GE Cardiocap 5 Service Manual
Here you can view all the pages of manual GE Cardiocap 5 Service Manual. The GE manuals for Health/Medical Alert are available online for free. You can easily download all the documents as PDF.
Page 81
Contents Chapter 4. Troubleshooting 4.1 Messages........................................................................\ ....4-1 4.2 Troubleshooting charts.........................................................4-7 4.2.1 Start-up troubleshooting........................................................................\ .....4-7 4.2.2 General........................................................................\ ..............................4-8 4.2.3...
Page 83
Troubleshooting 4. TROUBLESHOOTING See the User’s Reference Manual for more messages and troubleshooting procedures. 4.1 Messages MESSAGE POSSIBLE CAUSE ACTION Acknowl. alarms silenced Acknowledged alarms are silenced (Silence Alarms key pressed during silencing period). No action required. Air leakage Hose and/or cuff leaking. Damaged cuff, cuff connector, or hose double connector. Replace cuff, cuff connector, and/or hose double connector. Hose and cuff OK, but tube disconnected or...
Page 84
Cardiocap/5 Technical Reference Manual MESSAGE POSSIBLE CAUSE ACTION Call service: Error X (X = the NIBP hardware 1 ROM checksum error; memory failure Change NIBP board. error number shown in next column) 2 + 15 V failure. Check short circuits. Change NIBP board. 3 – 15 V failure. Check short circuits. Change NIBP board. 4 EEPROM protection switch error. Change NIBP calibration protection to OFF by turning the gear wheel and selecting Protection OFF in the NIBP Calibration menu....
Page 85
Troubleshooting MESSAGE POSSIBLE CAUSE ACTION Cuff loose (NIBP) Hose and/or cuff not connected. Connect the hose and the cuff. Cuff loosely wrapped. Tighten the cuff. Leakage in cuff or hose. Replace cuff/hose. Leakage inside module. Check/fix internal tubing and air chamber. Pump does not work. Check pump connector; if O.K., replace pump. No pulses during the last three measurements Check cuff positioning. Cuff occlusion NIBP cuff and/or hose occlusion due to kinked cuff tube or kinked...
Page 86
Cardiocap/5 Technical Reference Manual MESSAGE POSSIBLE CAUSE ACTION NIBP measurement removed Signal cable is loose or not attached. Check signal cable connections from Mother board to CPU board. No P1/P2 transducer Invasive pressure or transducer cable not connected. Connect invasive pressure or transducer cable. No SpO2 probe SpO2 sensor is not connected to the monitor. Check sensor connections. Faulty or wrong type of SpO2 sensor (sensor is not specified for use with the installed pulse...
Page 87
Troubleshooting MESSAGE POSSIBLE CAUSE ACTION Response too weak (NMT) The maximum gain is insufficient to increase the response signal amplitude to a measurable level. This can occur if: • Stimulation current is too weak. • Stimulation electrodes are not connected or they are improperly placed on the nerve. • Recording electrodes are disconnected. • Electrode(s) dry and should be replaced. • Skin at electrode site is improperly prepared. Reposition the stimulating or recording electrodes....
Page 88
Cardiocap/5 Technical Reference Manual MESSAGE POSSIBLE CAUSE ACTION Unstable zero pressure Pressure is unstable when starting the NIBP measurement. Calm the patient and retry. Voltage error Erroneous voltage level detected. Replace power supply unit. Weak pulsation Weak or unstable oscillation pulses due to: • Marked arrhythmia; too few pulses detected; weak or unusual blood circulation. • Artifacts (accurate diastolic pressure difficult to measure) or marked drop in diastolic pressure. •...
Page 89
Troubleshooting 4.2 Troubleshooting charts 4.2.1 Start-up troubleshooting No picture on screen NOTES ⇓ Check that the mains power is 100 to 240VAC. If the mains power is too low or high the power supply does not work. Check that the mains fuses have not blown. Fuse type has to be T2AH. Check that the monitor is not in the SHUT DOWN mode. Detach the mains cord from the monitor for at least 30 seconds to reset possible SHUT DOWN mode. OK but still no display ⇓ Check that the green LED on...
Page 90
Cardiocap/5 Technical Reference Manual 4.2.2 General SITUATION POSSIBLE CAUSE ACTION Keyboard and ComWheel are not working Keyboard and ComWheel cables not connected to CPU board. Detach monitor rear cover and check the cable connections to the CPU. Faulty CPU. Replace the CPU if the cable connections are OK Keyboard is not working but the ComWheel works Keyboard cable not properly connected to the CPU board. Detach monitor rear cover and check the keyboard cable connection to CPU....
All GE manuals