Preparing for a Biopsy LOGIQ 5 Basic User Manual 17-39 Direction 2300002-100 Rev. 2 Biopsy Need Path Verification To verify that the path of the needle is accurately indicated within the guidezone on the system monitor, perform the following: • Properly install the bracket and biopsy guide. • Scan in a container filled with water. • Display the biopsy guidezone on the monitor. • Ensure that the needle echo falls within the guidezone markers.
Page 822
Probes and Biopsy 17-40 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 The Biopsy Procedure 1. Place coupling gel on the scanning surface of the probe/ sheath/biopsy guide assembly. 2. Activate the biopsy guidezone on the system through the B-Mode Touch Panel. When using multi-angle guides, ensure that the proper guidezone angle is displayed. Figure 17-23. B-Mode Touch Panel 3. Scan to locate the target. Center the target in the electronic guidezone path. NOTE: Enabling color flow would...
Page 823
Preparing for a Biopsy LOGIQ 5 Basic User Manual 17-41 Direction 2300002-100 Rev. 2 Post Biopsy When the biopsy is complete, remove the needle barrel, needle clip and probe sheath. Properly dispose of these items in accordance with current facility guidelines. The biopsy bracket can be sterilized in a recommended disinfecting agent and reused. E8C Probe Biopsy Guide When the E8C probe is attached and active, the needle guide type is the TR5° (Civco disposable guide with a 5° offset angle). Figure...
Page 824
Probes and Biopsy 17-42 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Preparation (continued) To install the sheath: 1. Remove the sheath from its package. Do not unroll the sheath. NOTE: Remember to rinse all sanitary probe sheaths of powder before placing on the probe. Powder can degrade the displayed image. 2. Place an adequate amount of ultrasound gel inside the sheath tip (the gel is between the sheath inner surface and the probe aperture). NOTE: Ensure that only acoustic coupling...
Page 825
Preparing for a Biopsy LOGIQ 5 Basic User Manual 17-43 Direction 2300002-100 Rev. 2 Preparation (continued) 7. Place an adequate amount of ultrasound gel on the gel-filled sheath tip’s outer surface. Figure 17-27. Civco Disposable Biopsy Guide 5° Angle 8. Ensure the guide is properly seated and secure by pushing forward on the needle insertion end of the guide until the attachment node is firmly in place in it’s hole.
Page 827
LOGIQ 5 Basic User Manual 18-1 Direction 2300002-100 Rev. 2 Chapter 18 User Maintenance This chapter supplies system data, assistance information, and system care and maintenance instructions.
Page 828
User Maintenance 18-2 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 System Data LOGIQ 5 Features/Specifications Table 18-1: Physical Attributes Physical Attributes Dimensions and Weight • Height: Minimum 1350 mm (53 in). Maximum 1450 mm (56 in) • Width: 520 mm (20 in) • Depth: 990 mm (39 in) • Weight: approx. 180 kg (397 lb.) Keyboard • Keyboard width: 520 mm • Keyboard Height: 900 mm, adjustable • Full alphanumeric keyboard • Ergonomic hard key operations • Interactive Back-Lighting • Task...
Page 829
System Data LOGIQ 5 Basic User Manual 18-3 Direction 2300002-100 Rev. 2 LOGIQ 5 Features/Specifications (continued) Table 18-2: System Overview Applications• Abdominal • Obstetrical • Gynecological • Cardiac • Musculoskeletal • Vascular • Urological • Small parts and Superficial • Breast • Pediatric and Neonatal • Orthopedic • Intraoperative • Transcranial Scanning Methods • Electronic Sector • Electronic Convex • Electronic Linear • Steered Linear Operating Modes • B-Mode • M-Mode • M Color Flow Mode •...
Page 830
User Maintenance 18-4 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 LOGIQ 5 Features/Specifications (continued) Table 18-3: System Parameters Post-Processing• B/M-Mode (Grey Map; Rejection; Colorized B and M; Frame Averaging) • D-Mode (Grey Map; Rejection; Colorized D) • Color Flow (Color Map; CFM Display Threshold) Pre-Processing • B/M-Mode (Gain; TGC; Dynamic Range; Transmission Focus Position; Transmission Focus Number; Transmission Frequency; Edge Enhancement; Frame Rate Control; Image...