Page 521
Cardiology Measurements LOGIQ 5 Basic User Manual 10-51 Direction 2300002-100 Rev. 2 Mitral Valve (continued) Two distance measurement• Mitral Valve E-Peak to A-Peak Ratio (A-C and D-E) (MV E/A Ratio) Making a two distance measurement1. Select MV E/A Ratio; an active caliper displays. 2. Make the first distance measurement: a. To position the active caliper at the start point, move the Trackball. b. To fix the start point, press Set. The system fixes the first caliper and displays a second active...
Page 522
Cardiology 10-52 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Mitral Valve (continued) One time interval/ slope measurements• Mitral Valve Acceleration Time (MV AccT) • Mitral Valve Deceleration Time (MV DecT) Making a time interval/slope measurement1. Select the measurement; an active caliper displays. 2. To position the caliper at the start point, move the Trackball. 3. To fix the start point, press Set. The system fixes the first caliper and displays a second active caliper. 4. To...
Page 523
Cardiology Measurements LOGIQ 5 Basic User Manual 10-53 Direction 2300002-100 Rev. 2 Mitral Valve (continued) One time interval measurements• Mitral Valve Ejection Time ((MV Trace) • Mitral Valve A-Wave Duration (MV A Dur) • Mitral Valve Time to Peak (MV Trace) • Time Making a time interval measurement1. Select the measurement. The system displays an active caliper and a vertical dotted line. 2. To position the caliper at the start point, move the Trackball. 3. To fix the first caliper, press...
Page 524
Cardiology 10-54 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Pulmonic Valve The following are cardiac Doppler Mode Pulmonic Valve measurements: One peak velocity measurements• Pulmonic Insufficiency Peak Pressure Gradient (PR Vmax) • Pulmonic Insufficiency End-Diastolic Pressure Gradient (PR Trace) • Pulmonic Valve Peak Pressure Gradient (PV Vmax) • Pulmonic End-Diastolic Pressure Gradient (PR Trace) • Pulmonic Insufficiency Peak Velocity (PR Vmax) • Pulmonic Insufficiency...
Page 525
Cardiology Measurements LOGIQ 5 Basic User Manual 10-55 Direction 2300002-100 Rev. 2 Pulmonic Valve (continued) One velocity flow trace measurements• Pulmonary Artery Diastolic Pressure (PV Trace) • Pulmonic Insufficiency Mean Pressure Gradient (PR Trace) • Pulmonic Valve Mean Pressure Gradient (PV Trace) • Pulmonic Insufficiency Mean Velocity (PR Trace) • Pulmonic Insufficiency Mean Square Root Velocity(PR Trace) • Pulmonic Insufficiency Velocity Time Integral (PR Trace) • Pulmonic Valve Mean...
Page 526
Cardiology 10-56 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Pulmonic Valve (continued) One slope measurements• Pulmonic Insufficiency Pressure Half Time (PR PHT) • Pulmonic Valve Flow Acceleration (PV AccT) Making a slope measurement1. Select the measurement; an active caliper displays. 2. To position the caliper at the start point, move the Trackball. 3. To fix the start point, press Set. The system fixes the first caliper and displays a second active caliper. 4. To position the...
Page 527
Cardiology Measurements LOGIQ 5 Basic User Manual 10-57 Direction 2300002-100 Rev. 2 Pulmonic Valve (continued) Making a time interval measurement1. Select the measurement. The system displays an active caliper and a vertical dotted line. 2. To position the caliper at the start point, move the Trackball. 3. To fix the first caliper, press Set. The system displays a second active caliper. 4. To position the caliper at the end point, move the Trackball. 5. To complete the measurement, press Set. The...
Page 528
Cardiology 10-58 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Right Ventricle The following are cardiac Doppler Mode Right Ventricle measurements: One peak velocity measurements• Right Ventricle Outflow Tract Peak Pressure Gradient (RVOT Vmax) • Right Ventricle Systolic Pressure (RVOT Vmax) • Right Ventricle Outflow Tract Peak Velocity (RVOT Vmax) Making a peak velocity measurement1. Select the measurement; an active caliper with a vertical dotted line displays. 2. To position the...
Page 529
Cardiology Measurements LOGIQ 5 Basic User Manual 10-59 Direction 2300002-100 Rev. 2 Right Ventricle (continued) Making a velocity flow trace measurement1. Select the measurement; an active caliper displays. 2. To position the caliper at the start point, move the Trackball. 3. To fix the trace start point, press Set. The system displays a vertical dotted line. 4. To trace the envelope, move the Trackball. A line shows the traced area. 5. To complete the trace, press Set. The system displays a second...
Page 530
Cardiology 10-60 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Right Ventricle (continued) Making a time interval measurement1. Select the measurement. The system displays an active caliper and a vertical dotted line. 2. To position the caliper at the start point, move the Trackball. 3. To fix the first caliper, press Set. The system displays a second active caliper. 4. To position the caliper at the end point, move the Trackball. 5. To complete the measurement, press Set. The system...