Page 31

Contents 
Chapter 2. Installation and Functional Check  
2.1 Introduction ........................................................................\
. 2-1 
2.2 Installation ........................................................................\
.. 2-1 
2.2.1 Front panel components.......................................................................\
....... 2-1 
Patient connectors  panel (F-MX)................................................................... 2-2 
Patient connectors panel (F-MXG)...

Page 32

Cardiocap/5 Technical Reference Manual 
Table of Figures 
Figure 2-1. Cardiocap/ 5 monitor (F-MXG) ........................................................................\
. 2-1 
Figure 2-2. Patient connecti ons—Hemodynamic (F-MX) ...................................................... 2-2 
Figure 2-3. Patient connections—He modynamic with gas (F-MXG) ....................................... 2-3 
Figure 2-4. Rear panel (F-MXG)........................................................................\...

Page 33

Installation and Functional Check 
2. INSTALLATION AND FUNCTIONAL CHECK 
2.1 Introduction 
This chapter includes the information needed to install and check the monitor. Information for   
connecting other equipment, such as a printer or computer, is also included. If you need assistance 
concerning the installation, please contact your authorized distributor. 
2.2 Installation 
2.2.1 Front panel components 
 
Figure 2-1. Cardiocap/5 monitor (F-MXG) 
(1) Power On/standby key 
(2) External power...

Page 34

Cardiocap/5 Technical Reference Manual 
(8) Recorder (N-XREC option) 
NOTE:  The two-button recorder (shown) is for Cardiocap/5 monitors using software version 3.0 
or higher. A one-button recorder was available previously. 
(9) Patient connectors 
(10) Spirometry connectors 
(11) NIBP connector 
(12) D-fend housing (F-MXG only) 
Patient connectors panel (F-MX) 
NIBPECGSpO2P1P2T1
T2
1234567 
Figure 2-2. Patient connections—Hemodynamic (F-MX) 
(1) NIBP 
(2) ECG 
(3) SpO2  
NOTE:  Connector type depends...

Page 35

Installation and Functional Check 
2-3 
Patient connectors panel (F-MXG) 
P2P1
SpO2
ECG
NIBP Spi r omet r y
T1
T2
NM
T
1234
5698710 
Figure 2-3. Patient connections—Hemodynamic with gas (F-MXG) 
(1)  Spirometry (N-XV option) 
(2)  NMT (N-XNMT option) 
(3)  Invasive pressure, P2 (N-XP option) 
(4)  Temperature, T2 (N-XP option) 
(5) Temperature, T1 
(6)  Invasive pressure, P1 (N-XP option) 
(7) SpO
2  
NOTE:  Connector type depends on which SpO
2 option is installed: 
 
Connector for Datex-Ohmeda...

Page 36

Cardiocap/5 Technical Reference Manual 
2.2.2 Rear panel 
1
2
3
4
567891011
12
13
 
Figure 2-4. Rear panel (F-MXG) 
(1) Built-in handle 
(2) Gas outlet (F-MXG only), X6 
(3) Remote Control connector, X5 
(4) Ethernet connector, X4 
(5) Network connection LEDs 
(6) Network identification plug connector, X3 
(7) Serial communication interface/local printer connector, X2 
(8) Analog/digital output connector, X1  
(includes nurse call and defibrillator synchronization signals) 
(9) Mounting attachment 
(10)...

Page 37

Installation and Functional Check 
2.2.3 Connection to network 
If either Cardiocap/5 Network option (N-XNET or N-XDNET) is installed, you can connect the 
Cardiocap/5 to the Datex-Ohmeda Network and iCentral. Use the Monitor-Network cable to connect 
the monitor to the network. 
1. Power off the monitor.  
2. On the rear panel, connect one of the RJ-45 connectors to connector X4 and connect the 
Identification Plug to connector X3. 
3. Connect the other RJ-45 connector to the corresponding Datex-Ohmeda...

Page 38

Cardiocap/5 Technical Reference Manual 
• Attach the other end of the line to the ventilator reservoir. Make sure that the reservoir tube 
diameter is at least 2-3 times larger than the exhaust line.  
2.2.6 Scavenging through the anesthesia gas scavenging system 
Anesthesia machines are equipped with an anesthesia gas scavenging system (AGSS), and in some 
machines you can connect the sample gas outlet directly to that. 
For example, connect the sample gas outlet to the Datex-Ohmeda S/5 Avance:...

Page 39

Installation and Functional Check 
2.2.8 Returning gas to the patient circuit 
In some anesthesia machines, you can return the sample gas to the patient circuit, refer to the 
anesthesia machines manuals. For example, if you use the Datex-Ohmeda S/5 Anesthesia Delivery 
Unit (ADU), connect an optional adapter (881644, 5/pkg) to the patient breathing tubes. 
 
Figure 2-7. Gas return to patient circuit in AS/3 ADU 
2.3 Choosing the location 
The monitor can be placed on a flat surface or hung with the...

Page 40

Cardiocap/5 Technical Reference Manual 
2.4.2 Interfacing a computer 
It is possible to interface a computer to the Cardiocap/5. For further information, please contact your 
authorized GE Healthcare distributor.  
WARNING: Connecting the power supply cord of the computer to the wall socket may cause the 
computer leakage current to exceed the limit specified for medical equipment. Always connect 
the computer to an appropriate separating transformer. 
2.4.3 Interfacing other devices using the...