Page 71
Planned Maintenance 3.2 Planned maintenance checks 1. Visual inspection 1. Switch the monitor to standby and disconnect all cables from the rear of the monitor. 2. Check the external parts of the monitor. For example, check that the power cord receptacle is intact and the D-fend latch moves properly. 3. Check that the markings of the mains fuses correspond to the fuse plate. 2. Parts replacement 1. Dissamble the frame (see section 7.3.1). 2. Replace these items in the gas unit: occlusion...
Page 72
Cardiocap/5 Technical Reference Manual 3. Functional inspection 1. Connect the power cord and check that the Battery charge status LED turns on or flashes. 2. Switch on the monitor and check that it starts up as described below: • Both alarm LEDs flash On and Off. • The start-up sound is heard from the speaker. • The normal monitoring screen appears and no error messages are displayed. NOTE: “Check network connectors” is displayed if the N-XNET or N-XDNET option is installed. • The time and...
Page 73
Planned Maintenance 4. Check the capacity time of the backup battery: Disconnect the power cord (without switching the monitor to standby). Note the time and make sure that the monitor continues to run normally with the battery for at least 15 minutes. The battery indicator should appear on the screen: NOTE: You can continue the check while the monitor is powered by batteries. 5. Configure the monitor screen according to the monitor configuration so that all parameter information is displayed....
Page 74
Cardiocap/5 Technical Reference Manual 6. Non-invasive blood pressure (NIBP) measurement test NOTE: See the Service Menus chapter for illustrations of the service menus used for the NIBP test. 1. Return to the Modules service menu and select NIBP to enter the NIBP service menu. Check that NIBP board memories passed the internal memory test (RAM, ROM, and EEPROM all state OK). 2. From the NIBP service menu, select Calibrations to enter the Calibration service menu. Perform the Active Leak Test to...
Page 75
Planned Maintenance 8. Invasive blood pressure measurement test 1. Check the InvBP channels with a patient simulator. The values and waveforms should correspond to the simulator settings. NOTE: If you evaluate the measurement accuracy, remember to add the simulator’s accuracy specification to the one of the monitor. 2. Calibrate InvBP channels if necessary. 9. Gas measurement test NOTE: See the Service Menus chapter for illustrations of the service menus used for the gas measurement test....
Page 76
Cardiocap/5 Technical Reference Manual 11. Spirometry test 1. Perform the spirometry leakage test and calibration (see section 7.4.9). 2. With a sample line attached to the D-lite sensor, breathe through the wider side of the D-lite. Check that the flow waveform moves downward when you breathe in, and upward when you breathe out. 12. Gas sampling line and D-fend check 1. Block the tip of the sampling line with your finger and check that the “Sample line blocked” message is displayed on the...
Page 77
Planned Maintenance 16. Network test 1. Check that the Mon-Net cable connector and the Identification plug are clean and intact, then connect them to the monitor. Check the connection to the network by checking the states of the Network connection LEDs between the connectors: Yellow --> should flash intermittently Green --> should be lit continuously 2. Check that the monitor connects to the Datex-Ohmeda Network (that is, the network symbol is displayed on the upper right corner of the...
Page 78
Cardiocap/5 Technical Reference Manual 17. Data card test 1. Insert a memory card labeled Menu into the front-most memory card slot. Check that the “Menu card inserted” message appears in the message field. 2. Insert a memory card labeled Data into the second slot. Check that the Data card inserted message appears in the message field. 3. Check that some trend information is available in trend memory for monitored parameters. Erase the trends. Check that the trends are properly erased. Reload...
Page 79
Planned Maintenance Form Planned Maintenance Form Datex-Ohmeda Cardiocap/5 Customer Service Service Engineer Date Monitor Configuration Monitor model: Hemodynamic model F-MX Hemodynamic model with gas measurement F-MXG Measurement options Measurement options (SpO2) N-XP Two invasive pressure channels and second temperature (T2) N-XOSAT Datex-Ohmeda enhanced pulse oximetry N-XC CO2 N-XNSAT Nellcor compatible pulse oximetry N-XCO CO2, N2O, Patient Oxygen Data collection and...
Page 80
Planned Maintenance Form OK N.A. Fail 3. Functional inspection 4. ECG measurement test 5. Temperature measurement test 6. Non-invasive blood pressure (NIBP) measurement test 7. SpO2 measurement test 8. Invasive blood pressure measurement test 9. Gas measurement test 10. Anesthetic agent identification test 11. Spirometry test 12. Gas sampling line and D-fend check 13. Trend test 14. Watchdog test 15. Recorder test 16. Network test 17. Data card...