Page 211

Service Menus 
8.9.5 NIBP Buttons/Leds menu and data 
 
To enter this service menu from 
the Service View menu: 
1. Select Modules. 
2. Select NIBP.  
3. Select Buttons/Leds. 
The Auto ON/OFF, Manual ON/OFF, STAT ON/OFF, and Measur. ON/OFF menu selections have no 
effect. 
Buttons/Leds Data screen 
See NIBP service menu, data, and checks earlier in this chapter for information on Pressure, Zero, 
Protect handle, Calibr. prot., +15 V power, AD0 to AD7, Timeouts, etc. 
The Auto On/Off and Start Cancel...

Page 212

Cardiocap/5 Technical Reference Manual 
8.9.6 NIBP Pneumatics menu and data 
 
To enter this service menu from 
the Service View menu: 
1. Select Modules. 
2. Select NIBP.  
3. Select Pneumatics. 
Start Pump Starts the pump; changes to Stop Pump when the pump is on. 
Open Exh1 Opens exhaust valve 1; changes to Close Exh1 when the valve is open. 
Open Exh2 Opens exhaust valve 2; changes to Close Exh2 when the valve is open. 
Set Valve Adjusts the pulse-valve opening between 0 (fully closed) and 255...

Page 213

Service Menus 
Interval 20 mmHg -> 185 mmHg check 
1. Wrap an adult cuff around a pipe and connect the cuff to the monitor.  
2. Select Start Pump at different combinations of the valves open/closed and push the ComWheel.  
The time it takes for the pressure to go from 20 mmHg to 185 mmHg (2.7 to 24.7 kPa) is counted 
and displayed. When all valves are closed, the pump should pump the pressure in about 1 to 4 
seconds.  
3. To stop the pump, choose Stop Pump and push the ComWheel. 
8.9.7 NIBP Watchdog...

Page 214

Cardiocap/5 Technical Reference Manual 
8.10 NMT menu and data 
 
To access this menu from the Service 
View menu: 
1. Select Modules. 
2. Select NMT. 
 
Start Curr. test A test that checks the difference between a given current and the measured 
current.  
 Connect a 3 kΩ resistance between the stimulus electrodes before starting the 
test. All currents that are checked will be displayed on the service data screen. 
After the current value, the test status is shown.  
 If the test is not passed,...

Page 215

Service Menus 
T1%, etc. The measured response. A value of 1000 corresponds to 100%. Each is a ratio of 
the Absolute T1, etc. to the Reference (Ref). As drugs are administered, the 
typical response for T1 lessens, then drops more rapidly for each successive 
measured response (T2%, T3%, then T4%).  
Ratio% The common label for TOF/DBS%.  
Absolute T1, etc. The voltage measured from the A/D converter. The area-under-the-curve of the 
response to stimulation as the muscle-relaxant anesthetic is being...

Page 216

Cardiocap/5 Technical Reference Manual 
8.10.1 NMT Setup Menu 
 
To access this menu from the Service View menu: 
1. Select Modules. 
2. Select NMT. 
3. Select NMT Setup. 
 
Start-up Starts an NMT measurement. 
Continue Continues an NMT measurement. 
Stimulus Mode Selects the mode of stimulation:  TOF (train of four), DBS (double burst 
stimulation), or ST (single twitch). 
Cycle Time Sets the time intervals for the selected Stimulus Mode. For TOF and DBS, the 
selections are Manual, 10, 12, 15, or 20...

Page 217

Service Menus 
8.11 M-NSAT menu and data (N-XNSAT or N-XOSAT options) 
 
To enter this service menu from the Service 
View menu: 
1. Select Modules. 
2. Select M-NSAT. 
PR Pulse rate value [bpm] calculated from pleth. 
SpO2% Oxygen saturation value multiplied by 100. 
NoProbe 
PulseSearch 
NoPulse 
CheckProbe 
Not applicable to the Cardiocap/5. 
MP-203 Error Yes if an error message is sent by the SpO2 board. 
NOTE:  MP-203 Error is used to represent any version of the Datex-Ohmeda or 
Nellcor SpO2...

Page 218

Cardiocap/5 Technical Reference Manual 
8.12 More Modules menu 
 
To enter this service menu from the Service View menu: 
1. Select Modules. 
2. Select More Modules.  
NOTE: The Interface and Battery selections in the More Modules menu are not used in Cardiocap/5. 
8-32  

Page 219

Service Menus 
8.12.1 Memory Module menu and Module Status data 
 
To enter this service menu from 
the Service View menu: 
1. Select Modules. 
2. Select More Modules.  
3. Select Memory Module. 
Module Status screen 
NOTE: All values can be - - - to indicate “No data is available.” 
Module present Indicates whether the module is firmly attached to the monitor (YES or NO). 
Module active Indicates whether module services are available (YES or NO). 
ROM Status of the memory module’s ROM memory (OK or...

Page 220

Cardiocap/5 Technical Reference Manual 
8.12.2 Memory Module menu and Communication data 
 
To enter this service menu from the 
Service View menu: 
1. Select Modules. 
2. Select More Modules.  
3. Select Memory Module.  
4 Select Communication. 
Communication screen 
Interface status 
The status of the data link between the monitor and the memory module. If memory module is 
properly attached, the status should always be ACTIVE. If status blinks between ACTIVE and 
CLOSED, a communications error has...