Quality Assurance
LOGIQ 5 Basic User Manual 18-35
Direction 2300002-100 Rev. 2
Penetration
DescriptionPenetration is the ability of an imaging system to detect and 
display weak echoes from small objects at large depths. See 
Figure 18-9 for more information.
Penetration can be affected by the systems: 
• Transmitter/receiver
•  Degree of probe focusing
•  Attenuation of the medium
•  Depth and shape of reflecting object
•  Electromagnetic interference from local surroundings. 
BenefitWeak reflecting...

Page 862

User Maintenance
18-36 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Functional resolution
DescriptionFunctional resolution is an imaging systems ability to detect and 
display the size, shape, and depth of an anechoic structure, as 
opposed to a pin target. See Figure 18-9 for more information.
The very best possible image is somewhat less important than 
reproducibility and stability over time. Routine tests at the same 
settings should produce the same results.
BenefitThe data obtained will...

Page 863

Quality Assurance
LOGIQ 5 Basic User Manual 18-37
Direction 2300002-100 Rev. 2
Contrast resolution
DescriptionContrast resolution is the ability of an imaging system to detect 
and display the shape and echogenic characteristics of a 
structure. See Figure 18-9 for more information.
Specific values measured are less important than stability over 
time. Routine tests at the same settings should produce the 
same results.
BenefitA correct diagnosis is dependent upon an imaging systems 
ability to...

Page 864

User Maintenance
18-38 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Gray Scale photography
DescriptionPoor photography will cause loss of low level echoes and the 
lack of contrast between large amplitude echoes. See 
Figure 18-9 for more information.
BenefitWhen photographic controls and film processors are properly 
adjusted, weak echoes, as well as strong echoes, are accurately 
recorded on film.
Procedure1.  Adjust the camera according to the manufacturers 
instructions until the hard copy...

Page 865

Quality Assurance
LOGIQ 5 Basic User Manual 18-39
Direction 2300002-100 Rev. 2
Setting up a Record Keeping System
Preparation
The following is needed: 
•  Quality Assurance binder.
•  Hard copy or electronic file of images.
•  Quality Assurance Checklists. 
•  Display the following information while testing quality 
assurance: 
• Acoustic Output
• Gain
• Depth
• Probe
• Dynamic Range
•  Set up new patient to be the name of the test.
•  Annotate the following: 
•  Any control where its value is NOT...

Page 866

User Maintenance
18-40 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Ultrasound Quality Assurance Checklist  (For Future Validation)
Table 18-8: Ultrasound Quality Assurance Checklist (Part 1)
Performed By Date
System Serial Number
Probe Type Probe Model Serial Number
Phantom Model Serial Number Room Temperature
Acoustic Output Gain Focal Zone
Gray Map TGC Depth
Monitor Setting
Peripheral Settings
Other Image Processing Control Settings
Table 18-9: (Part 2)
Te s t
Baseline
Va l ue
Range
Tested...

Page 867

Assistance
LOGIQ 5 Basic User Manual 18-41
Direction 2300002-100 Rev. 2
Assistance
Supplies/Accessories
The following supplies/accessories have been verified to be 
compatible with the system:
Peripherals
CAUTIONDO NOT connect any probes or accessories without approval 
by GE.
Table 18-10: Peripherals and Accessories
AccessoryUnits
3.5” MO Drive Each
Sony VCR Model SVO-9500MD/9500MDP Each
Panasonic VCR Model AG-MD835/MD835P/
MD835EEach
Sony B/W Printer Model UP-895MDW Each
Mitsubishi B&W Printer Model...

Page 868

User Maintenance
18-42 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Console
LOGIQ 5 Probes
Table 18-11: Console Accessories
AccessoryUnits
Footswitch Each
TV Probe Holder Each
UPS (Battery Pack) Each
Table 18-12: Probes and Accessories
AccessoryUnits
3C Each
3.5C Each
5C Each
E8C Each
8C Each
7L Each
10L Each
12L Each
3S Each
P2D Each
P6D Each 

Page 869

Assistance
LOGIQ 5 Basic User Manual 18-43
Direction 2300002-100 Rev. 2
Gel 
Disinfectant
Table 18-13: Gel
AccessoryUnits
Thermasonic Gel Warmer Holds three plastic bottles 
(250ml or 8 oz)
Aquasonic 100 Scan Gel 5 liter jug
250 ml plastic bottles (12/
case)
Scan Ultrasound Gel 8 oz plastic bottles (12/
case)
1 gallon plastic jug
Four 1-gallon plastic jugs
Table 18-14: Gel
AccessoryUnits
Aquasonic 100 Scan Gel 5 liter jug
250 ml plastic bottles (12/
case)
Scan Ultrasound Gel 8 oz plastic bottles (12/...

Page 870

User Maintenance
18-44 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Ultrasound Probe and Cord Sheath Sets
Physio Accessories
Patient Electrodes
Contact the distributor, GE affiliate or sales representative for 
approved peripherals.Table 18-16: Probe and Cord Sheath Sets
AccessoryUnits
Sterile Ultrasound Probe Sheath Set 20 Per Set
Sterile Ultrasound Cord Sheath Set 20 Per Set
Sanitary Rectal/Vaginal Probe Cover 20 Per Set
Sterile Combination Probe and Cord Cover Set 12 Per Set
Sterile...