2 Cleaning and Sterilization 1006-0939-0002-17 2-17 SS S S ttt t ee e e pp p p 66 6 6 Reinstall the canisters: a. Ensure seals and rim are free of soda lime dust. b. Place the canisters in the drain dish. c. Turn the canister release handle counterclockwise to its locked position while gently applying pressure on the canisters to hold them against the canister guides. SS S S ttt t ee e e pp p p 77 7 7 Test the Bag circuit for leaks according to the Breathing System tests in the Appendix...
Page 33
1006-0939-0003-1 3-1 III I mm m m pp p p oo o o rrr r ttt t aa a a nn n n ttt t Datex-Ohmeda strongly recommends that you use O2 monitoring with this equipment. Refer to local standards for mandatory monitoring. III I mm m m pp p p oo o o rrr r ttt t aa a a nn n n ttt t European Standard EN 740 requires anesthetic agent monitoring when anesthetic vaporizers are in use. III I mm m m pp p p oo o o rrr r ttt t aa a a nn n n ttt t European Standard EN 740 requires CO 2 monitoring during ventilation. ww w...
Page 34
Aestiva 3-21006-0939-000 3-2 Breathing system setup To assemble the patient circuit, start at step 5. Step 1 Install the main manifold. Step 2 Install the Bellows. •Make sure the release button aligns correctly. Step 3 Gently close the control panel. Step 4 Install the bag arm. • Install a rebreathing bag on the bag arm. AA.96p024 AA.96p022 AA.96p012 AA.96p019
Page 35
3 Setup and Connections 1006-0939-0003-33-3 Step 5 Identify the breathing circuit module (Circle or Mapleson). Look at the tabs on the rear or the label on the front. Step 6 Install the breathing circuit module. You will feel some resistance. Continue to push. Step 7 Completely tighten the thumbscrew. Step 8 Install the O2 sensor. •Completely tighten the sensor. •Connect the cable. Mapleson CircleAB.23.211 AA.96p017 AA.96p016 AA.96p013, 014
Page 36
Aestiva 3-41006-0939-000 3-4 Step 9 Install the flow sensor module. •You will hear a “click” when it locks into position. •Refer to Note 1 or the User Maintenance section. Step 10 Close the front panel. Step 11 Connect the patient circuit. AA.96p011 AA.96p009 Circle Module Mapleson D / Bain Module AB.23.104 AB.23.152 To-fro
Page 38
Aestiva 3-61006-0939-000 3-6 Canister setup Each canister holds 1.35 kg of loose absorbent or 1.13 kg of commercially packaged (pre-pack) absorbent. Datex-Ohmeda recommends Sodium or Bar- ium hydroxide based absorbent. w WW W W AA A A RR R R NN N N III I NN N N GG G G Change absorbent often to prevent the build up of non-metabolic gases when the system is not in use. Inspect absorbent at the end of a case. During non-use, absorbent can go back to the original appearance. Refer to the absorbent...
Page 39
3 Setup and Connections 1006-0939-0003-73-7 Step 3 Push in the buttons and install the top dish and drain dish. Pull down to make sure the top dish is locked in place. Step 4 Fill the canisters with absorbent and install them. Push down on the drain dish if necessary. Step 5 Turn the release counter-clockwise. Step 6 Push in the release. AA.96p054 AA.96p029 AA.96p028 AA.96p026
Page 40
Aestiva 3-81006-0939-000 3-8 Pneumatic and electrical connections wWW W W AA A A RR R R NN N N III I NN N N GG G G Equipment connected to the electrical outlets can increase the leakage current. Regularly test the leakage current. wCC C C AA A A UU U U TT T T III I OO O O NN N N Use only medical grade gas supplies. Other types of gas supplies may contain water, oil, or other contaminants. The Aestiva gas supplies also supply these devices through internal connec- tions: •The venturi suction regulator...