Service Menus 
VO2 Gain Adjusts the VO2 gain. 
Y deadspace Adjusts the Y-deadspace. 
N2 injection Selects On (1) or Off (0); setting is zero (0) during normal operation. 
Record Data Prints the shown service data and board information (id. serial number and 
software id.) to the recorder (N-XREC option). 
Spirometry calibration 
Spirometry calibration is not needed in routine clinical use, but is recommended as part of the yearly 
Planned Maintenance procedures. Use a calibration pump or spirometry...

Page 202

Cardiocap/5 Technical Reference Manual 
8.7 ESTP : ECG service menu and data 
 
To enter this service menu from the 
Service View menu: 
1. Select Modules. 
2. Select ESTP : ECG. 
Power Freq Sets power frequency; 50 Hz/60 Hz. 
Filter Low Sets filter low frequency; 0.05 Hz/0.5 Hz. 
Filter High Sets filter high frequency; 30 Hz (40 Hz if power freq is 60 Hz) /100 Hz or 150 Hz 
Service Data screen 
Power freq and Cable type show the values chosen or detected. 
Filter low and high defines the selected...

Page 203

Service Menus 
8.7.1 ECG Setup menu 
 
To enter this service menu from the 
Service View menu: 
1. Select Modules. 
2. Select ESTP : ECG.  
3. Select ECG Setup.  
Filter Filters the ECG signal high-frequency noise and slow-respiratory artifacts. 
 Monit (monitor) filter is used in routine monitoring. It effectively filters the artifacts 
caused by the electrosurgery unit and respiration.  
 Diagn (diagnostic) filter is used if more accurate waveform information is needed (of 
P-wave or AV block, for...

Page 204

Cardiocap/5 Technical Reference Manual 
8.8 ESTP : STP service menu and data 
 
To enter this service menu 
from the Service View menu: 
1. Select Modules. 
2. Select ESTP : STP. 
Record Data Records, or prints, the service data shown and board information (ID, serial number 
and software ID) on the recorder paper. 
Temp Test Activates the automatic temperature test for temperature channels T1 and T2. The result 
appears in the service data screen. NOTE: Select Temp Test twice to start the test....

Page 205

Service Menus 
Ired int. (infrared intensity) range is 40 to 255 and Red int. (red intensity) range is 40 to 255. 
NOTE: Balance between LEDs is adjusted by changing the infrared/red intensity.  
DC gain shows the adjusted DC signal gain. 
IDC is the value of the infrared signal and RDC is the DC value of the red signal. 
AC gain is the gain of infrared and red AC signals. AC gain values can be 1 (high gain) or 0 (low gain). 
Pre gain is a preamplifier gain for infrared and red signals. Pre-gain values...

Page 206

Cardiocap/5 Technical Reference Manual 
8.9 NIBP service menu, data, and checks 
 
To enter this service menu from 
the Service View menu: 
1. Select Modules. 
2. Select NIBP. 
The submenus shown in this menu (NIBP Demo, Calibrations, etc.) are described later in this section. 
Previous Menu Closes this menu and displays the previous menu.  
Service Data screen 
Pressure Measured pressure multiplied by 10. 
Zero Pressure at auto zeroing multiplied by 10 and changes between +20 and -20 mmHg 
(+2.7 to...

Page 207

Service Menus 
Bad c-s by mod—how many communication errors the module detected. 
Counting starts at power up and resets to zero at power off. Nonzero values do not 
indicate a failure, but continuous counting (more than 50 per second) indicates a 
serial communication failure or that the module is not installed. Other modules can 
cause communication errors that cause these numbers to rise. 
RAM The state of RAM memory. 
ROM Indicates whether the checksum in EPROM is in accordance with the one the...

Page 208

Cardiocap/5 Technical Reference Manual 
8.9.2 NIBP Calibration menu 
 
To enter this service menu 
from the Service View menu: 
1. Select Modules. 
2. Select NIBP.  
3. Select Calibrations. 
Active Leak Test Tests for leaks. 
Calibration Check Checks the calibration.  
For details, see NIBP calibration check in the Service Procedures chapter. 
Protection Software calibration protection (ON/OFF). When calibrating, Protection should 
be OFF. To change the Protection, you must activate the menu item by...

Page 209

Service Menus 
8.9.3  NIBP Safety Valve menu and data 
 
To enter this service menu from 
the Service View menu: 
1. Select Modules. 
2. Select NIBP.  
3. Select Safety Valve. 
Start Test Starts the Safety Valve Test; changes to Stop Test after you start the test. 
Safety Valve Data screen 
See NIBP service menu, data, and checks earlier in this chapter for information on Pressure, Zero, 
Protect handle, Calibr. prot., +15 V power, AD0 to AD7, Timeouts, etc. 
Max. press and 2 s after stop show the...

Page 210

Cardiocap/5 Technical Reference Manual 
8.9.4 NIBP Pulse Valve menu and data 
 
To enter this service menu from 
the Service View menu: 
1. Select Modules. 
2. Select NIBP.  
3. Select Pulse Valve. 
Start Test Starts the Pulse Valve test; changes to Stop Test after you start the test. 
Set Valve Lets you adjust the opening of the pulse valve. 
Pulse Valve Data screen 
See NIBP service menu, data, and checks earlier in this chapter for information on Pressure, Zero, 
Protect handle, Calibr. prot., +15 V...