Home > GE > Ultrasound System > GE Logiq P5 User Manual

GE Logiq P5 User Manual

Here you can view all the pages of manual GE Logiq P5 User Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.

Page 481

Cardiology Measurements
LOGIQ 5 Basic User Manual 10-11
Direction 2300002-100 Rev. 2
Left Atrium
The following are cardiac B-Mode Left Atrium measurements:
Two distance
measurements•  Left Atrium Diameter to AoRoot Diameter Ratio (LA/Ao 
Ratio)
Making two
distance
measurements1. Select LA/Ao Ratio; an active caliper displays.
2.  Make the first distance measurement:
a.  To position the active caliper at the start point, move the 
Trackball.
b.  To fix the start point, press Set.
The system fixes the...

Page 482

Cardiology
10-12 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Left Atrium (continued)
One distance
measurement•  Left Atrium Diameter (LA Diam)
•  Left Atrium Length (LA Major)
•  Left Atrium Width (LA Minor)
Making a one
distance
measurement1.  Select the measurement; an active caliper displays.
2.  To position the active caliper at the start point, move the 
Trackball.
3.  To fix the start point, press Set.
The system fixes the first caliper and displays a second 
active caliper.
4.  To...

Page 483

Cardiology Measurements
LOGIQ 5 Basic User Manual 10-13
Direction 2300002-100 Rev. 2
Left Atrium (continued)
One trace
measurements•  Left Atrium Area
• Diastolic (LAA(d)) 
• Systolic (LAA(s))
Making a trace
measurement1.  Select the measurement; an active caliper displays.
2.  To position the caliper at the start point, move the Trackball.
3.  To fix the trace start point, press Set.
The system displays a vertical dotted line.
4.  To trace the measurement area, move the Trackball.
A line shows the...

Page 484

Cardiology
10-14 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Left Ventricle
The following are cardiac B-Mode Left Ventricle measurements:
One distance
measurements•  Left Ventricle Mass
• Diastolic (LVP Wd)
• Systolic (LV P W s)
•  Left Ventricle Volume, Teichholz
• Diastolic (LVID d)
• Systolic (LV I D s)
•  Left Ventricle Volume, Cubic
• Diastolic (LVID d)
• Systolic (LV I D s)
•  Left Ventricle Internal Diameter
• Diastolic (LVID d)
• Systolic (LV I D s)
•  Left Ventricle Length
• Diastolic...

Page 485

Cardiology Measurements
LOGIQ 5 Basic User Manual 10-15
Direction 2300002-100 Rev. 2
Left Ventricle (continued)
Making a one
distance
measurement1.  Select the measurement; an active caliper displays.
2.  To position the active caliper at the start point, move the 
Trackball.
3.  To fix the start point, press Set.
The system fixes the first caliper and displays a second 
active caliper.
4.  To position the second active caliper at the end point, move 
the Trackball.
A dotted line connects the measurement...

Page 486

Cardiology
10-16 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Left Ventricle (continued)
One trace
measurements•  Left Ventricle Area, Two Chamber
• Diastolic (LVA (d))
• Systolic (LVA  (s ))
•  Left Ventricle Area, Four Chamber
• Diastolic (LVA (d))
• Systolic (LVA  (s ))
•  Left Ventricle Area, Short Axis
• Diastolic (LVA (d))
• Systolic (LVA  (s ))
•  Left Ventricle Endocardial Area, Width (LVA (d))
•  Left Ventricle Epicardial Area, Length 
• Diastolic (LVAe pi  (d))
• Systolic (LVA e p i...

Page 487

Cardiology Measurements
LOGIQ 5 Basic User Manual 10-17
Direction 2300002-100 Rev. 2
Left Ventricle (continued)
One time interval
measurements• Heart Rate, Teichholz 
•  Heart Rate for Two Chamber study
•  Heart Rate for Four Chamber study
•  Heart Rate for Two Chamber Area-Length study
•  Heart Rate for Two Chamber Method of Disk study
•  Heart Rate for Four Chamber Area-Length study
•  Heart Rate for Four Chamber Method of Disk study
•  Heart Rate for Bi-Plane Method of Disk study
Making a time...

Page 488

Cardiology
10-18 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Left Ventricle (continued)
Body surface area
and LV mass
measurements•  Left Ventricle Mass Index,
• Diastolic (LVP Wd)
• Systolic (LV P W s)
The system calculates body surface area from the 
patient’s height and weight. Left Ventricle Mass is a one 
distance measurement.
Making a one
distance
measurement1.  Select the measurement; an active caliper displays.
2.  To position the active caliper at the start point, move the...

Page 489

Cardiology Measurements
LOGIQ 5 Basic User Manual 10-19
Direction 2300002-100 Rev. 2
Left Ventricle (continued)
Body surface area
and stroke volume
measurements•  Left Ventricle Stroke Index, Single Plane, Two Chamber, 
Method of Disk (LV I D d, LV I D s)
•  Left Ventricle Stroke Index, Single Plane, Four Chamber, 
Method of Disk (LV I D d, LV I D s)
•  Left Ventricle Stroke Index, Bi-Plane, Bullet 
•  Left Ventricle Stroke Index, Bi-Plane, Method of Disk (LVAd, 
LVA s)
The system calculates body surface...

Page 490

Cardiology
10-20 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Mitral Valve
The following are cardiac B-Mode Mitral Valve measurements:
One distance
measurements•  Mitral Valve Annulus Diameter (MV Ann Diam)
•  E-Point-to-Septum Separation (EPSS)
Making a one
distance
measurement1.  Select the measurement; an active caliper displays.
2.  To position the active caliper at the start point, move the 
Trackball.
3.  To fix the start point, press Set.
The system fixes the first caliper and displays a...