Page 861
Quality Assurance LOGIQ 5 Basic User Manual 18-35 Direction 2300002-100 Rev. 2 Penetration DescriptionPenetration is the ability of an imaging system to detect and display weak echoes from small objects at large depths. See Figure 18-9 for more information. Penetration can be affected by the systems: • Transmitter/receiver • Degree of probe focusing • Attenuation of the medium • Depth and shape of reflecting object • Electromagnetic interference from local surroundings. BenefitWeak reflecting...
Page 862
User Maintenance 18-36 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Functional resolution DescriptionFunctional resolution is an imaging systems ability to detect and display the size, shape, and depth of an anechoic structure, as opposed to a pin target. See Figure 18-9 for more information. The very best possible image is somewhat less important than reproducibility and stability over time. Routine tests at the same settings should produce the same results. BenefitThe data obtained will...
Page 863
Quality Assurance LOGIQ 5 Basic User Manual 18-37 Direction 2300002-100 Rev. 2 Contrast resolution DescriptionContrast resolution is the ability of an imaging system to detect and display the shape and echogenic characteristics of a structure. See Figure 18-9 for more information. Specific values measured are less important than stability over time. Routine tests at the same settings should produce the same results. BenefitA correct diagnosis is dependent upon an imaging systems ability to...
Page 864
User Maintenance 18-38 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Gray Scale photography DescriptionPoor photography will cause loss of low level echoes and the lack of contrast between large amplitude echoes. See Figure 18-9 for more information. BenefitWhen photographic controls and film processors are properly adjusted, weak echoes, as well as strong echoes, are accurately recorded on film. Procedure1. Adjust the camera according to the manufacturers instructions until the hard copy...
Page 865
Quality Assurance LOGIQ 5 Basic User Manual 18-39 Direction 2300002-100 Rev. 2 Setting up a Record Keeping System Preparation The following is needed: • Quality Assurance binder. • Hard copy or electronic file of images. • Quality Assurance Checklists. • Display the following information while testing quality assurance: • Acoustic Output • Gain • Depth • Probe • Dynamic Range • Set up new patient to be the name of the test. • Annotate the following: • Any control where its value is NOT...
Page 866
User Maintenance 18-40 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Ultrasound Quality Assurance Checklist (For Future Validation) Table 18-8: Ultrasound Quality Assurance Checklist (Part 1) Performed By Date System Serial Number Probe Type Probe Model Serial Number Phantom Model Serial Number Room Temperature Acoustic Output Gain Focal Zone Gray Map TGC Depth Monitor Setting Peripheral Settings Other Image Processing Control Settings Table 18-9: (Part 2) Te s t Baseline Va l ue Range Tested...
Page 867
Assistance LOGIQ 5 Basic User Manual 18-41 Direction 2300002-100 Rev. 2 Assistance Supplies/Accessories The following supplies/accessories have been verified to be compatible with the system: Peripherals CAUTIONDO NOT connect any probes or accessories without approval by GE. Table 18-10: Peripherals and Accessories AccessoryUnits 3.5” MO Drive Each Sony VCR Model SVO-9500MD/9500MDP Each Panasonic VCR Model AG-MD835/MD835P/ MD835EEach Sony B/W Printer Model UP-895MDW Each Mitsubishi B&W Printer Model...
Page 868
User Maintenance 18-42 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Console LOGIQ 5 Probes Table 18-11: Console Accessories AccessoryUnits Footswitch Each TV Probe Holder Each UPS (Battery Pack) Each Table 18-12: Probes and Accessories AccessoryUnits 3C Each 3.5C Each 5C Each E8C Each 8C Each 7L Each 10L Each 12L Each 3S Each P2D Each P6D Each
Page 869
Assistance LOGIQ 5 Basic User Manual 18-43 Direction 2300002-100 Rev. 2 Gel Disinfectant Table 18-13: Gel AccessoryUnits Thermasonic Gel Warmer Holds three plastic bottles (250ml or 8 oz) Aquasonic 100 Scan Gel 5 liter jug 250 ml plastic bottles (12/ case) Scan Ultrasound Gel 8 oz plastic bottles (12/ case) 1 gallon plastic jug Four 1-gallon plastic jugs Table 18-14: Gel AccessoryUnits Aquasonic 100 Scan Gel 5 liter jug 250 ml plastic bottles (12/ case) Scan Ultrasound Gel 8 oz plastic bottles (12/...
Page 870
User Maintenance 18-44 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Ultrasound Probe and Cord Sheath Sets Physio Accessories Patient Electrodes Contact the distributor, GE affiliate or sales representative for approved peripherals.Table 18-16: Probe and Cord Sheath Sets AccessoryUnits Sterile Ultrasound Probe Sheath Set 20 Per Set Sterile Ultrasound Cord Sheath Set 20 Per Set Sanitary Rectal/Vaginal Probe Cover 20 Per Set Sterile Combination Probe and Cord Cover Set 12 Per Set Sterile...