System Care and Maintenance LOGIQ 5 Basic User Manual 18-15 Direction 2300002-100 Rev. 2 Trackball (continued) 4. Lift off the inner retainer and Trackball from the keyboard. Figure 18-2. Removing Inner Retainer and Trackball a. Inner Retainer b. Trackball 5. Wipe off any oil or dust from the Trackball using a cleaner or dry cloth. 6. Wipe off any oil or dust from the Trackball housing using a cleaner or cotton swab. (a) (b) CAUTIONWhen cleaning the Trackball housing, make sure not to spill or...
Page 842
User Maintenance 18-16 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Trackball (continued) Figure 18-3. Cleaning Trackball and Housing a. Trackball Housing b. Cotton Swab (a) (b)
Page 843
System Care and Maintenance LOGIQ 5 Basic User Manual 18-17 Direction 2300002-100 Rev. 2 Trackball (continued) 7. Wipe off any oil or dust from the two rollers using a cleaner or cotton swab. Figure 18-4. Cleaning Rollers a. Cotton Swab b. Rollers CAUTIONWhen cleaning the roller, make sure not to spill or spray any liquid into the Trackball housing (keyboard or system). Use either ethanol, isopropyl alcohol or VCR head cleaner to clean the Trackball assembly. Avoid other solvents that may damage...
Page 844
User Maintenance 18-18 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Trackball (continued) 8. Insert the Trackball into the housing. 9. Place the Trackball and inner retainer into the housing with its stopper facing down. Lift off the inner retainer and Trackball from the keyboard. Figure 18-5. Set Inner Retainer and Trackball a. Inner Retainer b. Trackball c. Stopper (a) (b) (c)
Page 845
System Care and Maintenance LOGIQ 5 Basic User Manual 18-19 Direction 2300002-100 Rev. 2 Trackball (continued) 10. Install the Trackball retainer ring onto the inner retainer, then rotate it clockwise until its notches are set in the horizontal position. Figure 18-6. Rotating the Retainer Ring a. Retainer Ring Figure 18-7. Notches set in the horizontal position a. Notches (Set Horizontally) (a) (a)
Page 846
User Maintenance 18-20 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Monitor and Touch Panel For monitor and touch screen, use a non-ammonia (just isopropyl and water) lens cleaner. These are available at most computer outlet stores. DO NOT use Windex, Screen-Clean, etc., because these contain ammonia, which will remove the anti-glare coating on the monitor. Video Page Printer To clean the external surface of the video page printer: 1. Turn off the power. If possible, disconnect the power...
Page 847
System Care and Maintenance LOGIQ 5 Basic User Manual 18-21 Direction 2300002-100 Rev. 2 Other Maintenance Replacing illuminated key caps/lamps Contact a local Service Representative when a key cap or lamp needs to be replaced. Uninterruptible Power Supply (UPS) Option LifetimeThe lifetime of a UPS is limited and varies according to circumstances. It lasts for three years under normal conditions; however, it may last one year if the UPS is used over twelve hours per day and if the room temperature is...
Page 848
User Maintenance 18-22 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Cleaning the LOGIQ 5 air filter Clean the systems air filters to ensure that a clogged filter does not cause the system to overheat and reduce system performance and reliability. It is recommended the filters be cleaned quarterly (once every three months). LocatingThe LOGIQ 5 has two air filters. The front air filter is located in the front of the system and the back air filter is located behind the system. Figure 18-8....
Page 849
System Care and Maintenance LOGIQ 5 Basic User Manual 18-23 Direction 2300002-100 Rev. 2 Cleaning the LOGIQ 5 air filter (continued) CleaningTo clean the filter: 1. Be sure to shake the filter in an area away from the system. 2. Wash the filter in a mild soapy solution, rinse and air dry or dry with a cloth. NOTE: Allow wet filter to dry thoroughly before installing. Installing1. Slide the filter back into the system. 2. Secure the filter with the Velcro tab.
Page 850
User Maintenance 18-24 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Quality Assurance Introduction A good Quality Assurance Evaluation program consists of periodic systematic actions that provide the user with adequate confidence that their diagnostic ultrasound system will produce consistently high quality images and quantitative information. Therefore, it is in the best interests of every ultrasound user to routinely monitor equipment performance. The frequency of Quality Assurance...