7 Theory of Operation and Specifications 
1006-0939-0007-77-7
Breathing system
schematic
AB.23.106 

Page 112

Aestiva
7-81006-0939-000
7-8
Electrical block diagram
O2 Sensor Flow Sensors Bag/Vent Switch Canister Release Breathing Circuit Module
Absorber Control Panel CO
2 Bypass RS232C
System switch
O2 supply fail and   flush switches
AB.90.089
Tec 6 Outlet 100-120 Vac
220-240 Vac
50/60 Hz
Task Light
Connector Board
Connector Board
Front PanelPneumatic
EngineSensor Interface
Board MicroController
Board BatteryTwo Stage
FilterToroid
AC Inlet Module
- Fuses
- Voltage Selector
Assembly 

Page 113

7 Theory of Operation and Specifications 
1006-0939-0007-97-9
Pneumatic circuits
Gas supplies
(items 1-6)Refer to Figure 7-8 for location of items. Gas goes into the system through a 
pipeline or cylinder connection. All connections have indexed fittings, filters, 
and check (one way) valves. Gauges show the cylinder and pipeline pressures. 
A regulator decreases the cylinder pressures to the appropriate system 
pressure. A pressure relief valve helps protect the system from high pressures.
To help...

Page 114

Aestiva
7-101006-0939-000
7-10
Mixed gas (item 26)The mixed gas goes from the flowmeter outlet through the vaporizer that is ON, 
to the fresh gas outlet, and into the breathing system. A pressure relief valve 
sets the maximum outlet pressure.
Key to Numbered
Components1. Pipeline pressure gauge
2. Pipeline connection
3. Cylinder pressure gauge
4. Cylinder connection
5. Cylinder pressure regulator
6. Pressure relief (opens at approx. 918 kPa)
7. Supply connections for the ventilator
8. System switch
9....

Page 115

7 Theory of Operation and Specifications 
1006-0939-0007-117-11
Figure 7-7 • Pneumatic circuit
AA.96.131 

Page 116

Aestiva
7-121006-0939-000
7-12
Suction regulators (optional)
Venturi Suction
RegulatorSupply: Air or O2 from system gas supply
Drive gas consumption: 75 L/min
Maximum Vacuum: 457 mm Hg with pipeline drive gas at 345 kPa
Minimum flow: 20 L/min
Sound level 53 dBA
Accuracy: 5% of full scale
Continuous
Suction RegulatorSupply: External vacuum
Vacuum levels: 0-200 mm Hg and full line vacuum
Maximum flow: >20 L/min
Accuracy: 5% of full scale
O2 flowmeter (optional)
Supply: O2 from system gas supply
flow rates:...

Page 117

7 Theory of Operation and Specifications 
1006-0939-0007-137-13
Breathing system
There are two types of breathing circuit modules (Figure 7-8):
1. Circle
2. Mapleson D or Bain (With Mapleson D/Bain circuits one flow sensor measures 
Inspiratory and expiratory flows. The other measures fresh gas.)
Tabs on the circuit modules fit into infrared switches on the breathing system bulkhead. 
The ventilator uses these switches to identify the circuit module.  
Direct fresh gas
connections
(common and
auxiliary...

Page 118

Aestiva
7-141006-0939-000
7-14
1.  O
2 sensor (ventilator) Or plug
2. Inspiratory port
3. Tabs (identify the breathing circuit module to ventilator)
4. Expiratory port
5. To-Fro port
6. Fresh gas port
7. Flow sensor module connector (used with ventilator)
8. Optical switches (use tabs to identify circuit module)
9. Internal connection (exhaled gas flows to absorber during expiration)
10. Internal connection (to Bag/Vent switch)
11. Auxiliary Common Gas Outlet (some models)
12. Selector switch (Auxiliary...

Page 119

7 Theory of Operation and Specifications 
1006-0939-0007-157-15
Breathing system specifications
Volume 5500 mL
Absorbent 1.35 kg (400 mL) / canister
Connections    
Breathing circuit modules: Circle or Bain modules
Common gas outlet: ISO 5356 type connector (Standard 22 mm OD or 15 mm ID conical 
friction fit connectors). Remove the breathing circuit module to access.
Auxiliary outlet: ISO 5356 type connector on the front of the breathing system (Standard 
22 mm OD or 15 mm ID conical friction fit...

Page 120

Aestiva
7-161006-0939-000
7-16
Pressure required to open inspiratory and expiratory valves
Dry:          0.49 cmH
2O
Moist: 0.91 cmH
2O
Breathing system leakage (average during use)
APL Valve
Approximately 0 to 70 cmH
2O
Pressure Flow Data (APL Valve Completely Open)
System volume: 5500 mL 
Gas scavengingPassive scavenging:
Positive pressure relief: 10 cmH
2O
Negative pressure relief: 0.5 cmH
2O
Outlet connector: 30 mm male taper swivel
Active scavenging:
Particle filter at the outlet has a pore size of...