Page 571
Vascular Measurements LOGIQ 5 Basic User Manual 11-11 Direction 2300002-100 Rev. 2 A/B Ratio In B-Mode and M-Mode, you can calculate A/B ratio by diameter or by area. DiameterTo calculate A/B ratio by diameter: 1. From the Generic Touch Panel, select A/B Ratio. 2. Select ratio(Diam). The system displays an active caliper. 3. Make a distance measurement of the first velocity. The system displays an active caliper for the second distance measurement. 4. Make a distance measurement of the second...
Page 572
Vascular 11-12 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 M-Mode Measurements Figure 11-4. M-Mode Touch Panel NOTE: The following instructions assume that you first scan the patient and then press Freeze.
Page 573
Vascular Measurements LOGIQ 5 Basic User Manual 11-13 Direction 2300002-100 Rev. 2 % Stenosis In M-Mode, you measure % Stenosis by diameter. NOTE: Always take the measurement from a cross-sectional view of the vessel. 1. Select % Stenosis. The system displays an active caliper. 2. Make a distance measurement of the inner area of the blood vessel: a. To position the active caliper at the start point, move the Trackball. b. To fix the start point, press Set. The system fixes the first caliper and...
Page 574
Vascular 11-14 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 A/B Ratio In M-Mode you can measure A/B ratio by diameter, time, or velocity. DiameterTo calculate A/B ratio by diameter: 1. Select A/B Ratio. 2. Select ratio(Diam). The system displays an active caliper. 3. Make a distance measurement of the first velocity. a. To position the active caliper at the start point, move the Trackball. b. To fix the start point, press Set. The system fixes the first caliper and displays a second...
Page 575
Vascular Measurements LOGIQ 5 Basic User Manual 11-15 Direction 2300002-100 Rev. 2 A/B Ratio (continued) TimeTo calculate A/B ratio by time: 1. Select A/B. 2. Select ratio(Time). The system displays an active caliper. 3. To position the caliper at the A point, move the Trackball. 4. To fix the measure point, press Set. The system displays a second active caliper. 5. To position the second caliper at the B point, move the Trackball. 6. To complete the measurement, press Set. The system displays the...
Page 576
Vascular 11-16 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Doppler Mode Measurements Figure 11-5. D-Mode Touch Panel Example Control Assignment Cancel TransferAfter the Auto Vascular calculation results are assigned to a particular vessel, the user can cancel the assignment and assigned parameters are removed from Worksheet and Report page. When Cancel Transfer occurs, a message appears on the screen to indicate the value was erased from Worksheet and Report page. Vessel locationIf the...
Page 577
Vascular Measurements LOGIQ 5 Basic User Manual 11-17 Direction 2300002-100 Rev. 2 Control Assignment (continued) Modify Auto CalcsWhen you select this key, the Modify Calculation menu is displayed as below. In this menu, you select parameters to display in the Auto Vascular Calculation window. Only parameters that can be used by the calculation are displayed. Select Save as Default to save the selected parameters as the default calculations for this application. Select Return to return to the...
Page 578
Vascular 11-18 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Naming format for vessels When you want to measure a vessel, on the Touch Panel you select the folder for the vessel. Many vessel folders are labeled with an abbreviation. The following table lists abbreviations used for naming vascular vessels. Table 11-2: Vascular Vessel Abbreviations AcronymName ACA Anterior Cerebral Artery AComA Anterior Communicating Artery Axill Axillary Artery Axill V Axillary Vein BA Basilar Artery Basil V...
Page 579
Vascular Measurements LOGIQ 5 Basic User Manual 11-19 Direction 2300002-100 Rev. 2 PCA Posterior Cerebral Artery PComA Posterior Communicating Artery SMA Superior Mesenteric Artery SMV Superior Mesenteric Vein SUBC Subclavian Artery SUBC V Subclavian Vein Sup Femoral Superficial Femoral Artery TCD Transcranial Doppler TIPS Transjugular Intrahepatic Portosystemic Shunt VERT Vertebral Artery Table 11-2: Vascular Vessel Abbreviations AcronymName
Page 580
Vascular 11-20 LOGIQ 5 Basic User Manual Direction 2300002-100 Rev. 2 Auto Vascular Calculation Overview Auto Vascular Calculation enables the LOGIQ 5 to detect and identify a cardiac cycle. It allows you to assign measurements and calculations during live timeline imaging, while the image is frozen, or in CINE. Peak values are detected for venous flow. During cardiac cycle detection, the system identifies the cardiac cycle using calipers, vertical bars, and/or highlighting of timeline data. Use of...