Page 191
Service Menus 8.3.2 Communication menu and data This service menu shows information about external communication. To enter this service menu from the Service View menu: 1. Select Monitor. 2. Select Communication. Network (with N-XNET or N-XDNET option) Shows information about communication with the Datex-Ohmeda Network. • Statistics In shows the number of received data packets and data bytes. • Statistics Out shows the number of transmitted data packets and data bytes. • Data errors shows...
Page 192
Cardiocap/5 Technical Reference Manual Communication – Remote Access menu Use the Remote Access menu to set the interface with an external PC that runs the Virtual Support™ software, a service tool for remote diagnostics. To enter this service menu from the Service View menu: 1. Select Monitor. 2. Select Communication. 3. Select Remote Access. Access Mode When set to Normal (the default), the PC can only read service menu data from the monitor. When set to Extended (future option), the PC...
Page 193
Service Menus 8.3.3 Configuration menu To enter this service menu from the Service View menu: 1. Select Monitor. 2. Select Configuration. Frame Chooses the correct frame configuration: F-MX or F-MXG. Set Chooses the parameter configuration that corresponds to the monitor hardware. The lower part of the menu screen shows the parameters included in each set. Network Shows the network setting for the monitor: • Network (Net) • Network and DataCard (Net+DC) • Neither (NONE) 8-7
Page 194
Cardiocap/5 Technical Reference Manual 8.3.4 Watchdog Tests menu To enter this service menu from the Service View menu: 1. Select Monitor. 2. Select Watchdog Tests. Watchdog This test ensures that the watchdog functions properly. This test shows running seconds with an accuracy of 0.1 seconds. When the safety element is working properly, the monitor restarts 1.5 seconds after the start of the test. In case of a malfunction, >20 s is displayed and the test will interrupt. In this case, the...
Page 195
Service Menus 8.3.5 Country Settings menu To enter this service menu from the Service View menu: 1. Select Monitor. 2. Select Country Settings. Language Sets the language used in this monitor. National Reqs Sets specific software features to fulfill national requirements in Germany or France. Power Frequency Sets the power frequency (50 Hz or 60 Hz). This setting is used to filter out possible power frequency interference from the parameter measurements. Time format Sets the time format of...
Page 196
Cardiocap/5 Technical Reference Manual 8.4 Keyboard menu and data The Keyboard menu is used to test the alarm indicator LEDs and function keys. To enter this service menu from the Service View menu, select Keyboard. Upper Led Checks upper alarm indicator LED (red). Lower Led Checks lower alarm indicator LED (yellow). Dummy Press Tests the function keys. See the description for Direct action keys below. Keyboard Type Not used in Cardiocap/5. Service Data screen Service Data counts whenever a...
Page 197
Service Menus 8.5 Modules menu To enter this service menu from the Service View menu, select Modules. The modules menu provides access to the measurement parameter menus used for the Cardiocap/5: • Gas Unit • ESTP : ECG • ESTP : STP • NIBP • NMT • M-NSAT (for N-XNSAT or N-XOSAT pulse oximetry option) • More Modules Each of these menus and screens are illustrated and described in the sections that follow. NOTE: The P/PT, PP, and COP selections in the Modules menu are not used in...
Page 198
Cardiocap/5 Technical Reference Manual 8.6 Gas Unit menu To access this menu from the Service View menu: 1. Select Modules. 2. Select Gas Unit. 8.6.1 General menu and data To access this menu from the Service View menu: 1. Select Modules. 2. Select Gas Unit. 3. Select General. General Service Data screen Module configuration Shows the available measurement options (the measurement options that are detected by the measurement unit). Timeouts, Bad checksums, and Bad c-s by mod are...
Page 199
Service Menus Counting starts at power on and is reset to zero at power off. Nonzero values do not indicate a failure, but continuous counting (more than 50 per second) indicates a serial communication failure. Also, other measurement units can cause these numbers to rise. 8.6.2 Gases menu and data To access this menu from the Service View menu: 1. Select Modules. 2. Select Gas Unit. 3. Select Gases. Calibrate ID Not used in Cardiocap/5. Noise Meas Activates the noise measurement. Sample...
Page 200
Cardiocap/5 Technical Reference Manual Sample Flow Calculated from differential pressure and adjusted by the measurement unit. Zero—value as measured during initialization when the pump is off. Gain—sample flow measurement can be calibrated by adjusting the gain. Ambient Ambient pressure is measured every 30 min. Amb-Work Ambient pressure—sampling system internal pressure. Fall time CO2 and O2 in ms. For N2O and AA, same as CO2. CO2-O2 Delay In ms. No delay between CO2, N2O, and agents. Pump...