Preparing for a Biopsy
LOGIQ 5 Basic User Manual 17-39
Direction 2300002-100 Rev. 2
Biopsy Need Path Verification
To verify that the path of the needle is accurately indicated within 
the guidezone on the system monitor, perform the following:
•  Properly install the bracket and biopsy guide.
•  Scan in a container filled with water.
•  Display the biopsy guidezone on the monitor.
•  Ensure that the needle echo falls within the guidezone 
markers. 

Page 822

Probes and Biopsy
17-40 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
The Biopsy Procedure
1.  Place coupling gel on the scanning surface of the probe/
sheath/biopsy guide assembly.
2.  Activate the biopsy guidezone on the system through the 
B-Mode Touch Panel.  When using multi-angle guides, 
ensure that the proper guidezone angle is displayed.
 Figure 17-23. B-Mode Touch Panel
3.  Scan to locate the target.  Center the target in the electronic 
guidezone path.
NOTE:  Enabling color flow would...

Page 823

Preparing for a Biopsy
LOGIQ 5 Basic User Manual 17-41
Direction 2300002-100 Rev. 2
Post Biopsy
When the biopsy is complete, remove the needle barrel, needle 
clip and probe sheath.  Properly dispose of these items in 
accordance with current facility guidelines.
The biopsy bracket can be sterilized in a recommended 
disinfecting agent and reused.
E8C Probe Biopsy Guide
When the E8C probe is attached and active, the needle guide 
type is the TR5° (Civco disposable guide with a 5° offset angle).
 Figure...

Page 824

Probes and Biopsy
17-42 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Preparation (continued)
To install the sheath:
1.  Remove the sheath from its package. Do not unroll the 
sheath.
NOTE:  Remember to rinse all sanitary probe sheaths of powder 
before placing on the probe. Powder can degrade the 
displayed image.
2.  Place an adequate amount of ultrasound gel inside the 
sheath tip (the gel is between the sheath inner surface and 
the probe aperture).
NOTE:  Ensure that only acoustic coupling...

Page 825

Preparing for a Biopsy
LOGIQ 5 Basic User Manual 17-43
Direction 2300002-100 Rev. 2
Preparation (continued)
7.  Place an adequate amount of ultrasound gel on the gel-filled 
sheath tip’s outer surface.
 Figure 17-27. Civco Disposable Biopsy Guide 5° Angle
8.  Ensure the guide is properly seated and secure by pushing 
forward on the needle insertion end of the guide until the 
attachment node is firmly in place in it’s hole. 

Page 826

Probes and Biopsy
17-44 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2 

Page 827

LOGIQ 5 Basic User Manual 18-1
Direction 2300002-100 Rev. 2
Chapter 18
User Maintenance
This chapter supplies system data, assistance 
information, and system care and maintenance 
instructions.  

Page 828

User Maintenance
18-2 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
System Data
LOGIQ 5 Features/Specifications
Table 18-1: Physical Attributes
Physical Attributes
Dimensions and Weight
• Height: Minimum 1350 mm (53 in). Maximum 1450 
mm (56 in)
• Width: 520 mm (20 in)
• Depth: 990 mm (39 in)
• Weight: approx. 180 kg (397 lb.)
Keyboard
• Keyboard width: 520 mm
• Keyboard Height: 900 mm, adjustable
• Full alphanumeric keyboard
• Ergonomic hard key operations
• Interactive Back-Lighting
• Task...

Page 829

System Data
LOGIQ 5 Basic User Manual 18-3
Direction 2300002-100 Rev. 2
LOGIQ 5 Features/Specifications (continued)
Table 18-2: System Overview
Applications• Abdominal
• Obstetrical
• Gynecological
• Cardiac
• Musculoskeletal
• Vascular
• Urological
• Small parts and Superficial
• Breast
• Pediatric and Neonatal
• Orthopedic
• Intraoperative
• Transcranial
Scanning Methods
• Electronic Sector
• Electronic Convex
• Electronic Linear
• Steered Linear
Operating Modes
• B-Mode
• M-Mode
• M Color Flow Mode
•...

Page 830

User Maintenance
18-4 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
LOGIQ 5 Features/Specifications (continued)
Table 18-3: System Parameters
Post-Processing• B/M-Mode (Grey Map; Rejection; Colorized B and 
M; Frame Averaging)
• D-Mode (Grey Map; Rejection; Colorized D)
• Color Flow (Color Map; CFM Display Threshold)
Pre-Processing
• B/M-Mode (Gain; TGC; Dynamic Range; 
Transmission Focus Position; Transmission Focus 
Number; Transmission Frequency; Edge 
Enhancement; Frame Rate Control; Image...
Start reading GE Logiq P5 User Manual

Related Manuals for GE Logiq P5 User Manual

All GE manuals