Home > GE > Ultrasound System > GE Logiq P5 User Manual

GE Logiq P5 User Manual

Here you can view all the pages of manual GE Logiq P5 User Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.

Page 831

System Data
LOGIQ 5 Basic User Manual 18-5
Direction 2300002-100 Rev. 2
LOGIQ 5 Features/Specifications (continued)
Table 18-4: Measurements and Calculations
B-Mode• Distance
• Circumference/ Area
• Angle
• Ratios 
• Depth from Probe Surface
M-Mode
• Distance
• Time 
• Slope
• Heart Rate
Doppler Measurements/Calculations
• Velocity
• Frequency
• Time
• Acceleration
• Heart Rate
• Auto Doppler Trace function with automatic 
calculations
• Time Averaged Max./ Mean Velocity
• Ratios
• PI (Pulsatility...

Page 832

User Maintenance
18-6 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
LOGIQ 5 Features/Specifications (continued)
Table 18-5: Probes
• 3C Convex Probe (Applications: Abdomen, OB/
GYN, Urology)
• 3.5C Convex Probe (Applications: Abdomen, OB/
GYN, Urology)
• 5C Convex Probe(Applications: Abdomen, OB/
GYN, Urology)
• E8C Micro Convex Endocavity -Probe 
(Applications: Transvaginal, Transrectal, 
Obstetrics, Gynecology, Urology)
• 3S Phased Array Sector Probe (Applications: 
Cardiac, Transcranial)• 7L...

Page 833

System Data
LOGIQ 5 Basic User Manual 18-7
Direction 2300002-100 Rev. 2
Clinical Measurement Accuracy
Basic Measurements
The following information is intended to provide guidance to the 
user in determining the amount of variation or measurement 
error that should be considered when performing clinical 
measurements with this equipment.  Error can be contributed by 
equipment limitations and improper user technique.  Be sure to 
follow all measurement instructions and develop uniform 
measurement...

Page 834

User Maintenance
18-8 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Basic Measurements (continued)
Table 18-7: System Measurements and Accuracies
MeasurementUnitsUseful RangeAccuracy
Limitations
or
Conditions
Depth mm Full Screen±5% or 1 mm
Distance:
Axial mm Full Screen±5% or 1 mm
Lateral mm Full Screen±5% or 2 mm Linear Probes
Lateral mm Full Screen±5% or 4 mm Convex Probes
Lateral mm Full Screen±5% or 4 mm Sector Probes
Circumference:±5% or 1 mm
Trace mm Full Screen±10% or 1 mm
Ellipse mm...

Page 835

System Data
LOGIQ 5 Basic User Manual 18-9
Direction 2300002-100 Rev. 2
Clinical Calculation Accuracy
Estimate the overall inaccuracy of a combined measurement 
and calculation by including the stated inaccuracy from the basic 
measurement accuracy statements.
Calculation formulas and databases are provided as a tool to 
assist the user, but should not be considered an undisputed 
database, in making a clinical diagnosis.  The user is 
encouraged to research the literature and judge the equipment...

Page 836

User Maintenance
18-10 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
System Care and Maintenance
Overview
Refer to Section 10 of the LOGIQ 5 Service Manual for any 
additional maintenance guidance.
Contact the local Service Representative for parts or periodic 
maintenance inspections.
Inspecting the System
Examine the following on a monthly basis:
•  Connectors on cables for any mechanical defects.
•  Entire length of electrical and power cables for cuts or 
abrasions.
•  Equipment for loose or...

Page 837

System Care and Maintenance
LOGIQ 5 Basic User Manual 18-11
Direction 2300002-100 Rev. 2
Weekly Maintenance
The system requires weekly care and maintenance to function 
safely and properly.  Clean the following:
• Monitor
• Operator control panel
• Footswitch
•  Video Cassette Recorder (VCR)
•  Video Page Printer
Failure to perform required maintenance may result in 
unnecessary service calls. 

Page 838

User Maintenance
18-12 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Cleaning the system
Prior to cleaning any part of the system:
1.  Turn off the system power. If possible, disconnect the power 
cord. See ‘Power Off’ for more information.
System Cabinet
To clean the system cabinet:
1.  Moisten a soft, non-abrasive folded cloth with a mild, 
general purpose, non-abrasive soap and water solution.
2.  Moisten a soft, non-abrasive folded cloth.
3.  Wipe down the top, front, back, and both sides of...

Page 839

System Care and Maintenance
LOGIQ 5 Basic User Manual 18-13
Direction 2300002-100 Rev. 2
Footswitch
To clean the footswitch:
1.  Moisten a soft, non-abrasive folded cloth with a mild, 
general purpose, non-abrasive soap and water solution.
2.  Wipe the external surfaces of the unit then dry with a soft, 
clean, cloth.
VCR
To clean the VCR:
1.  Turn off the VCR power. If possible, disconnect the power 
cord.
2.  Wipe the external surfaces of the unit with a soft, clean, dry 
cloth.
NOTE:  Do not use a wet...

Page 840

User Maintenance
18-14 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Discs
To clean the discs:
•  If the disc is dirty, you can clean it with a commercially 
available dust removal spray. Be careful not to scratch the 
disc when wiping it off.
NOTE:  CD type cleaner cannot be used.
Trackball
1.  Power off the system.
2.  Place your fingers onto the notches of the Trackball retainer 
ring.
3.  Rotate the retainer ring counterclockwise until it can be 
removed from the keyboard.
 Figure 18-1....