GE Aestiva 5 Service Manual
Here you can view all the pages of manual GE Aestiva 5 Service Manual. The GE manuals for Medical Equipment are available online for free. You can easily download all the documents as PDF.
Page 81
5 Alarms and Troubleshooting 1006-0939-0005-55-5 Apnea Alarm StandbyLow Normal condition after End Case, power-up, or ACGO change from On to OffMonitoring resumes after first breath (mechanical) or 2 breaths within 30 seconds (non-mechanical).- - - Apnea Alarm Off Low The cardiac bypass option is selected (alarm limit menu).Apnea alarms are normally turned off when this option is selected.- - - Aux Gas Outlet On Medium (low after acknowledged)The outlet selection switch is set to the...
Page 82
Aestiva 5-61006-0939-000 5-6 Cal Flow Sensors Low The last flow sensor calibration failed.Calibrate the flow sensors. Look for water in the flow sensor tubes. Dry if necessary.Contact a qualified service representative. Calibrate O 2 SensorLow O 2%>110% Does the sensor measure 21 % O 2 in room air?Calibrate O 2 sensor. Canister open Medium The canister release is open (large leak) without a bypass mechanism.Close the canister release. A switch checks the release position. Cannot Drive...
Page 83
5 Alarms and Troubleshooting 1006-0939-0005-75-7 Control Settings Input has FailedMinimum monitoring (Medium)Ventilator malfunction. Ventilate manually. Monitoring is still available.Contact a qualified service representative. CPU Failure Minimum shutdown (High)Ventilator malfunction. Ventilate manually. Monitoring is not reliable.Contact a qualified service representative. CPU Internal Error Minimum shutdown (High)Ventilator malfunction. Ventilate manually. Monitoring is not...
Page 84
Aestiva 5-81006-0939-000 5-8 Heliox Mode is On Low Control setting on ventilation setup menu.When Heliox is used, the ventilator must adjust volume calculations.- - - High O 2Medium O2% > alarm high limit setting.Is the limit set correctly? What is the O 2 flow? Did you just push Flush? Does the sensor see 21% O 2in room air?Calibrate O 2 sensor. Replace O 2 sensor. High Paw High Paw is greater than Plimit. The ventilator cycles to expiration.Are Plimit and other controls set correctly? Look...
Page 85
5 Alarms and Troubleshooting 1006-0939-0005-95-9 Internal Ventilator Clock Too FastMinimum shutdown (High)Ventilator malfunction. Ventilate manually. Monitoring is not reliable.Contact a qualified service representative. Internal Ventilator Clock Too SlowMinimum shutdown (High)Ventilator malfunction. Ventilate manually. Monitoring is not reliable.Contact a qualified service representative. Invalid Circuit ModuleLow The system does not recognize the type of circuit module installed....
Page 86
Aestiva 5-101006-0939-000 5-10 Low Paw Medium Paw does not rise at least 4 cm from the lowest pressure measured during the last 20 seconds.Are circuit connections OK? Look at the Paw gauge on the absorber.Look for circuit disconnection. Low Ve Medium Exhaled minute volume
Page 87
5 Alarms and Troubleshooting 1006-0939-0005-115-11 Monitoring Only Medium A severe malfunction prevents mechanical ventilation. Other alarms may also occur. Ventilate manually. Cycle system power (On- Standby- On). If the alarm clears, restart mechanical ventilation.Contact a qualified service representative. No Circuit Module Low The patient circuit module is not installed.Install a module. Refer to the Setup section.Optical sensors look for tabs on the back of the module. Is the module...
Page 88
Aestiva 5-121006-0939-000 5-12 O2 Sensor out of circLow O2 sensor not installed in breathing circuit module; sensor not measuring gas in breathing circuit.Install a breathing circuit module and an O 2 sensor.- - - On Battery - Power OK?Medium (low after acknowledged)The mains supply is not connected or has failed and the system is using battery power.Ventilate manually to save power. At full charge, the battery permits approx. 30 minutes of mechanical ventilation.Make sure power is...
Page 89
5 Alarms and Troubleshooting 1006-0939-0005-135-13 Pres/Vol Mon InactiveMedium Outlet selection switch is set to auxiliary gas outlet.Connect the patient circuit to the auxiliary gas outlet or set the switch to the common gas outlet for normal operation.- - - Pressure Limit Switch FailureMinimum monitoring (Medium)A pressure safety switch activated at a Paw
Page 90
Aestiva 5-141006-0939-000 5-14 Sustained Paw High Paw > sustained pressure limit for 15 seconds 3.Check tubing for kinks, blockages, disconnects.Calibrate the flow sensors. System Leak? Medium Delivered volumes do not match set volumes.If you are using Heliox, select Heliox on the ventilator setup menu. Look for leaks in the manifold. Compare set to delivered volumes.Calibrate the flow sensors. Drain water buildup from the breathing system. Vaux_ref Out-of- RangeMinimum shutdown...