Page 391
OB Measurements and Calculations LOGIQ 5 Basic User Manual 9-25 Direction 2300002-100 Rev. 2 Fetal Trunk Area (FTA) (continued) Trace1. Select FTA; a trace caliper displays. 2. To position the trace caliper at the start point, move the Trackball. 3. To fix the trace start point, press Set. The trace caliper changes to an active caliper. 4. To trace the measurement area, move the Trackball around the anatomy. A dotted line shows the traced area. 5. To complete the measurement, press Set. The system...
Page 392
OB/GYN 9-26 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Fetal Trunk Area (FTA) (continued) Two distances1. Select FTA; an active caliper displays. 2. Make the first distance measurement: a. To position the active caliper at the start point, move the Trackball. b. To fix the start point, press Set. The system fixes the first caliper and displays a second active caliper. c. To position the second active caliper at the end point, move the Trackball. A dotted line connects the measurement...
Page 393
OB Measurements and Calculations LOGIQ 5 Basic User Manual 9-27 Direction 2300002-100 Rev. 2 Foot Length To measure foot length, make one distance measurement: 1. Select Ft; an active caliper displays. 2. To position the active caliper at the start point, move the Trackball. 3. To fix the start point, press Set. The system fixes the first caliper and displays a second active caliper. 4. To position the second active caliper at the end point, move the Trackball. A dotted line connects the...
Page 394
OB/GYN 9-28 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Gestational Sac To calculate the gestational sac, you make three distance measurements in two scan planes. To display two scan planes, press the L or R key. Get an image in each scan plane and press Freeze. 1. Select GS; an active caliper displays. a. To position the active caliper at the start point, move the Trackball. b. To fix the start point, press Set. The system fixes the first caliper and displays a second active...
Page 395
OB Measurements and Calculations LOGIQ 5 Basic User Manual 9-29 Direction 2300002-100 Rev. 2 Head Circumference (HC) To calculate head circumference, you make an ellipse, trace, or two distance measurements. Ellipse1. Select HC; an active caliper displays. 2. If necessary, select HC until Ellipse is displayed above the key. 3. To position the active caliper, move the Trackball. 4. To fix the start point, press Set. The system fixes the first caliper and displays a second active caliper. 5. To...
Page 396
OB/GYN 9-30 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Head Circumference (HC) (continued) Trace1. Select HC; an active caliper displays. 2. If necessary, select HC until Trace is displayed above the key. 3. To position the trace caliper, move the Trackball. 4. To fix the trace start point, press Set. The trace caliper changes to an active caliper. 5. To trace the measurement area, move the Trackball around the anatomy. 6. To complete the measurement, press Set. The system displays the...
Page 397
OB Measurements and Calculations LOGIQ 5 Basic User Manual 9-31 Direction 2300002-100 Rev. 2 Head Circumference (HC) (continued) Two distances1. Select HC; an active caliper displays. 2. If necessary, select HC until Trace is displayed above the key. 3. To make the first distance measurement: a. To position the active caliper at the start point, move the Trackball. b. To fix the start point, press Set. The system fixes the first caliper and displays a second active caliper. c. To position the...
Page 398
OB/GYN 9-32 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Heart Rate To calculate the heart rate: 1. Select Heart Rate. The system displays an active caliper. 2. To position the caliper at a recognizable point in the first cycle, move the Trackball. 3. To fix the first caliper, press Set. The system displays a second active caliper. 4. To position the caliper at the identical point in the second or third cycle, move the Trackball. Figure 9-3. Two Heart Beat Reference 5. To complete the...
Page 399
OB Measurements and Calculations LOGIQ 5 Basic User Manual 9-33 Direction 2300002-100 Rev. 2 Humerus Length To measure humerus length, make one distance measurement: 1. Select HL; an active caliper displays. 2. To position the active caliper at the start point, move the Trackball. 3. To fix the start point, press Set. The system fixes the first caliper and displays a second active caliper. 4. To position the second active caliper at the end point, move the Trackball. A dotted line connects the...
Page 400
OB/GYN 9-34 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Occipitofrontal Diameter To measure occipitofrontal diameter, make one distance measurement: 1. Select OFD; an active caliper displays. 2. To position the active caliper at the start point, move the Trackball. 3. To fix the start point, press Set. The system fixes the first caliper and displays a second active caliper. 4. To position the second active caliper at the end point, move the Trackball. A dotted line connects the...