GE Cardiocap 5 Service Manual
Here you can view all the pages of manual GE Cardiocap 5 Service Manual. The GE manuals for Health/Medical Alert are available online for free. You can easily download all the documents as PDF.
Page 22
Cardiocap/5 Technical Reference Manual 1.6.5 Impedance respiration Respiration range: 4 to 120 respirations/minute Accuracy: ± 5% or ± 5 bpm 1.6.6 Pulse oximetry, standard Display update time: 5 seconds Averaging time: adjustable Plethysmographic waveform scaling: adjustable SpO2 Calibration range: 50 to 100% Calibrated against functional saturation Measurement range: 40 to 100% Measurement accuracy (% SpO2 ±1 SD): 80 to 100% ± 2 digits; 50 to 80% ± 3 digits; Below 50%...
Page 23
Overview Power adapter: IPX0, ordinary equipment. Power adapter for Transport Vehicles: IPX1. • EQUIPMENT not suitable for use in the presence of a FLAMMABLE ANAESTHETIC MIXTURE with air or with oxygen or nitrous oxide. • CONTINUOUS OPERATION according to the mode of operation. • CISPR 11: Group 1, Class A. Group 1 contains all ISM (Industrial, scientific and medical) equipment in which there is intentionally generated and/or used conductively coupled radio-frequency energy which is necessary...
Page 24
Cardiocap/5 Technical Reference Manual 1.7.3 Pulse oximetry, Nellcor compatible (N-XNSAT) Display update time: 5 seconds Averaging time: 5 to 7 seconds Plethysmographic waveform scaling: automatic SpO2 Calibrated against functional saturation Measurement range: 1 to 100% Measurement accuracy (% SpO2 ±1 SD): 70 to 100% (± 2 digits to ± 3.5 digits, depending on the sensor) Below 70% unspecified See the Users Reference Manual (Pulse Oximetry chapter) for a list of approved sensors and...
Page 25
Overview Automatic compensation for pressure, CO2-N2O and CO2-O2 collision broadening effect Warm-up time: 2 minutes for operation with CO2, O2, and N2O 5 minutes for operation of anesthetic agents 30 minutes for full specifications Autozeroing interval: immediately after “Calibrating gas sensor” message and 2, 5, 10, 15, 30, 45, 60 minutes after start-up, then every 60 minutes Respiration rate (RR) Measurement range: 4 to 60 breaths/minute Detection criteria: 1 % variation in CO2...
Page 26
Cardiocap/5 Technical Reference Manual MAC Range: 0 to 9.9 MAC Equation: MAC(AA)=%(ETAA) x(AA) ETNO 100 2+% where x(AA): Hal = 0.75%, Enf = 1.7%, Iso = 1.15%, Sev = 2.05%, Des = 6.0%. Normal conditions After 30-minute warm-up period: Ambient temperature: 18 to 28 °C, within ± 5 °C of calibration Ambient pressure: 500 to 800 mmHg, ± 50 mmHg of calibration Ambient humidity: 20 to 80% RH, ± 20% RH of calibration Non-disturbing gases: Ethanol C2H5OH (< 0.3%) Acetone (< 0.3%) Methane CH4...
Page 27
Overview Parameter Accuracy under Condition n CO2 ± (0.3 vol% + 4% of reading); at 5 vol% error ± 0.5 vol% O2 ± (2 vol% + 2% of reading) N2O ± (3 vol% + 3% of reading Agents (Des, Enf, Hal, Iso, Sev) ± (0.2 vol% + 10% of reading) Parameter Accuracy under Condition o CO2 ± (0.4 vol% + 7% of reading); at 5 vol% error ± 0.75 vol% O2 ± (3 vol% + 3% of reading) N2O ± (3 vol% + 5% of reading) Agents (Des, Enf, Hal, Iso, Sev) ± (0.3 vol% + 10% of reading) Parameter Accuracy under...
Page 28
Cardiocap/5 Technical Reference Manual Detection through D-lite™ or Pedi-lite™ flow sensor and gas sampler: Measurement D-lite flow sensor (adult) Pedi-lite flow sensor (pediatric) Tidal volume Measurement range Resolution Accuracy (typical value) 150 to 2000 ml 1 ml ± 6% or 30 ml (whichever is larger) 15 to 300 ml 1 ml ± 6% or 4 ml (whichever is larger) Minute volume Measurement range Resolution Accuracy (typical value) 2 to 20 l/minute 0.1 l/minute ± 6% 0.5 to 5 l/minute...
Page 29
Overview Stimulator Stimulus pulse: Square wave, constant current Pulse width: 100, 200 or 300 µs Stimulus current range (supramax and manual): 10 to 70 mA with 5 mA steps Stimulus current accuracy: 10% or ±3 mA (whichever is greater) Maximum load: 3 kΩ Maximum voltage: 300 V Regional block mode Stimulation modes: Single twitch Intervals: 1 second, 2 seconds, 3 seconds Stimulus pulse: Square wave, constant current Pulse width: 40 µs Stimulus current range: 0 to 5.0 mA with 0.1 mA...
All GE manuals