GE Aestiva 5 Service Manual
Here you can view all the pages of manual GE Aestiva 5 Service Manual. The GE manuals for Medical Equipment are available online for free. You can easily download all the documents as PDF.
Page 51
1006-0939-0004-1 4-1wWW W W AA A A RR R R NN N N III I NN N N GG G G TO PREVENT FIRES: •Use lubricants approved for anesthesia or O 2 equipment, such as Krytox®1. •Do not use lubricants that contain oil or grease. They burn or explode in high O 2 concentrations. •All covers used on the system must be made from antistatic (conductive) materials. Static electricity can cause fires. wWW W W AA A A RR R R NN N N III I NN N N GG G G Obey infection control and safety procedures. Used equipment may contain...
Page 52
Aestiva 4-21006-0939-000 4-2 Repair policy Do not use malfunctioning equipment. Make all necessary repairs or have the equipment serviced by an authorized Datex-Ohmeda service representative. After repair, test the equipment to ensure that it is functioning properly, in accordance with the manufacturer’s published specifications. To ensure full reliability, have all repairs and service done by an authorized Datex- Ohmeda service representative. If this cannot be done, replacement and maintenance of...
Page 53
4 User Maintenance 1006-0939-0004-34-3 Manifold maintenance Step 1 Complete the basic disassembly (Refer to the Cleaning and Sterilization section). Step 2 (APL valve) 1 Remove the diaphragm. Replace it if: •The disk, cage, or diaphragm is damaged. •The diaphragm looks worn. •The diaphragm fits over the rim. Step 3 (Seals) Loosen the thumbscrew and remove the plate. • Replace the seal if damaged or worn. AB.23.094 AB.23.069 AB.23.070
Page 54
Aestiva 4-41006-0939-000 4-4 Step 4 (Seals con.) Loosen the thumbscrews. Step 5 (Seals con.) Open the manifold (top middle, and bottom parts). Step 6 (Seals con.) Replace the parts if they are damaged or worn. AB.23.073 AB.23.074 Side TopBottom AB.23.071
Page 55
4 User Maintenance 1006-0939-0004-54-5 Step 7 (Bag/Vent)2 Replace the parts if they are damaged or worn. Step 8 Put the parts together •Make sure all gaskets are fully seated. •Fully tighten the thumbscrews. 1.When you turn the APL knob, you change the pressure that is necessary to push the disk up and permit gas to exit. 2. The Bag/Vent switch pushed down plungers to close the other path. In the Bag position, the vent path is closed, etc. Before you use the system, complete the preoperative test...
Page 56
Aestiva 4-61006-0939-000 4-6 Expiratory valve maintenance Disassemble the manifold and examine the seals at 12 autoclave cycles. These conditions indicate a possible leak in the exhalation valve: •A decrease in airway pressure during the inspiratory pause in the volume control mode of ventilation. •Alarms for low tidal volume. •The ventilator circuit has a leak, but the bellows pass the bellows tests in this section. Step 1 Remove the exhalation valve (Cleaning and Sterilization section). Step 2...
Page 57
4 User Maintenance 1006-0939-0004-74-7 Before you use the system, complete the preoperative test procedure. Refer to "Appendix - Preoperative Tests" section of Part 1 of the Operation Manual. Step 3 Unscrew the lock ring (one quarter turn) and remove the cap. Step 4 Put the parts together •Align the dimples in the cap with the valve seat. •To install the top, align it with the base and push down. AB.23.072 Dimples AB.23.021
Page 58
Aestiva 4-81006-0939-000 4-8 Receiver maintenance (active gas scavenging only) A blocked filter decreases flow through the scavenging system. In extreme cases this can open the high-pressure relief valve and vent to room air. Note: Passive gas scavenging does not require maintenance. w WW W W AA A A RR R R NN N N III I NN N N GG G G This system operates correctly at the electrical interference levels of IEC 601-1-2. Higher levels can cause nuisance alarms that may stop mechanical ventilation. •Always...
Page 59
4 User Maintenance 1006-0939-0004-94-9 Flow sensor maintenance Before you use the system, complete the preoperative test procedure. Refer to "Appendix - Preoperative Tests" section of Part 1 of the Operation Manual. Step 1 Remove the flow sensor module (Cleaning and Sterilization section). Push in the tab and remove the cover. Step 2 Install the new sensor with the arrows up. Keep the tubes straight. Step 3 Slide on the cover. Do not pinch the tubes. AA.96p.037 AA.96p.039 AA.96p.038
Page 60
Aestiva 4-101006-0939-000 4-10 Breathing circuit maintenance Disassemble the patient circuit module to replace check valves or seals. Step 1 Remove the patient circuit module (Cleaning and Sterilization section). Step 2 Hold in the tab. Step 3 Remove the rear cover. Note: The rear seal is keyed to fit a specific module. AA.96p.017 AA.96p.041 AA.96p.042