Page 831
System Data LOGIQ 5 Basic User Manual 18-5 Direction 2300002-100 Rev. 2 LOGIQ 5 Features/Specifications (continued) Table 18-4: Measurements and Calculations B-Mode• Distance • Circumference/ Area • Angle • Ratios • Depth from Probe Surface M-Mode • Distance • Time • Slope • Heart Rate Doppler Measurements/Calculations • Velocity • Frequency • Time • Acceleration • Heart Rate • Auto Doppler Trace function with automatic calculations • Time Averaged Max./ Mean Velocity • Ratios • PI (Pulsatility...
Page 832
User Maintenance 18-6 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 LOGIQ 5 Features/Specifications (continued) Table 18-5: Probes • 3C Convex Probe (Applications: Abdomen, OB/ GYN, Urology) • 3.5C Convex Probe (Applications: Abdomen, OB/ GYN, Urology) • 5C Convex Probe(Applications: Abdomen, OB/ GYN, Urology) • E8C Micro Convex Endocavity -Probe (Applications: Transvaginal, Transrectal, Obstetrics, Gynecology, Urology) • 3S Phased Array Sector Probe (Applications: Cardiac, Transcranial)• 7L...
Page 833
System Data LOGIQ 5 Basic User Manual 18-7 Direction 2300002-100 Rev. 2 Clinical Measurement Accuracy Basic Measurements The following information is intended to provide guidance to the user in determining the amount of variation or measurement error that should be considered when performing clinical measurements with this equipment. Error can be contributed by equipment limitations and improper user technique. Be sure to follow all measurement instructions and develop uniform measurement...
Page 834
User Maintenance 18-8 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Basic Measurements (continued) Table 18-7: System Measurements and Accuracies MeasurementUnitsUseful RangeAccuracy Limitations or Conditions Depth mm Full Screen±5% or 1 mm Distance: Axial mm Full Screen±5% or 1 mm Lateral mm Full Screen±5% or 2 mm Linear Probes Lateral mm Full Screen±5% or 4 mm Convex Probes Lateral mm Full Screen±5% or 4 mm Sector Probes Circumference:±5% or 1 mm Trace mm Full Screen±10% or 1 mm Ellipse mm...
Page 835
System Data LOGIQ 5 Basic User Manual 18-9 Direction 2300002-100 Rev. 2 Clinical Calculation Accuracy Estimate the overall inaccuracy of a combined measurement and calculation by including the stated inaccuracy from the basic measurement accuracy statements. Calculation formulas and databases are provided as a tool to assist the user, but should not be considered an undisputed database, in making a clinical diagnosis. The user is encouraged to research the literature and judge the equipment...
Page 836
User Maintenance 18-10 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 System Care and Maintenance Overview Refer to Section 10 of the LOGIQ 5 Service Manual for any additional maintenance guidance. Contact the local Service Representative for parts or periodic maintenance inspections. Inspecting the System Examine the following on a monthly basis: • Connectors on cables for any mechanical defects. • Entire length of electrical and power cables for cuts or abrasions. • Equipment for loose or...
Page 837
System Care and Maintenance LOGIQ 5 Basic User Manual 18-11 Direction 2300002-100 Rev. 2 Weekly Maintenance The system requires weekly care and maintenance to function safely and properly. Clean the following: • Monitor • Operator control panel • Footswitch • Video Cassette Recorder (VCR) • Video Page Printer Failure to perform required maintenance may result in unnecessary service calls.
Page 838
User Maintenance 18-12 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Cleaning the system Prior to cleaning any part of the system: 1. Turn off the system power. If possible, disconnect the power cord. See ‘Power Off’ for more information. System Cabinet To clean the system cabinet: 1. Moisten a soft, non-abrasive folded cloth with a mild, general purpose, non-abrasive soap and water solution. 2. Moisten a soft, non-abrasive folded cloth. 3. Wipe down the top, front, back, and both sides of...
Page 839
System Care and Maintenance LOGIQ 5 Basic User Manual 18-13 Direction 2300002-100 Rev. 2 Footswitch To clean the footswitch: 1. Moisten a soft, non-abrasive folded cloth with a mild, general purpose, non-abrasive soap and water solution. 2. Wipe the external surfaces of the unit then dry with a soft, clean, cloth. VCR To clean the VCR: 1. Turn off the VCR power. If possible, disconnect the power cord. 2. Wipe the external surfaces of the unit with a soft, clean, dry cloth. NOTE: Do not use a wet...
Page 840
User Maintenance 18-14 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Discs To clean the discs: • If the disc is dirty, you can clean it with a commercially available dust removal spray. Be careful not to scratch the disc when wiping it off. NOTE: CD type cleaner cannot be used. Trackball 1. Power off the system. 2. Place your fingers onto the notches of the Trackball retainer ring. 3. Rotate the retainer ring counterclockwise until it can be removed from the keyboard. Figure 18-1....