Page 401
OB Measurements and Calculations LOGIQ 5 Basic User Manual 9-35 Direction 2300002-100 Rev. 2 % Stenosis In B-Mode, you can calculate % Stenosis by diameter or by area. Diameter NOTE: When you use diameter to calculate the %stenosis, always take the measurement from a cross-sectional view of the vessel. To calculate percent stenosis by diameter: 1. Select % Stenosis. 2. Select %sten(Diam). The system displays an active caliper. 3. Make a distance measurement of the inner area of the blood vessel:...
Page 402
OB/GYN 9-36 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 % Stenosis (continued) AreaTo calculate percent stenosis by area: 1. Select % Stenosis. 2. Select %sten(Area). The system displays a trace caliper. 3. Make a trace measurement of the inner area of the blood vessel: a. To position the trace caliper at the start point, move the Trackball. b. To fix the trace start point, press Set. The trace caliper changes to an active caliper. c. To trace the measurement area, move the Trackball...
Page 403
OB Measurements and Calculations LOGIQ 5 Basic User Manual 9-37 Direction 2300002-100 Rev. 2 Transverse Abdominal Diameter To measure transverse abdominal diameter, make one distance measurement: 1. Select TA D; an active caliper displays. 2. To position the active caliper at the start point, move the Trackball. 3. To fix the start point, press Set. The system fixes the first caliper and displays a second active caliper. 4. To position the second active caliper at the end point, move the...
Page 404
OB/GYN 9-38 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Thorax Transverse Diameter To measure thorax transverse diameter, make one distance measurement: 1. Select ThD; an active caliper displays. 2. To position the active caliper at the start point, move the Trackball. 3. To fix the start point, press Set. The system fixes the first caliper and displays a second active caliper. 4. To position the second active caliper at the end point, move the Trackball. A dotted line connects the...
Page 405
OB Measurements and Calculations LOGIQ 5 Basic User Manual 9-39 Direction 2300002-100 Rev. 2 Tibia Length To measure tibia length, make one distance measurement: 1. Select Tibia; an active caliper displays. 2. To position the active caliper at the start point, move the Trackball. 3. To fix the start point, press Set. The system fixes the first caliper and displays a second active caliper. 4. To position the second active caliper at the end point, move the Trackball. A dotted line connects the...
Page 406
OB/GYN 9-40 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Vo l u m e The system can make a volume calculation from any of the following measurements: • One distance • Two distances • Three distances To make a volume calculation using one or two distances: 1. Select Volume. 2. Make one or two distance measurements: a. To position the active caliper at the start point, move the Trackball. b. To fix the start point, press Set. The system fixes the first caliper and displays a second active...
Page 407
OB Measurements and Calculations LOGIQ 5 Basic User Manual 9-41 Direction 2300002-100 Rev. 2 Volume (continued) To make a volume calculation using three distances: NOTE: Three distances should be done in the dual format mode (side by side images). One measurement is usually made in the sagittal plane and two measurements in the axial plane. To use the dual format mode, press the L or R key on front panel. 1. Select Volume; an active caliper displays. 2. Make the first distance measurement: a. To...
Page 408
OB/GYN 9-42 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 M-Mode Measurements In M-Mode you can measure % stenosis, A/B ratio, and heart rate. % Stenosis In M-Mode, you measure % Stenosis by diameter. NOTE: Always take the measurement from a cross-sectional view of the vessel. 1. Select % Stenosis. The system displays an active caliper. 2. Make a distance measurement of the inner area of the blood vessel: a. To position the active caliper at the start point, move the Trackball. b. To...
Page 409
OB Measurements and Calculations LOGIQ 5 Basic User Manual 9-43 Direction 2300002-100 Rev. 2 A/B Ratio In M-Mode you can measure A/B ratio by diameter, time, or velocity. DiameterTo calculate A/B ratio by diameter: 1. Select A/B Ratio. 2. Select ratio(Diam). The system displays an active caliper. 3. Make a distance measurement of the first velocity. a. To position the active caliper at the start point, move the Trackball. b. To fix the start point, press Set. The system fixes the first caliper and...
Page 410
OB/GYN 9-44 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 A/B Ratio (continued) TimeTo calculate A/B ratio by time: 1. Select A/B. 2. Select ratio(Time). The system displays an active caliper. 3. To position the caliper at the A point, move the Trackball. 4. To fix the measure point, press Set. The system displays a second active caliper. 5. To position the second caliper at the B point, move the Trackball. 6. To complete the measurement, press Set. The system displays the two time...