit base load
    5 o'clock 15
    2)Output 15: Drying unit control
    6 o'clock 16
    2)Output 16: Drying unit fan
    7 o'clock 17
    2)Output 17: DK3), no function
    Output 17: DHK
    4)1, spray valve
    8 o'clock 18Output 18: DK
    3)drain
    Output 18: DHK
    4)drain
    9 o'clock 19
    2)Output 19: DK3), no function
    Output 19: DHK
    4)2, spray valve
    Table 47:Service LevelS 24
    1)If themin°Cbutton is pressed and held during testing of outputs 12 and 13, then if thec
    button is pressed, the circulation pump is also activated.
    2)Outputs 14 to 17 and 19 are only used in washer-disinfectors with drying unit (CD versions).3)DK: Steam condenser functioning with heat exchanger principle.4)DHK: Steam and hot-air condenser operating with spray valves.
    2.9.5 S 25, Machine Number Setting
    For service mode 2 access, see Service Mode 2 Summary, 070 4.3.
    x
    Service Level S 25
    Program
    Selector Switch
    SettingDisplay Function
    1 o'clock e.g., 1 Last digit of machine number (0–9)
    2 o'clock e.g., 2 Next-to-last digit of machine number (0–9)
    3 o'clock ... ...
    4 o'clock .......
    etc. ... ...
    7 o'clock e.g., 7 Second digit of machine number (0–9)
    8 o'clock e.g., 8 First digit of machine number (0–9)
    9 o'clock—Vacant
    10 o'clock—Vacant
    Descriptive Technical Documentation
    070-44 DTD no. 62-7881
    This information should not be duplicated or passed on without Miele approval. All rights reserved. 18.09.2013, US_am 
    						
    							Service Level S 25
    Program
    Selector Switch
    SettingDisplay Function
    11 o'clock00 Machine number deleted
    PP Machine number programmed
    Table 48:Service LevelS 25
    Note
    The set machine number is recorded in a printout via the serial interface.
    2.9.6 S 26, Special Functions
    For service mode 2 access, see Service Mode 2 Summary, 070 4.3.
    From P 05: Program interruption possible.
    From P 05: Display for machines without steam condenser (DHK), display only for
    machines with pulsed heating.
    Dispensing relay monitoring
    x
    Service Level S 26
    Program
    Selector Switch
    SettingDisplay Function
    1 o'clock PE
    1)
    Constant display: Programming mode
    accessible
    Flashing display: Programming mode blocked
    Flashing display: Service mode 2 access more
    difficult
    2 o'clock20Program cancellation is allowed at any time
    during a cleaning programmed by selecting
    “STOP”
    21No program cancellation during a cleaning
    program by selecting“STOP”
    3 o'clock30Without additional steam condenser (DHK)
    pause
    31With additional steam condenser (DHK) pause
    (if drying not active)
    32With additional steam condenser (DHK) pause
    (if heating element drying active)
    4 o'clock40 Drying using drying unit (TA)
    41Drying using heater element pulsing (G 7884
    only)
    Descriptive Technical Documentation