Page 61
4 User Maintenance 1006-0939-0004-114-11 Do the steps in the opposite order to assemble the module. Before you use the system, complete the preoperative test procedure. Refer to "Appendix - Preoperative Tests" section of Part 1 of the Operation Manual. Step 4 Push in the tabs and remove the top cover. Step 5 Remove the seals. Turn and pull to remove the check valves. AA.96p.043 AA.96p.044 AA.96p.045
Page 62
Aestiva 4-121006-0939-000 4-12 Bellows maintenance Step 1 Remove the bellows from the breathing system. Step 2 Turn the housing counter-clockwise and lift. Step 3 Remove the bottom edge of the bellows from the rim. AA.96p.023 AA.96p.051 AA.96p.052
Page 63
4 User Maintenance 1006-0939-0004-134-13 wWW W W AA A A RR R R NN N N III I NN N N GG G G Do not disassemble the pressure relief valve. This can damage the seat or diaphragm and cause injury to the patient. Step 4 Remove the disk from the bellows. Step 5 Remove the ring from inside the top of the bellows. Step 6 Push the latch toward the center and remove the rim. Step 7 Remove the pressure relief assembly. AA.96p.031 AA.96p.032 AA.96p.033 AA.96p.034
Page 64
Aestiva 4-141006-0939-000 4-14 Do these steps in the opposite order to assemble the bellows. If you see a dust-like powder on the housing or the bellows, apply a thin layer of KRYTOX lubricant to the ribs of the bellows housing. Make sure the lubricant is applied smoothly and there are no lumps. Make sure that: •The arrow on the seal points up. •You hear a double click when you install the rim. •The rim is locked in position. •The inner ring is correctly installed inside the top of the bellows. •Only...
Page 65
4 User Maintenance 1006-0939-0004-154-15 Bellows tests wWW W W AA A A RR R R NN N N III I NN N N GG G G Objects in the breathing system can stop gas flow to the patient. This can cause injury or death: •Do not use a test plug that is small enough to fall into the breathing system. •Make sure that there are no test plugs or other objects caught in the breathing system. wWW W W AA A A RR R R NN N N III I NN N N GG G G The bellows assembly test does not replace the preoperative tests. Always complete...
Page 66
Aestiva 4-161006-0939-000 4-16 Step 2 Invert the bellows. They must not fall more than 100 mL/min. If it does: •The ports are not tightly sealed. •The bellows is incorrectly installed. •The seal inside the bellows is not correctly installed (with its groove pointed up). •Parts are damaged. Step 3 Remove the plugs from the ports. Permit the bellows to fully extend. Step 4 Close port C. AA.96p.124 C
Page 67
4 User Maintenance 1006-0939-0004-174-17 If the result for all the bellows tests was “passed,” install it in the system. Before you use the system, complete the preoperative test procedure. Refer to "Appendix - Preoperative Tests" section of Part 1 of the Operation Manual. Step 5 The bellows must not fall more than 100 mL/min. If it does: •The port is not tightly sealed. •The bellows or the pressure relief valve is not correctly installed. •Parts are damaged.
Page 68
Aestiva 4-181006-0939-000 4-18 O2 sensor calibration - 21% O2 Note: The 21% O2 calibration must be done before the 100% O2 calibration. During O2 calibration the screen replaces O 2 data with - - -. Mechanical ventilation must be off to perform calibration. This procedure takes three minutes or less. Step 1 Push the menu key. Step 2 Turn the knob to select Setup/Calibration (highlight). Step 3 Push the knob to show the next screen. AB.29.013 AB29.002 AB.90.036 AB.90.037 AB.90.091
Page 69
4 User Maintenance 1006-0939-0004-194-19 The screen shows “Calibrating. . .”, followed by the result (“Complete” or “Failure”). Step 4 Turn, then push the knob to select O 2 Sensor Cal. Step 5 Select 21%. Then, push the knob. Step 6 Complete the steps shown on the screen. • Remove the O 2 sensor from the circuit. •Do not twist or stress the cable. •Make sure the cable is connected. Step 7 Select Start Cal. Then, push the knob. AB29.002 AB.29.012AB.90.048 AB.90.049 AB.48.080
Page 70
Aestiva 4-201006-0939-000 4-20 If the calibration fails: •perform the calibration again. •do a 100% O 2 sensor calibration. If this passes, repeat the 21% O2 calibration. If the calibration passes, install the O 2 sensor. If necessary, do the 100% O2 calibration. Before you use the system, complete the preoperative test procedure. Refer to "Appendix - Preoperative Tests" section of Part 1 of the Operation Manual.