GYN Measurements
LOGIQ 5 Basic User Manual 9-95
Direction 2300002-100 Rev. 2
To Start a Gynecology Exam
To begin a gynecology exam, you enter patient data or, if the 
patient data from a previous exam is saved in the system, find 
the patient information. 
For details about how to start an exam, See ‘To Start an 
Obstetrics Exam’ for more information.
After you complete the patient information, you can begin the 
scan. 
1.  To change from the Patient Data Entry screen to the Scan 
screen, do one of the...

Page 462

OB/GYN
9-96 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
B-Mode Measurements
In B-Mode, you make the measurements in the General 
Gynecology study. These measurements include: 
•  Uterine length, width, and height 
•  Ovarian length, width, and height
• Ovarian follicle
• Endometrium thickness
 Figure 9-30. General Gynecology study 

Page 463

GYN Measurements
LOGIQ 5 Basic User Manual 9-97
Direction 2300002-100 Rev. 2
Follicle measurements
You can make left and right ovary follicle measurements from 
one, two, or three distances. 
One distance1.  To select the left or right, adjust the Side knob.
2. Select Follicle; an active caliper displays.
3.  To position the active caliper at the start point, move the 
Trackball.
4.  To fix the start point, press Set.
The system fixes the first caliper and displays a second 
active caliper.
5.  To...

Page 464

OB/GYN
9-98 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Follicle measurements (continued)
Two distances1.  To select the left or right, adjust the Side knob.
2. Select Follicle; an active caliper displays.
3.  Make the first distance measurement:
a.  To position the active caliper at the start point, move the 
Trackball.
b.  To fix the start point, press Set.
The system fixes the first caliper and displays a second 
active caliper.
c.  To position the second active caliper at the end point,...

Page 465

GYN Measurements
LOGIQ 5 Basic User Manual 9-99
Direction 2300002-100 Rev. 2
Follicle measurements (continued)
Three distances1.  To select the left or right, adjust the Side knob.
2. Select Follicle; an active caliper displays.
3.  Make the first distance measurement:
a.  To position the active caliper at the start point, move the 
Trackball.
b.  To fix the start point, press Set.
The system fixes the first caliper and displays a second 
active caliper.
c.  To position the second active caliper at the...

Page 466

OB/GYN
9-100 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Endometrium thickness (Endo)
To measure the endometrium thickness, make one distance 
measurement.
1. Select Endo; an active caliper displays.
2.  To position the active caliper at the start point, move the 
Trackball.
3.  To fix the start point, press Set.
The system fixes the first caliper and displays a second 
active caliper.
4.  To position the second active caliper at the end point, move 
the Trackball.
A dotted line connects the...

Page 467

GYN Measurements
LOGIQ 5 Basic User Manual 9-101
Direction 2300002-100 Rev. 2
Ovary length, width, and height 
You can measure the length, width, and height of the left and 
right ovaries. Each measurement is a typical distance 
measurement made in the appropriate scan plane. Typically, the 
width and height are measured on the axial plane while the 
length is measured on the sagittal plane.
To measure ovarian length, width, or height:
1.  Scan the patients right or left ovary in the appropriate plane....

Page 468

OB/GYN
9-102 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Uterus length, width, and height
Each of these is a standard distance measurement. Length is 
typically measured in the sagittal plane. Width and height are 
measured in the axial plane.
To measure uterus length, width, or height:
1.  Scan the patient in the appropriate scan plane. 
2. Select the UT folder, then select UT L, UT W, or UT H. 
An active caliper displays.
3.  Perform a standard distance measurement: 
a.  To position the...

Page 469

GYN Measurements
LOGIQ 5 Basic User Manual 9-103
Direction 2300002-100 Rev. 2
M-Mode Measurements
M-Mode measurements for the Gynecology exam are identical 
to M-Mode measurements for the Obstetrics exam. These 
measurements include % stenosis, A/B ratio, and heart rate.
For details regarding these measurements, See ‘M-Mode 
Measurements’ for more information.
 Figure 9-31. M-Mode Generic study 

Page 470

OB/GYN
9-104 LOGIQ 5 Basic User Manual
Direction 2300002-100 Rev. 2
Doppler Mode Measurements
Doppler measurements for the Gynecology exam are identical 
to Doppler measurements for the Obstetrics exam. These 
measurements include the following vessels: uterine, ovarian, 
umbilical, middle cerebral artery, aorta, placenta, and 
descending aorta. For each vessel, you can make any of the 
following measurements: peak systole, mid diastole, end 
diastole, heart rate, TAMAX, pulsatility index, resistive...