Cardiology Measurements
LOGIQ 5 Basic User Manual 10-81
Direction 2300002-100 Rev. 2
Generic Study
B-Mode
The Cardiology M-Mode Generic exam category includes the 
following measurements: 
• Caliper
• Point
• Area (Trace)
• Volume
• Volume (d)
• Volume (s)
•  Dist (Distance) Ratio
• Area Ratio
• Ellipse
• Open Trace
• R-R
•  VF Diam (Diameter)
 Figure 10-4. Cardiology B-Mode - Generic Touch Panel 

Page 552

Cardiology
10-82 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
M-Mode
The Cardiology M-Mode Generic exam category includes the 
following measurements: 
• LV Study
• LA/Ao
• RV Study
• D-E Excursion
• Depth
• Slope Caliper
• Caliper
• Area Ratio
• Ellipse
• Time
• Heart Rate
 Figure 10-5. Cardiology M-Mode - Generic Touch Panel 

Page 553

Cardiology Measurements
LOGIQ 5 Basic User Manual 10-83
Direction 2300002-100 Rev. 2
Doppler Mode
The Cardiology Doppler Mode Generic exam category includes 
the following measurements: 
• Point
• Caliper
• Manual Trace
•  SD Auto Calcs
• VSD Vmax
•  MV E/A Ratio
• PHT
• Time
• Heart Rate
 Figure 10-6. Cardiology Doppler Mode - Generic Touch Panel 

Page 554

Cardiology
10-84 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
ECG Option
Overview
A physiological input panel is available for the LOGIQ 5.  This 
panel has inputs for ECG, PCG and auxiliary signals.
The physiological module consists of four channels: ECG, PCG, 
AUX1 and AUX2 (AUX 2 is currently not available). The two 
channels, AUX1 and AUX2, are both capable of handling 
external ECG signals from other diagnostic ECG devices. AUX2 
is also capable of handling a pressure/pulse signal.
The...

Page 555

ECG Option
LOGIQ 5 Basic User Manual 10-85
Direction 2300002-100 Rev. 2
ECG Cable
The ECG Cable is a modular cable consisting of two different 
cable parts:
•  Single cable with a system connection at one end and a 
cable splitter at the other.
•  A triple color-coded electrode cable to be inserted into the 
splitter device. Each electrode cable hooks up to the 
appropriate stick-on electrode by a color-coded clip type 
connector.
The cable splitter device has a drawing defining the color codes, 
names...

Page 556

Cardiology
10-86 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Physiological Trace Monitor Display
The scanned image is synchronized with the ECG and PCG 
trace. In Doppler or M-Mode, the traces are synchronized with 
that particular mode’s sweep.
The user can control the gain, position and sweep speed of the 
traces using the Touch Panel controls.
 Figure 10-9. Physiological Trace Monitor Display
1. PCG
2. AUX
3. ECG
4.  Auto Heart Rate Display 

Page 557

ECG Option
LOGIQ 5 Basic User Manual 10-87
Direction 2300002-100 Rev. 2
ECG Touch Panel
The ECG Touch Panel Menu provides for control of the 
physiological input signals.
Without the ECG option, the ECG Touch Panel menu is not 
displayed.
 Figure 10-10. ECG Touch Panel Menu
Sweep Speed
Description Change the speed of the trace. The sweep speed of the physio 
signal on the B-Mode image can be set independent of the 
timeline (Doppler and M-Mode) sweep speed.
Va l u e 1  -  1 6 . 

Page 558

Cardiology
10-88 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Timer Trigger
Description Enables intermittent imaging based on a timer.
NOTE:  If Timer Trigger is turned on, the ECG Trigger is set to None.
Value On or Off.
ECG Trigger
Description Enables intermittent imaging based on the ECG. The trigger 
location(s) relative to the R trigger are set with the Delay Time 
key.
NOTE:  If other than None is selected, Timer Trigger is turned off.
Value None, Trig1, Trig2, and Both.
AUX Display...

Page 559

ECG Option
LOGIQ 5 Basic User Manual 10-89
Direction 2300002-100 Rev. 2
ECG Display
Description Provides the ability to turn on the ECG trace and Auto Heart 
Rate for display on the monitor.
Adjusting When the key is selected, the ECG trace and Auto Heart Rate 
toggle between on and off.
Value On or Off
ECG Trigger Period
Description The control specifies the number of heart cycles (R-waves) that 
are skipped between ECG triggers. 
The default is 1 or no skipping.
Adjusting Adjust the corresponding...

Page 560

Cardiology
10-90 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
AUX Gain/Position
Description Allows for the amplitude control of the AUX trace or allows for 
the vertical positioning of the AUX trace on the image display.
Adjusting Press the knob to toggle between Gain and Position. The default 
is Gain.
Value Gain: 0.01 - 4.00
Position: 0.00 - 100.00
PCG Gain/Position
Description Allows for the amplitude control of the PCG trace or allows for 
the vertical positioning of the PCG trace on the...