Page 121

Frames and Software 
5.6 Display, input, and output 
5.6.1 Display 
 
NOTE: The LCD display backlight circuit runs on high voltage. Do not touch the inverter board 
when powered. 
The display consists of 10.4 inch TFT LCD display unit, backlight board, and inverter board.  
The CPU board drives the display. The picture data, synchronization, and +5 V supply voltage are 
connected from the CPU board to the display unit via the display adapter board. 
The +12 V supply voltage connects to the backlight...

Page 122

Cardiocap/5 Technical Reference Manual 
5.6.4 I/O board 
The I/O board contains four different I/O connectors: 
• 44-pin female D connector for digital/analog output. 
• 9-pin male D connector for RS-232 serial I/O. 
• 9-pin female D connector for connecting NET ID plug. 
• 5-pin 180 DIN connector for connecting RemCo keyboard. 
Data from the CPU is connected through a 40-pin flat cable connector. Power from the Mother board is 
connected through a 16-pin flat cable connector. 
The NET function...

Page 123

Frames and Software 
5.7 Recorder and recorder board (N-XREC option) 
You can use the optional built-in recorder to do the following:  
• Record the graphical or numerical trend and print up to 24 hours of trend data. 
• Record up to three real-time waveforms simultaneously. 
• Display recordings of numerical information in horizontal and vertical plane.  
NOTE:  Thermal paper printouts may be destroyed when exposed to light, heat, alcohol, etc. Make a 
photocopy for your archives. 
 
Figure 5-11....

Page 124

Cardiocap/5 Technical Reference Manual 
 
 
5-26  

Page 125

Contents 
Chapter 6. Measurement Parameters  
6.1 Introduction........................................................................\
.6-1 
6.1.1 Hemodynamic parameters and related options.............................................6-1 
6.1.2 Airway gases and associated options...........................................................6-2 
6.2 Measurement principles for hemodynamic parameters...........6-3 
6.2.1 NIBP........................................................................\...

Page 126

Cardiocap/5 Technical Reference Manual 
ECG filtering ........................................................................\
...................... 6-18 
Respiration section........................................................................\
............ 6-18 
Microprocessor  section ........................................................................\
...... 6-19 
Isolated section.......................................................................\
.................. 6-19 
Power supply...

Page 127

Contents 
Table of Figures 
Figure 6-1. Hemodynamic parameters measurement unit (NESTPR)....................................6-1 
Figure 6-2. Airway gases measurement unit......................................................................6-2 
Figure 6-3. Absorption of infrared light in the finger............................................................6-4 
Figure 6-4. Comparative light absorption........................................................................\
..6-6 
Figure 6-5. Extinction...

Page 128


				            

Page 129

Measurement Parameters 
6. MEASUREMENT PARAMETERS 
6.1 Introduction 
This chapter provides information about the Cardiocap/5 standard measurement features and all 
measurement options.  
6.1.1 Hemodynamic parameters and related options 
The Cardiocap/5 Parameter unit (NESTPR) provides measurement of general hemodynamic 
parameters.  
 
Figure 6-1. Hemodynamic parameters measurement unit (NESTPR) 
NESTPR represents the measurement parameters listed below: 
N = Non-invasive blood pressure (NIBP) 
E = ECG...

Page 130

Cardiocap/5 Technical Reference Manual 
6.1.2 Airway gases and associated options 
The Cardiocap/5 Airway Gas options (N-XC, N-XCO, N-XCAiO) provide airway and respiratory 
measurements through the airway gas measurement unit.  
 
Figure 6-2. Airway gases measurement unit 
The features of each option are noted in the chart below. 
NOTE:  One or more of the airway gas options is required for the Patient Spirometry (N-XV) and 
NeuroMuscular Transmission (N-XNMT) options.  
 
Option 
 
CO2
 
N2O 
 
O2...