Service Procedures 7.3.6 AC/DC unit replacement 1. Remove the 3 screws that hold the AC/DC unit to the backplate unit. 2. Disconnect AC/DC–DC/DC cable from the DC/DC board. 3. Disconnect the ground wire. 4. Lift the AC/DC unit carefully up. 7.3.7 DC/DC board replacement You may move the AC/DC unit out of the way first to improve access to the DC/DC board. 1. Disconnect the DC/DC–AC/DC board cable from the DC/DC board. 2. Disconnect the DC/DC–Mother board (red & black color) cable from the...
Page 172
Cardiocap/5 Technical Reference Manual 7.3.10 CPU battery replacement CAUTION: Discard the lithium battery according to local regulations. Before replacing the battery, determine the monitor software version. Look on the small sticker on the rear of the monitor. The software version follows the dash after the part number (for example, in 898716–1.0, the monitor software is version 1.0; in 898716–2.4, it is version 2.4). If you cannot determine the software version from the sticker, look at the...
Page 173
Service Procedures 7.3.11 CPU board replacement Important: The new CPU board does not contain software or network options. If the network is required, make sure the CPU board supports the network. You must load the software after you replace the board. Refer to the Spare Parts chapter. 1. Disconnect the following flat cables from the CPU board:
Page 174
Cardiocap/5 Technical Reference Manual 7.3.13 Enhanced pulse oximetry (N-XOSAT) or SpO2 interface board replacement 1. Disconnect the Mother board cable from the SpO2 interface board. 2. Remove the 3 screws that secure the assembly cover to the front panel. 3. Remove the 4 screws that secure the SpO2 interface board to the assembly cover. 4. To remove the pulse oximetry board cover from the SpO2 interface board, gently bend the tabs to an upright position and lift the cover off. 5. Remove the 4...
Page 176
Cardiocap/5 Technical Reference Manual 7.4 Adjustments and calibration You use service menus to perform procedures in this section. Once you have entered the Service View menu, you can enter and exit the specific service menus at which you make the adjustments. To enter the Service View menu, from which you access all other service menus: 1. Press the ComWheel and select Monitor Setup. 2. Select Install/Service and enter the password (16-4-34). 3. Select Service and enter the password (26-23-8)...
Page 177
Service Procedures 7.4.2 NIBP calibration check NIBP pressure measurement electronics are calibrated at the factory. The processor maintains zeroing pressure automatically. If the zero point of the pressure transducer drifts more than specified, an error message is given and the NIBP board should be recalibrated or replaced. Checking the primary pressure channel calibration 1. To check the primary pressure channel calibration, press the NIBP key and select NIBP Setup to open the NIBP Setup menu....
Page 178
Cardiocap/5 Technical Reference Manual 7.4.3 NIBP calibration A hole in the middle of the monitor bottom plate provides access to a gear wheel for the Temperature and NIBP calibration protection switches. Turning the gear wheel moves a slide that operates the switches. CAUTION: Turning the gear wheel more than 1/4 turn can damage the calibration switch. Figure 7-4. Calibration switch 1. Detach the NIBP hoses from the front panel connector to enable proper zeroing. 2. To enter the NIBP...
Page 179
Service Procedures 11. Turn the gear wheel approximately 1/4 turn counterclockwise until Protection is highlighted. Choose Protection ON and push the ComWheel. Turn the gear wheel approximately 1/4 turn clockwise until Calibration and Protection become faint. 12. Perform the NIBP calibration check (earlier in this chapter) to verify the new calibration. 7.4.4 Temperature calibration Always calibrate the temperature: • When measured test values deviate more than ± 0.1 °C. • After replacing the...
Page 180
Cardiocap/5 Technical Reference Manual 7.4.6 General checks for gas system components These faults can occur in the sampling system: • Leaks or blockages in the tubing. • Sampling pump or magnetic valves failure. • Diminished flow rates because of dirt accumulation in the internal tubing. The following checks should help in localizing the fault. Whenever you suspect the sampling system and always after you work on the sampling system, check and (if necessary) adjust the flow rate. CAUTION: The...