Home > GE > Ultrasound System > GE Logiq P5 User Manual

GE Logiq P5 User Manual

Here you can view all the pages of manual GE Logiq P5 User Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.

Page 451

OB Table Editor
LOGIQ 5 Basic User Manual 9-85
Direction 2300002-100 Rev. 2
O B  Ta b l e  Te m p l a t e s
Tool Type:
• Distance: 2D Caliper
•  Circumference: 2D Ellipse, 2D Trace, 2D Caliper
Te m p l a t e  1
Table 9-4: Template 1 (based on Hadlock)
Template 1: SD Range Table
Fetal Age Table Table Format MEAS MEAN SD
Table Unit mm Week Week
Table Range 1SD
Graph Range 1SD
Measurement Result Value [cm]
GA [#w#d]
Min [#w#d]
Max [#w#d]
Fetal Growth Table Table Format AGE MEAN SD
Table Unit Week mm Week...

Page 452

OB/GYN
9-86 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Te m p l a t e  2
Te m p l a t e  3
Table 9-5: Template 2 (based on Tokyo)
Template 2: SD Range Table
Fetal Age Table Table Format MEAS MEAN SD
Table Unit mm Day Day
Table Range 1SD
Graph Range 1SD
Measurement Result Value [cm]
GA [#w#d]
SD: day(+/-)
Fetal Growth Table Table Format AGE MEAN SD
Table Unit Day mm Day
Others are same as above
Table 9-6: Template 3 (based on Osaka)
Template 3: SD Table
Fetal Age Table Table Format MEAS MEAN...

Page 453

OB Table Editor
LOGIQ 5 Basic User Manual 9-87
Direction 2300002-100 Rev. 2
Te m p l a t e  4
Table 9-7: Template 4 (based on several European tables)
Template 4: 5%-95% Table
Fetal Age Table Table Format MEAS MEAN SD
Table Unit mm WeekDay mm
Table Range 5%:95%
Graph Range 5%:95%
Measurement Result Value [cm]
GA [#w#d]
GP [%]
GP is calculated by Fetal Growth Table. If you did 
not edit Growth Table, GP is not calculated by the 
system,
Fetal Growth Table Table Format AGE MEAN SD
Table Unit WeekDay mm...

Page 454

OB/GYN
9-88 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Te m p l a t e  5
Table 9-8: Template 5 (based on several European tables)
Template 5: 5% - 95% Table
Fetal Age Table Table Format MEAS MEAN SD
Table Unit mm WeekDay mm
Table Range 1SD
Graph Range 5%:95%
Measurement Result Value [cm]
GA [#w#d]
GP [%]
GP is calculated by Fetal Growth Table. If you did not 
edit Growth Table, GP is not calculated by the system,
Fetal Growth Table Table Format AGE MEAN MAX
Table Unit WeekDay mm mm
Table...

Page 455

OB Table Editor
LOGIQ 5 Basic User Manual 9-89
Direction 2300002-100 Rev. 2
Te m p l a t e  6
Table 9-9: Template 6 (based on several European tables)
Template 6: 10% - 90% Table
Fetal Age Table Table Format MEAS MIN MEAN SD
Table Unit mm WeekDay WeekDay WeekDay
Table Range 10%-90%
Graph Range 10%-90%
Measurement Result Value [cm]
GA [#w#d]
GP [%]
GP is calculated by Fetal Growth Table. If you did not edit Growth 
Table, GP is not calculated by the system,
Fetal Growth Table Table Format AGE MIN MEAN SD...

Page 456

OB/GYN
9-90 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Te m p l a t e  7
Table 9-10: Template 7 (Based on several european tables)
Template 7: 10% - 90% Table
Fetal Age Table Table Format MEAS MEAN SD
Table Unit mm WeekDay mm
Table Range 1SD
Graph Range 10%:90%
Measurement Result Value [cm]
GA [#w#d]
GP [%]
GP is calculated by Fetal Growth Table. If you did not 
edit Growth Table, GP is not calculated by the system,
Fetal Growth Table Table Format AGE MEAN MAX
Table Unit WeekDay mm mm
Table...

Page 457

OB Table Editor
LOGIQ 5 Basic User Manual 9-91
Direction 2300002-100 Rev. 2
OB Table Edit Menu
The data you enter in the OB Table Edit Menu depends on 
whether the table type is Fetal Age or Fetal Growth.
Fetal Age Table
If you are creating or editing a Fetal Age table, the OB Table Edit 
Menu is as follows:
 Figure 9-28. OB Table Edit Menu: Fetal Age Table 

Page 458

OB/GYN
9-92 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Fetal Age Table (continued)
Complete the field1.  Input value to Min, Max and Interval of the Parameters field. 
The system automatically fills in the MEAS column.
Input value to the columns of MEAN and SD.
NOTE:  To move between the fields in the table, use the up, down, 
left, and right arrow keys.
NOTE:  You must enter a minimum of two rows of data. Any lines 
with a blank cell are not saved. 
To save the Table Data, move the Trackball...

Page 459

OB Table Editor
LOGIQ 5 Basic User Manual 9-93
Direction 2300002-100 Rev. 2
Fetal Growth Table
If you are creating or editing a Fetal Growth table, the OB Table 
Edit Menu is as follows:
 Figure 9-29. OB Table Edit Menu: Fetal Growth Table 
Complete the field1.  Input value to the required columns.
NOTE:  To move between the fields in the table, use the up, down, 
left, and right arrow keys.
NOTE:  You must enter a minimum of two rows of data. Any lines 
with a blank cell are not saved. 
2.  To save the...

Page 460

OB/GYN
9-94 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
GYN Measurements
Introduction
The Gynecology exam category includes the following three 
studies:
•  Generic. This study is common to all exam categories. See 
‘Generic Measurements’ for more information. 
•  General Gynecology. This study includes uterine, ovarian, 
ovarian follicle, and endometrium measurements.
•  OB/GYN Vessel. This study includes the following vessels: 
uterine, ovarian, umbilical, middle cerebral artery, aorta,...